Fetal MRI is a specialized prenatal imaging technique that uses magnetic resonance imaging to evaluate a developing fetus when ultrasound findings are inconclusive or when more detailed anatomical information is required. Unlike ultrasound, which relies on sound waves, fetal MRI uses strong magnetic fields and radiofrequency pulses to create high-resolution, multiplanar images of fetal organs, the placenta, and surrounding maternal structures without ionizing radiation. It has become an essential second-line tool in maternal-fetal medicine.

The technique gained widespread clinical use in the late 1990s after the development of ultrafast single-shot sequences such as HASTE and SSFSE, which freeze fetal motion in a fraction of a second. Before these sequences existed, fetal motion blurred almost every image, making MRI of the unborn child impractical. Today, modern 1.5T scanners produce diagnostic-quality fetal images in scan times short enough to complete a full study in 30 to 45 minutes without sedation.

Most fetal MRI examinations are performed after 18 weeks gestation, with the majority scheduled between 20 and 32 weeks when fetal anatomy is sufficiently developed and there is still adequate amniotic fluid for tissue contrast. Earlier scans are technically possible but yield smaller anatomic structures, while late-third-trimester imaging can be limited by crowding and decreased fluid volume. Timing is always tailored to the specific clinical question.

Common indications include suspected central nervous system anomalies, congenital diaphragmatic hernia, complex genitourinary malformations, neck masses that may compromise the airway at delivery, and abnormalities of the placenta such as suspected placenta accreta spectrum. Fetal MRI also plays a critical role in surgical planning for in-utero procedures and for delivery teams preparing an EXIT (ex-utero intrapartum treatment) procedure. For broader context, see our overview of the history of MRI.

Safety is the foremost concern for expectant parents. Extensive clinical experience and multiple professional society statements support the safety of non-contrast fetal MRI at 1.5T after the first trimester. Gadolinium-based contrast agents, however, are generally avoided during pregnancy because they cross the placenta and have been associated with adverse outcomes in some observational studies. The decision to image is always weighed against the diagnostic benefit and the alternative of repeat ultrasound.

The mother enters the magnet feet-first, lies supine or in a left lateral decubitus position to reduce caval compression, and breathes normally. No special preparation, fasting, or sedation is typically required. A phased-array body coil is wrapped over the abdomen, and the technologist communicates with the patient throughout the exam. The acoustic noise is buffered by ear protection, and the room is kept cool and dim for comfort.

Because fetal MRI is operator- and protocol-dependent, results are best interpreted by radiologists with specific fetal imaging fellowship training, working in concert with maternal-fetal medicine specialists. The combined report often guides decisions about delivery location, neonatal subspecialty involvement, and prenatal counseling. Understanding what fetal MRI can and cannot show helps parents engage meaningfully in these conversations during a stressful time.

Fetal MRI by the Numbers

โฑ๏ธ
30-45 min
Typical Scan Duration
๐Ÿ“…
18-32 wks
Optimal Gestational Window
๐Ÿงฒ
1.5 Tesla
Standard Field Strength
๐Ÿ“Š
30-50%
Add Info Beyond Ultrasound
๐Ÿ›ก๏ธ
0
Ionizing Radiation
Test Your Fetal MRI Knowledge โ€” Free MRI Quiz

Top Clinical Indications for Fetal MRI

๐Ÿง  CNS Anomalies

Ventriculomegaly, agenesis of the corpus callosum, posterior fossa malformations, cortical development disorders, and intracranial hemorrhage are the most common reasons for referral, accounting for nearly half of all fetal MRI studies performed worldwide.

๐Ÿซ Chest & Airway

Congenital diaphragmatic hernia, CPAM, bronchopulmonary sequestration, and cervical masses threatening the airway are evaluated to plan delivery strategy, including EXIT procedures and immediate neonatal surgical intervention.

๐Ÿฉบ Abdomen & GU

Complex urinary tract dilation, bowel obstruction, abdominal wall defects, and ambiguous genitalia benefit from MRI when ultrasound is limited by maternal habitus, fetal position, or oligohydramnios that obscures visualization.

๐Ÿ“‹ Placenta

Placenta accreta spectrum and vasa previa are increasingly imaged with MRI in high-risk pregnancies to map the depth of myometrial invasion, helping surgical teams plan for cesarean hysterectomy or uterine-sparing approaches.

๐Ÿ”ฌ Fetal Surgery Planning

Open myelomeningocele repair, fetoscopic spina bifida closure, and twin-twin transfusion syndrome cases rely on MRI for precise anatomic mapping before in-utero procedures performed at specialized fetal therapy centers.

Fetal MRI exploits the same physical principles as adult MRI: hydrogen protons within tissue water align with a strong external magnetic field, then emit radiofrequency signals that are spatially encoded by gradient coils to form an image. The challenge is that the fetus moves unpredictably, and even small motion blurs traditional sequences. Engineers solved this by developing single-shot techniques that acquire an entire slice in under a second, effectively freezing motion.

The workhorse sequence is single-shot fast spin echo, known as HASTE on Siemens systems and SSFSE on GE platforms. A complete T2-weighted slice is acquired in roughly 400 to 1000 milliseconds, which is faster than the typical fetal kick. Stacks of these images are obtained in axial, coronal, and sagittal planes relative to the fetal anatomy of interest, not maternal anatomy, requiring real-time replanning by the technologist between acquisitions.

Balanced steady-state free precession sequences, such as TrueFISP or FIESTA, provide bright-blood and bright-fluid contrast that highlights vascular structures, amniotic fluid, and the heart. Echo-planar imaging gives susceptibility contrast useful for detecting hemorrhage or calcification. Diffusion-weighted imaging assesses microstructure in the developing brain and characterizes lesions when bleeding or ischemia is suspected.

Most centers use 1.5T magnets because they offer the best balance of signal-to-noise ratio, specific absorption rate margins, and image uniformity for body imaging during pregnancy. 3T scanners are used selectively, especially for high-resolution brain imaging, but they generate more heat, suffer from more susceptibility artifact, and have stricter SAR limits that can extend total scan time. Field strength choice is institution-dependent.

Coil selection matters as much as field strength. A combination of anterior and posterior phased-array coils is wrapped around the maternal abdomen to provide uniform signal across the gravid uterus. Older body coils alone produce inferior images, particularly in the late third trimester or in patients with elevated BMI. Modern receive arrays with parallel imaging acceleration cut scan time and reduce motion-related artifacts. For comparison with other contrast strategies, see MRI with and without contrast.

Patient positioning is critical for both comfort and image quality. After roughly 20 weeks, supine positioning can compress the inferior vena cava and reduce cardiac output, causing maternal nausea or hypotension. Many protocols default to left lateral decubitus or use a wedge under the right hip. The technologist confirms maternal comfort before each acquisition and is prepared to pause and reposition immediately if symptoms develop.

Acoustic noise is a frequent parental concern. The gradient coils generate sound levels of 100 to 110 decibels during certain sequences, but the maternal abdominal wall and amniotic fluid attenuate this significantly by the time it reaches the fetus. Studies measuring intrauterine sound pressure confirm fetal exposure remains below thresholds associated with hearing damage. Ear protection is provided for the mother throughout the examination.

FREE MRI Knowledge Questions and Answers
Test core MRI knowledge including fetal imaging principles, sequences, and clinical applications.
FREE MRI Physics Questions and Answers
Practice MRI physics concepts including single-shot sequences, SAR, and field strength considerations.

Fetal MRI Sequences Explained

๐Ÿ“‹ T2 Single-Shot

T2-weighted single-shot fast spin echo is the foundation of every fetal MRI protocol. Acquired in under a second per slice, it provides excellent contrast between amniotic fluid, fetal soft tissues, and dense structures like bone. Pathologies such as ventriculomegaly, cystic lung lesions, and bowel dilation appear strikingly bright on these images because they contain fluid, which has a long T2 relaxation time.

Radiologists typically obtain three orthogonal stacks of T2 images aligned to the fetus rather than to the mother. Slice thickness ranges from 3 to 5 millimeters, with no interslice gap. The technologist replans each stack after reviewing the previous one, because the fetus may shift position between acquisitions. Total scan time for T2 imaging usually runs 15 to 20 minutes across all planes and regions of interest.

๐Ÿ“‹ T1 & Echo-Planar

T1-weighted imaging is reserved for specific questions: detection of fat-containing lesions such as sacrococcygeal teratomas, identification of meconium in the distal bowel, or characterization of hemorrhage. Because T1 sequences are inherently slower, gradient-recalled echo techniques with breath-hold or fast spoiled sequences are typically used. Image quality is more vulnerable to fetal motion than T2, so acquisitions are kept brief and targeted.

Echo-planar imaging provides susceptibility-weighted contrast in just a few seconds per stack. It is highly sensitive to blood breakdown products, making it ideal for detecting intracranial or intraventricular hemorrhage in the fetal brain. EPI also serves as the backbone for diffusion-weighted imaging, which adds microstructural information particularly valuable when ischemic injury or developmental abnormalities of the cortex are suspected during third-trimester evaluation.

๐Ÿ“‹ Balanced SSFP

Balanced steady-state free precession sequences, marketed as TrueFISP, FIESTA, or bSSFP depending on the vendor, produce striking bright-blood and bright-fluid contrast. They are useful for evaluating vascular anatomy, the fetal heart, and complex cystic lesions where flow and fluid characterization matter. Acquisitions complete in under a second, and the cinematic quality of these images often helps surgical teams visualize spatial relationships.

However, bSSFP is sensitive to magnetic field inhomogeneity and can produce dark band artifacts at tissue interfaces. It is therefore used as a complementary tool rather than a primary sequence. In placenta accreta spectrum imaging, bSSFP highlights abnormal vascular channels and bulging contours of the placenta into the bladder or pelvic sidewall, complementing T2 findings of dark intraplacental bands and disorganized architecture.

Fetal MRI vs Prenatal Ultrasound: Strengths and Limitations

Pros

  • Superior soft-tissue contrast for brain, lung, and placental tissue characterization
  • Large field of view captures entire fetus and maternal pelvis in a single image
  • Not limited by maternal body habitus, oligohydramnios, or fetal position
  • Provides multiplanar reformatting without dependence on operator skill at acquisition
  • No ionizing radiation, considered safe after the first trimester
  • Adds clinically significant information in 30 to 50 percent of complex CNS cases

Cons

  • Substantially higher cost than ultrasound, often not covered without strict indication
  • Limited availability outside tertiary academic medical centers and fetal therapy programs
  • Cannot be used in patients with non-MRI-conditional implants or severe claustrophobia
  • Image quality degraded by excessive fetal motion despite ultrafast sequences
  • Gadolinium contrast generally avoided, limiting characterization of some lesions
  • Requires specialized training to interpret accurately, increasing risk of misinterpretation
FREE MRI Registry Questions and Answers
ARRT-style registry practice covering fetal imaging safety, screening, and protocol considerations.
MRI MRI Anatomy and Pathology
Practice identifying fetal anatomy and common pathology patterns seen on MRI examinations.

Patient Preparation Checklist for Fetal MRI

Confirm gestational age is at least 18 weeks before scheduling whenever possible
Complete the standard MRI safety screening form including all implants and prior surgeries
Remove all metallic objects including jewelry, hairpins, underwire bras, and body piercings
Wear comfortable two-piece clothing without metal zippers, or change into a hospital gown
Empty the bladder immediately before the scan to improve comfort during the examination
Eat a light meal beforehand to prevent nausea, but avoid heavy foods that may cause reflux
Inform the technologist about any history of claustrophobia or anxiety before entering the room
Position the patient in left lateral decubitus if supine causes maternal symptoms during scanning
Use double hearing protection with earplugs and headphones to minimize acoustic exposure
Have a partner or support person available in the control room when institutional policy allows
Roughly one in three complex CNS cases gets a new diagnosis

Published series consistently show that fetal MRI alters the diagnosis or adds clinically actionable information in approximately 30 to 50 percent of fetuses referred for suspected central nervous system anomalies. In a subset of these cases, the new finding directly changes prenatal counseling, delivery planning, or decisions about fetal intervention. This is why MRI has shifted from optional adjunct to standard of care in many high-volume fetal medicine centers across North America and Europe.

Safety in pregnancy is the most common question expectant parents ask about fetal MRI, and the answer is reassuring but nuanced. The American College of Radiology, the American College of Obstetricians and Gynecologists, and the International Society for Magnetic Resonance in Medicine have all issued statements supporting the use of non-contrast MRI at 1.5T during any stage of pregnancy when the diagnostic information cannot be obtained safely with ultrasound and clinical management would be affected.

The first trimester historically received more caution because organogenesis is occurring and the long-term effects of magnetic field exposure during this period have been less studied. In practice, fetal MRI before 18 weeks is rarely performed, partly because the structures of interest are too small to resolve and partly because clinicians prefer to defer until after the most vulnerable embryologic window. A large 2016 cohort study in JAMA found no increased risk of adverse outcomes with first-trimester MRI exposure.

Specific absorption rate, or SAR, measures radiofrequency energy deposition that can heat tissue. Pregnant patients are scanned in normal operating mode, which limits whole-body SAR to 2 watts per kilogram. Modern scanners monitor SAR continuously and automatically adjust sequence parameters if limits are approached. Maternal core temperature does not measurably rise during clinical fetal MRI, and intrauterine temperature remains within physiologic ranges based on phantom and animal studies.

Gadolinium-based contrast agents are a separate safety consideration. These agents cross the placenta, enter the amniotic fluid, are swallowed by the fetus, re-excreted into the amniotic cavity, and recirculate over hours to days. A 2016 JAMA study reported an association between gadolinium exposure during pregnancy and a small increased risk of stillbirth, neonatal death, and rheumatologic conditions in childhood. Most centers therefore avoid gadolinium unless the benefit clearly outweighs the risk.

Acoustic noise during MRI raises understandable concern about fetal hearing. Measurements taken with hydrophones in ex vivo preparations and modeled in pregnant phantoms show that the amniotic fluid and abdominal wall attenuate sound by roughly 30 decibels by the time it reaches the fetal ear. The result is exposure well below the 90-decibel threshold associated with hearing risk. Follow-up audiologic studies in children scanned in utero have not demonstrated hearing impairment.

Claustrophobia and anxiety are practical maternal safety issues. Up to 10 percent of patients experience significant discomfort inside the bore, and a smaller proportion cannot complete the examination. Strategies include thorough pre-scan education, use of wide-bore 70-centimeter magnets, prism glasses to see outside the bore, music through MR-compatible headphones, and the presence of a support person. Pharmacologic anxiolysis is generally avoided during pregnancy.

Finally, screening is essential. Maternal implants, cardiac devices, and prior surgical history must be reviewed against the manufacturer's MRI conditional labeling before the patient enters the room. Some older devices remain contraindicated. Many newer implants are MR-conditional at 1.5T but not 3T. The screening process for pregnant patients is no different from any other MRI candidate and is the responsibility of trained MR safety personnel.

Interpretation of fetal MRI requires more than general radiology training. The developing brain, lung, and skeleton look fundamentally different from postnatal anatomy at every gestational age, and what is normal at 22 weeks may be pathologic at 32 weeks. Reference atlases such as those by Glenn, Griffiths, and Prayer have become essential desk references for fetal radiologists, providing week-by-week comparisons of normal development across organ systems and imaging planes.

The report typically opens with biometry: biparietal diameter, head circumference, abdominal circumference, and femur length, compared against gestational age. These measurements are then cross-referenced with the patient's last menstrual period or earlier ultrasound dating to confirm consistency. Discrepancies prompt review of the referring obstetric history and may suggest growth restriction, macrosomia, or dating uncertainty that affects subsequent measurements throughout the examination.

Organ-by-organ assessment follows a structured template. The fetal brain is reviewed for ventricular size, cortical development, posterior fossa anatomy, midline structures, and signal abnormalities. The chest evaluates lung volumes, mediastinal position, and diaphragmatic integrity. The abdomen surveys bowel, kidneys, bladder, and abdominal wall. The spine is screened for closure defects, and the extremities for major limb anomalies when clinically relevant.

The placenta and amniotic fluid receive specific attention. Placental thickness, location, signal characteristics, and any abnormal vascular features are described, especially in patients with prior cesarean deliveries who are at risk for accreta spectrum. Amniotic fluid volume is noted qualitatively or with a deepest pocket measurement, because oligohydramnios changes the relevance of certain fetal findings such as renal anomalies. For related reading, see common MRI findings.

Results are communicated through the standard radiology report but, more importantly, through direct discussion with the maternal-fetal medicine team. Many centers hold weekly fetal imaging conferences where radiologists, MFM specialists, neonatologists, geneticists, and pediatric surgeons review cases together. This multidisciplinary format prevents single-discipline interpretation bias and ensures that complex findings translate accurately into prenatal counseling and delivery plans.

Turnaround times vary by institution. Routine fetal MRI reports are typically finalized within 24 hours, while urgent studies performed for delivery planning or evaluation of acute fetal compromise are read in real time with the technologist and ordering provider. The radiologist is often available at the scanner for protocol modifications during the acquisition itself, especially when unexpected findings emerge that warrant additional sequences.

Patients receive their results from the ordering clinician, not the radiologist directly, in keeping with standard prenatal care practice. This allows the maternal-fetal medicine specialist to contextualize imaging findings within the broader clinical picture, including ultrasound history, family history, genetic testing results, and the patient's own values and goals for the pregnancy. Imaging is one input among several in shared decision-making.

Sharpen Your MRI Physics Skills โ€” Free Practice Test

Preparing for a fetal MRI as a patient is straightforward, but a few practical tips can substantially improve the experience. Wear comfortable clothing without metallic fasteners, and consider bringing a change of clothes if you prefer not to use a hospital gown. Hydrate normally in the hours before the exam but empty your bladder right before entering the scanner. Eat a light meal beforehand, because lying still on an empty stomach can trigger nausea that disrupts imaging.

For technologists, the most important practical tip is to plan each sequence relative to the fetus and not the mother. The maternal pelvis is fixed, but the fetus can rotate between acquisitions. Localizers should be reacquired frequently, and three-plane stacks should be aligned to the specific fetal organ of interest. Communicating with the patient between sequences about her comfort and any sensations also helps build trust and reduces motion from anxiety.

Radiologists new to fetal MRI benefit from beginning with high-volume CNS cases under mentorship before progressing to body, placental, and surgical-planning examinations. Reading fetal MRI is pattern recognition built on a foundation of normal developmental anatomy that takes time to internalize. Daily review of normal cases alongside the pathologic cases is essential, because the eye must first recognize normal cortical lamination, sulcation, and ventricular contour before subtle abnormalities become apparent.

Referring clinicians can improve the value of every fetal MRI by providing a focused clinical question. A request for evaluation of suspected agenesis of the corpus callosum following ventriculomegaly seen on ultrasound is far more useful than a generic request for fetal brain imaging. The protocol can then be tailored, additional sequences added as needed, and the radiologist can address the specific question directly in the impression of the report. Reviewing the MRI medical abbreviation glossary can help non-imaging clinicians follow detailed reports.

Trainees preparing for radiology boards or MRI registry exams should understand the core sequences used in fetal imaging, the rationale for 1.5T over 3T in most cases, the safety profile of gadolinium during pregnancy, and the most common indications. Questions frequently address single-shot fast spin echo as the workhorse sequence, the avoidance of gadolinium, and the role of MRI as a problem-solving tool after ultrasound rather than a primary screening modality.

Quality improvement in fetal MRI programs depends on tracking key metrics: study completion rate, rate of non-diagnostic examinations, turnaround time, and concordance between MRI findings and postnatal imaging or surgical pathology. Programs that audit these metrics regularly identify protocol gaps, training needs, and equipment limitations. Such audits also support credentialing of new radiologists entering fetal imaging practice and accreditation by professional bodies that increasingly recognize fetal MRI as a subspecialty.

Finally, expectations should be set carefully with families before the examination. Fetal MRI is a powerful diagnostic tool, but it has limitations: it cannot make every prenatal diagnosis definitive, and findings sometimes raise more questions than they answer. Patients benefit from understanding in advance that the result may add clarity, add complexity, or remain inconclusive. Honest framing builds trust and supports the difficult decision-making that often follows complex prenatal findings.

MRI MRI Anatomy and Pathology 2
Continue practicing anatomy and pathology recognition across MRI sequences and body regions.
MRI MRI Anatomy and Pathology 3
Advanced anatomy and pathology questions including fetal, neuro, and musculoskeletal MRI cases.

MRI Questions and Answers

Is fetal MRI safe for my baby?

Non-contrast fetal MRI at 1.5T is considered safe after the first trimester and is endorsed by the American College of Radiology and ACOG when clinically indicated. There is no ionizing radiation, and decades of clinical experience have shown no measurable harm to the fetus. Most centers avoid imaging before 18 weeks, not because of proven risk but because the structures of interest are too small to evaluate meaningfully during early pregnancy.

At what gestational age is fetal MRI performed?

The majority of fetal MRI examinations are performed between 20 and 32 weeks gestation. This window provides sufficient fetal size for detailed anatomic evaluation while preserving enough amniotic fluid for tissue contrast. Earlier studies are technically possible but yield smaller anatomic structures, and late third-trimester scans may be limited by crowding and reduced amniotic fluid. Timing is always individualized to the specific clinical question and the urgency of management decisions.

How long does a fetal MRI take?

A typical fetal MRI takes 30 to 45 minutes, depending on the clinical question and the cooperation of the fetus. CNS studies tend to be shorter, while complex body or placental imaging may take longer. Single-shot ultrafast sequences allow each slice to be acquired in under a second, so the bulk of scan time is spent repositioning, replanning, and obtaining multiple orthogonal stacks tailored to the fetal anatomy rather than the maternal pelvis.

Do I need contrast for a fetal MRI?

In the vast majority of cases, no. Gadolinium-based contrast agents cross the placenta and have been associated with adverse outcomes in some observational studies, so they are generally avoided during pregnancy. Fetal MRI relies on the intrinsic contrast of fetal tissues, amniotic fluid, and the placenta on T2-weighted and other sequences. Contrast is reserved for very specific maternal indications, such as evaluation of suspected maternal malignancy, after multidisciplinary discussion and informed consent.

Will I need sedation for fetal MRI?

Maternal sedation is rarely used in modern fetal MRI. Single-shot sequences are fast enough to overcome most fetal motion, and maternal anxiety is usually managed with thorough pre-scan education, support persons, and comfort measures inside the bore. Fetal sedation through maternal administration of medications was used historically but is now largely abandoned because ultrafast imaging eliminated the need. Patients with severe claustrophobia are evaluated case by case, sometimes with mild anxiolysis under MFM guidance.

How does fetal MRI compare to ultrasound?

Ultrasound remains the primary prenatal imaging modality because it is portable, real-time, low-cost, and excellent for most evaluations. Fetal MRI complements ultrasound when findings are inconclusive, when maternal habitus or fetal position limits visualization, or when soft-tissue characterization is needed. MRI provides superior contrast for brain, lung, and placental tissues. The two modalities are not interchangeable, and most fetal MRI requests follow a detailed ultrasound that identified a concerning or complex finding.

Can fetal MRI diagnose Down syndrome?

Fetal MRI cannot diagnose Down syndrome or other chromosomal abnormalities. Genetic diagnoses require karyotyping or microarray analysis from amniocentesis, chorionic villus sampling, or noninvasive prenatal screening of maternal blood. MRI may identify structural anomalies sometimes associated with chromosomal conditions, such as cardiac defects or duodenal atresia, which can prompt genetic counseling. However, the imaging study itself is anatomic, not genetic, and a normal MRI does not rule out chromosomal disorders.

What is the cost of a fetal MRI?

Fetal MRI costs vary widely by region, institution, and insurance coverage, but in the United States the cash price typically ranges from 1,500 to 4,000 dollars. Insurance coverage depends on the documented clinical indication and prior ultrasound findings. Most insurers cover the study when ordered by a maternal-fetal medicine specialist for an accepted indication. Patients should verify coverage in advance and request a cost estimate from the imaging center before scheduling whenever possible.

Is the loud noise harmful to my baby's hearing?

Studies measuring intrauterine sound pressure show that the maternal abdominal wall and amniotic fluid attenuate MRI gradient noise by approximately 30 decibels by the time it reaches the fetal ear. The resulting exposure is well below thresholds associated with hearing damage in animal and human studies. Audiologic follow-up of children scanned in utero has not demonstrated hearing impairment attributable to MRI. The mother wears ear protection because she experiences the full noise level outside the uterus.

Can I have a fetal MRI if I have claustrophobia?

Many patients with mild to moderate claustrophobia complete fetal MRI successfully with preparation and support. Strategies include touring the scanner room beforehand, using a wide-bore 70-centimeter magnet when available, having a partner present in the control room, listening to music through MR-compatible headphones, and using prism glasses to see outside the bore. Pharmacologic anxiolysis is generally avoided during pregnancy. Patients with severe claustrophobia who cannot tolerate the scan may need alternative imaging or repeat targeted ultrasound.