During SPEX inductively coupled plasma (ICP) analysis, a matrix-matched calibration is performed, yet results for a high-salinity water sample still show a 12% positive bias compared to certified reference values. Which mechanism is MOST likely responsible?
-
A
Space-charge effects from high dissolved solids suppressing analyte ion transmission in the mass spectrometer interface
-
B
Incomplete nebulization of the high-salinity sample causing signal depression rather than enhancement
-
C
Spectral overlap from sodium emission lines directly interfering with the analyte wavelength
-
D
Incorrect internal standard selection causing under-correction of signal drift