Instrumentation & Calibration Flashcards
7 cards from real PCR practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 7 Instrumentation & Calibration flashcards as text
Which method best verifies temperature uniformity across a thermal cycler block?
Answer: A multi-probe temperature verification system placed in multiple wells
Multi-probe systems measure several well positions at the same time, which shows differences across the block.
In thermal cycler specifications, 'ramp rate' refers to:
Answer: The speed of temperature change between steps, in °C per second
Ramp rate is how fast the block heats or cools between setpoints, in °C/s.
What is the main purpose of a heated lid on a thermal cycler?
Answer: To prevent condensation and evaporation of the reaction onto the tube cap
Keeping the lid hotter than the block (often around 105°C) stops water from condensing under the cap, so the reaction volume stays constant.
A lab notices that the outer wells of a 96-well block consistently give later Cq values. What is the most likely instrument-related cause?
Answer: Edge effect from temperature non-uniformity at the block perimeter
Perimeter wells can heat and cool differently from center wells, which lowers amplification efficiency in a position-dependent way.
In many qPCR systems, the passive reference dye ROX is used to:
Answer: Normalize for non-PCR fluorescence variation such as small differences in well volume
ROX fluorescence stays constant during PCR, so dividing the reporter signal by it corrects for optical and volume differences between wells.
What is the purpose of a pure dye (spectral) calibration on a multicolor qPCR instrument?
Answer: To record each dye's spectral signature so overlapping signals can be separated
Pure dye spectra let the software separate the combined fluorescence into individual dye contributions.
A background calibration on a qPCR instrument is performed using:
Answer: A plate containing buffer or water without fluorescent dyes
Background calibration measures fluorescence from the plate, buffer, and optics so it can be subtracted from experimental data.
How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. We have not yet matched this exam's facts to an official source, so confirm fees, dates and rules with the exam body. Not reviewed by a subject-matter expert. Editorial policy