Laboratory Procedures Flashcards
6 cards from real CCMA practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 6 Laboratory Procedures flashcards as text
A CCMA performs a fingerstick glucose on a patient and gets a critically low result of 32 mg/dL, but the patient is alert, asymptomatic, and just finished eating lunch. The CCMA notices the patient's fingers were cold and the site was squeezed firmly to obtain enough blood. What is the MOST likely explanation and appropriate next step?
Answer: Excessive squeezing diluted the sample with tissue fluid; warm the hand and repeat the test before notifying the provider
Milking or excessively squeezing a cold, poorly perfused finger introduces interstitial (tissue) fluid into the sample, which dilutes the blood and produces falsely low glucose readings. Because the clinical picture (alert, asymptomatic, recently ate) contradicts the critical value, the correct action is to address the preanalytical error — warm the site to improve perfusion, perform a proper puncture without milking, and repeat the test. A critical value should never be acted upon or reported when there is strong evidence of a collection error without first verifying it.
A provider orders a lactic acid level, a lavender-top CBC, a light-blue-top PT/INR, and blood cultures on a septic patient. The lactic acid must be collected in a gray-top tube. According to CLSI order of draw, in what sequence should these be collected?
Answer: Blood cultures, light blue, lavender, gray
The CLSI order of draw is: blood cultures (sterile) first, then light blue (sodium citrate), then serum tubes (red/gold), then green (heparin), then lavender (EDTA), and finally gray (sodium fluoride/potassium oxalate) last. Blood cultures always come first to preserve sterility, coagulation tubes precede additive tubes that could cause carryover interference, and the gray top is always drawn last because its additives (fluoride and oxalate) would significantly interfere with nearly every other test if carried over.
A CCMA is running daily quality control on a CLIA-waived hemoglobin analyzer. The QC results for the past six days have all been within range, but each result has been progressively higher than the last (a steady upward pattern approaching the upper limit). What does this pattern represent and what should the CCMA do?
Answer: A trend, which suggests a developing systematic problem; document it and notify the supervisor before it goes out of range
Six or more consecutive QC values moving steadily in one direction is called a trend, and it indicates a gradually developing systematic error — such as deteriorating reagents, a failing lamp or sensor, or environmental drift. Even though the individual values are still 'in range,' a trend predicts imminent failure and must be investigated proactively (documented, supervisor notified, instrument/reagents checked). It is not random error (which shows scattered values), and it is not a shift (which is an abrupt jump to a new level that then stays there). Past patient results are not automatically discarded because QC never actually failed.
A CCMA collects a lavender-top tube for a CBC, and the automated analyzer flags a critically low platelet count of 18,000/µL. The patient has no bruising, petechiae, or bleeding history. Before the provider is notified of a critical value, what preanalytical cause should be ruled out FIRST?
Answer: Platelet clumping or a clotted specimen due to inadequate mixing after collection
The most common cause of a spurious (falsely low) platelet count is platelet clumping or micro-clot formation, typically caused by inadequate inversion of the EDTA tube immediately after collection, a slow or traumatic draw, or EDTA-dependent platelet aggregation. Platelets consumed in clumps are not counted by the analyzer, producing pseudothrombocytopenia. Since the clinical picture (no bruising, petechiae, or bleeding) doesn't match the critical result, the specimen should be examined for clots and the count verified — often with a redraw — before the critical value is reported. Tourniquet hemoconcentration and IV contamination affect other analytes and would not selectively produce this pattern.
A patient arrives for a 24-hour urine creatinine clearance collection and tells the CCMA she began the collection yesterday at 7:00 AM by collecting her first morning void into the container, then collected all urine afterward, including her 7:00 AM void this morning. What is wrong with this collection?
Answer: The first void at the start of the 24-hour period should have been discarded, not collected; the specimen is invalid
Correct 24-hour urine protocol requires the patient to void at the start time and DISCARD that specimen — because that urine was produced before the timed period began — then collect every void afterward, including the final void at exactly 24 hours. By including the first morning void (urine produced overnight, before the collection window), this patient added urine from outside the timed period, falsely elevating the total volume and analyte content. The specimen is invalid and the collection must be restarted with proper instructions. Refrigeration is also required for creatinine clearance, but the fundamental timing error alone invalidates this collection.
While performing a venipuncture with a butterfly (winged infusion) set, a CCMA draws a light-blue-top sodium citrate tube as the FIRST and only coagulation tube. The PT result is normal but the aPTT is unexpectedly prolonged, and the lab notes the tube appears slightly underfilled. What is the most likely cause?
Answer: Air in the butterfly tubing displaced blood volume, causing an underfilled tube and excess citrate-to-blood ratio
Butterfly tubing holds approximately 0.5 mL of dead-space air. When a citrate tube is drawn first through a butterfly, that air is pulled into the tube in place of blood, underfilling it. Coagulation tubes require an exact 9:1 blood-to-anticoagulant ratio; underfilling leaves excess citrate, which binds additional calcium during testing and falsely prolongs results — the aPTT being especially sensitive. This is why a discard tube (plain or another citrate tube) must be drawn first when using a butterfly for coagulation studies. No prior tube was drawn, ruling out EDTA carryover, and warfarin primarily prolongs the PT/INR, not selectively the aPTT.
Written and edited by Nathan Beja, founder of Practice Test Geeks, with AI assistance. Exam facts are checked against the official sources linked below. Editorial policy