Infection Control and Sterilization Flashcards
6 cards from real CCMA practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 6 Infection Control and Sterilization flashcards as text
A medical assistant is running the office autoclave and notices the Bowie-Dick type test pack (used with a prevacuum sterilizer) shows a non-uniform color change in the center of the test sheet. What does this result most specifically indicate?
Answer: Residual air was not fully removed, so steam failed to penetrate the load
A Bowie-Dick test evaluates air removal and steam penetration in prevacuum (dynamic-air-removal) sterilizers, not the actual killing of microorganisms. A non-uniform or patchy color change—especially in the center—indicates an air-removal problem (an air leak or trapped air pocket) preventing steam from penetrating evenly. It is not a measure of temperature, pressure, bioburden, or ink integrity.
During processing, an instrument with a hinged joint (such as a hemostat) must be run through a steam sterilization cycle. To ensure sterilant contact on all surfaces, how should the instrument be positioned, and why?
Answer: Open or unlocked, so steam can reach the box-lock and serrated surfaces
Hinged instruments must be sterilized in the open or unlocked position. If locked closed, the box-lock and mating surfaces are pressed together and steam (the sterilant) cannot contact those hidden surfaces, leaving them unsterilized. Locking closed, over-compressing, or leaving liquid in the pack all impair sterilant contact.
A biological indicator (spore test) is run weekly on the office autoclave. This week the processed BI vial shows growth (a positive result) after incubation, while the unprocessed control vial also shows growth. What is the correct interpretation and action?
Answer: The result is valid and indicates a sterilizer failure requiring recall and reprocessing of loads
The control vial is unprocessed and is SUPPOSED to grow—it verifies the spores were viable and the incubation/media worked. Because the control grew, the test is valid. A processed vial that also grows means the spores survived the cycle, so the sterilizer failed. The correct action is to remove the sterilizer from service and recall/reprocess loads back to the last negative BI. (A control that did NOT grow would be what invalidates the test.)
A patient with confirmed Clostridioides difficile infection has soiled an exam table. When selecting a disinfectant and technique for the surface, which choice reflects the key distinction the MA must make?
Answer: An EPA-registered sporicidal agent (e.g., diluted bleach), because C. difficile forms spores
C. difficile is a spore-forming organism, and its spores resist alcohols and standard quaternary ammonium compounds. Environmental surfaces contaminated by C. difficile require an EPA-registered sporicidal agent (commonly a diluted sodium hypochlorite/bleach solution). Alcohol is not sporicidal, and treating the surface as ordinary noncritical cleaning ignores the spore hazard.
An MA receives a flexible endoscope after a procedure and cannot immediately reprocess it. According to reprocessing principles, why is the pre-cleaning/point-of-use treatment step so time-critical, and what happens if it is delayed?
Answer: Delay allows organic material and biofilm to dry and adhere, blocking later high-level disinfection
Point-of-use pre-cleaning matters because dried organic soil and biofilm strongly adhere to lumens and surfaces. Once dried, this material shields microorganisms and physically blocks the disinfectant from contacting them, making subsequent high-level disinfection unreliable—no terminal step fully compensates for a scope that wasn't pre-cleaned promptly. The issue is bioburden drying/biofilm, not coatings, evaporating disinfectant, or an irrelevant delay.
An instrument is classified under the Spaulding system. It is a laryngoscope blade that contacts intact mucous membranes but does not penetrate sterile tissue or the vascular system. What is the MINIMUM level of reprocessing required, and why?
Answer: High-level disinfection, because it is a semicritical item contacting mucous membranes
Under the Spaulding classification, items that contact mucous membranes or non-intact skin but do not enter sterile tissue are semicritical and require, at minimum, high-level disinfection. Sterilization exceeds the minimum requirement; intermediate- and low-level disinfection are insufficient for a semicritical device because they may not eliminate all vegetative bacteria, viruses, and mycobacteria on a mucosa-contacting instrument.