EKG Administration Flashcards
6 cards from real CCMA practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 6 EKG Administration flashcards as text
During a routine 12-lead EKG, the tracing shows tall R waves in V1 and V2 with deep S waves in V5 and V6, an unusual pattern for the patient. Before alerting the provider to possible right ventricular hypertrophy, what should the medical assistant check first?
Answer: Whether the precordial leads were reversed or misplaced across the chest
Reversed or misplaced precordial (chest) leads are a classic cause of unexpected R-wave progression abnormalities, such as prominent R waves in V1–V2. Because lead misplacement can mimic pathology like right ventricular hypertrophy or posterior infarction, the medical assistant should verify correct anatomical placement of V1–V6 before the tracing is interpreted. Breathing rate and limb-lead position over bone affect artifact, not precordial R-wave patterns, and a gain error would enlarge all complexes equally rather than reverse R-wave progression.
A patient with dextrocardia requires a 12-lead EKG. In addition to placing the chest leads in mirror-image positions on the right side, what else must the medical assistant do to obtain an accurate tracing?
Answer: Reverse the right and left arm limb electrodes
In dextrocardia, the heart's electrical axis is mirrored. A true mirror-image EKG requires both right-sided precordial lead placement and reversal of the right and left arm electrodes so the frontal-plane leads reflect the reversed cardiac orientation. Changing paper speed or moving limb leads to the torso does not correct for the mirrored anatomy, and reversing the leg electrodes has essentially no effect since the left leg reference dominates and the right leg is the ground.
While performing an EKG on a patient with a continuous tremor from Parkinson's disease, the medical assistant notes persistent somatic artifact despite reassurance and repositioning. Which technique is MOST appropriate to minimize the artifact?
Answer: Move the limb electrodes proximally to the trunk-adjacent portions of the limbs
For patients with uncontrollable tremors, placing the limb electrodes higher on the limbs, closer to the trunk (e.g., upper arms and thighs), reduces the muscle-movement artifact transmitted to the electrodes while preserving an acceptable tracing. Having the patient grip an object actually increases muscle tension and worsens somatic artifact. Adjusting filters or gain distorts the tracing rather than addressing the source, and rerouting wires addresses AC interference (60-cycle), not somatic tremor artifact.
A provider orders a right-sided EKG for a patient with an acute inferior MI to evaluate for right ventricular involvement. Which lead is MOST diagnostic for right ventricular infarction, and where is it placed?
Answer: V4R — fifth intercostal space, right midclavicular line
V4R, placed at the fifth intercostal space on the right midclavicular line (the mirror image of standard V4), is the single most sensitive and specific lead for detecting right ventricular infarction; ST elevation ≥1 mm in V4R strongly suggests RV involvement. The distractors describe incorrect landmarks: 'V1R' would simply be the standard V2 position, V6R belongs at the right midaxillary (not anterior axillary) line, and the V3R description misstates its placement, which is midway between V1 and V4R.
During a treadmill stress test, the medical assistant observes the patient becoming pale and ataxic, and the monitor shows 2.5 mm of horizontal ST-segment depression in multiple leads. The patient denies chest pain. What is the medical assistant's BEST immediate action?
Answer: Alert the supervising provider immediately so the test can be stopped per protocol
Significant ST depression (≥2 mm horizontal or downsloping), ataxia, and pallor are absolute or near-absolute indications for test termination even without chest pain, and the supervising provider must be notified immediately — the MA does not independently continue or modify the protocol. Continuing or merely reducing speed ignores clear termination criteria, and an abrupt full stop without a cool-down walk (unless the patient is unstable enough to require it per provider direction) can cause venous pooling and syncope; the provider directs how the test is terminated.
A medical assistant applies a Holter monitor and instructs the patient on the diary. Two days later the recording returns with large segments of flat-line signal in one channel, though the patient reports wearing the device continuously. What is the MOST likely cause?
Answer: An electrode detached or lost skin contact due to inadequate skin preparation
A flat-line (absent) signal in a single channel of a Holter recording almost always indicates a loose or detached electrode on that channel, typically from poor skin preparation (failure to cleanse, dry-shave, or abrade the site) or inadequate adhesion during daily activities. Memory or battery failures would affect the entire recording across all channels, not one channel intermittently, and true sinus arrest of that duration would be a life-threatening event visible across all channels — not an isolated single-channel dropout.