Treatment Planning Flashcards
7 cards from real RT practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 7 Treatment Planning flashcards as text
For a fixed-SSD (SSD) treatment setup, which depth dose quantity is typically used in the monitor unit calculation?
Answer: Percent depth dose (PDD)
PDD is defined at a fixed SSD, so it is the natural quantity for SSD-based monitor unit calculations.
A beam is calibrated at 100 cm. Approximately what fraction of the dose rate is delivered at 110 cm, based on the inverse square law?
Answer: 0.826
(100/110)^2 = 0.826, so the dose rate drops to about 82.6%.
What is the equivalent square of a 10 cm x 20 cm rectangular field using the area/perimeter (4A/P) method?
Answer: 13.3 cm
4A/P = 4(200)/60 = 13.3 cm.
Which dosimetric quantity is most commonly used for monitor unit calculations in an isocentric (SAD) setup?
Answer: Tissue-maximum ratio (TMR)
TMR does not depend on SSD, so it suits SAD setups where the point of calculation is at isocenter.
The Mayneord F factor is used to:
Answer: Convert PDD from one SSD to another SSD
The Mayneord F factor approximately corrects PDD for a change in SSD by applying inverse-square terms.
For two beams with a hinge angle of 90 degrees, what is the ideal wedge angle?
Answer: 45 degrees
Wedge angle = 90 - (hinge angle / 2) = 90 - 45 = 45 degrees.
When matching two adjacent fields on the skin, the gap is calculated to make the fields meet:
Answer: At a chosen depth inside the patient
Skin gaps let divergent beam edges meet at a chosen depth, which avoids overlapping hot spots over critical structures.
How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. Exam facts are checked against the official sources listed below. Not reviewed by a subject-matter expert. Editorial policy
Sources & references
Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy
- arrt.org — Score validity, retake policy, test delivery, passing score, score release, fees, exam information · verified Oct 2026
- assets-us-01.kc-usercontent.com — Eligibility, exam format, time limit, exam information · verified Oct 2026