Which leak test method is most commonly used to verify the integrity of a body plethysmograph cabinet?
Answer: Sealing the cabinet and monitoring pressure drift over time
A timed pressure-decay (leak) test with the cabinet sealed detects leaks that would disrupt the constant-volume assumption required for accurate plethysmographic measurements.
What is the role of a HEPA filter placed on the exhalation port of a spirometer during patient testing?
Answer: Protects the flow sensor from contamination and reduces cross-infection risk
HEPA or bacterial/viral filters protect the spirometer internals from patient secretions and prevent cross-contamination between patients.
A spirometer's volume reading drifts upward by 4.2% over two weeks of biologic controls. What should the technologist do?
Answer: Take the spirometer out of service and recalibrate or service it
A consistent upward drift exceeding the ±3.5% ATS accuracy limit indicates equipment malfunction requiring removal from service and correction.
What information should be documented in the PFT lab's equipment logbook after each daily calibration?
Answer: Date, time, technologist ID, calibration values, pass/fail status, and any corrective actions
Comprehensive calibration records including operator, values, and any corrective actions support quality assurance audits and regulatory compliance.
Which parameter is checked during a resistance test of a pneumotachograph flow sensor?
Answer: Pressure drop across the sensor at a known flow rate
Pneumotachograph sensors measure flow by detecting the pressure differential across a resistance mesh; verifying pressure drop at a known flow confirms proper sensor function.
When must a spirometer's calibration be repeated within the same testing day?
Answer: After any significant temperature change in the testing room
Significant ambient temperature changes alter gas viscosity and sensor characteristics, requiring recalibration to maintain BTPS accuracy.
How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. Exam facts are checked against the official sources listed below. Not reviewed by a subject-matter expert. Editorial policy
Sources & references
Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy
- nbrc.org — Renewal, score validity, fees, exam format, passing score, registration deadline, retake fee, retake policy, test dates, time limit · verified Oct 2026
- test-takers.psiexams.com — Test delivery · verified Oct 2026