What environmental data must be entered into a spirometry system before testing to allow BTPS correction?
Answer: Ambient temperature, barometric pressure, and relative humidity
BTPS (Body Temperature, Pressure, Saturated) correction requires ambient temperature, barometric pressure, and humidity to convert measured gas volumes to body conditions.
A peak flow meter shows readings 15% below the lab spirometer for the same patient. What is the most appropriate first action?
Answer: Recalibrate or replace the peak flow meter
A consistent 15% discrepancy suggests instrument error; the peak flow meter should be checked against a reference or replaced before relying on its readings.
Which gas mixture is used to calibrate the DLCO analyzer?
Answer: A precision gas with known concentrations of CO and an inert tracer gas
DLCO analyzers require a certified precision calibration gas containing specific concentrations of CO and an inert tracer (e.g., methane or helium) for accurate two-point calibration.
What is the primary purpose of the mechanical pump ('super syringe') used in DLCO system calibration?
Answer: To verify the accuracy of inspired volume measurement at known flow rates
The super syringe delivers known volumes at controlled flow rates to verify that the spirometric component of the DLCO system measures inspired volume accurately.
After replacing a flow sensor, what is the minimum verification a RPFT should perform before resuming patient testing?
Answer: Full calibration including volume accuracy and linearity checks
Any hardware change requires a complete recalibration to ensure the new sensor performs within acceptable accuracy limits across its entire range.
What is the recommended ambient temperature range for performing pulmonary function tests?
Answer: 17–40°C (63–104°F)
ATS/ERS guidelines specify that spirometry should be performed when ambient temperature is between 17°C and 40°C to ensure reliable BTPS correction and sensor function.
How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. Exam facts are checked against the official sources listed below. Not reviewed by a subject-matter expert. Editorial policy
Sources & references
Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy
- nbrc.org — Renewal, score validity, fees, exam format, passing score, registration deadline, retake fee, retake policy, test dates, time limit · verified Oct 2026
- test-takers.psiexams.com — Test delivery · verified Oct 2026