Pipeline Installation Flashcards
7 cards from real RMGA practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 7 Pipeline Installation flashcards as text
According to NFPA 99, what is the minimum burial depth required for medical gas piping installed underground inside a healthcare facility?
Answer: 18 inches
NFPA 99 requires underground medical gas piping inside buildings to be buried at least 18 inches deep to protect against damage.
Which type of fitting is prohibited in medical gas pipeline systems according to NFPA 99?
Answer: Threaded fittings on copper tubing
Threaded fittings on copper tubing are prohibited in medical gas systems; all joints must use brazed or approved mechanical connections.
What is the required color coding for a nitrous oxide pipeline in the United States?
Answer: Blue
Nitrous oxide pipelines are coded blue per CGA and NFPA 99 standards for medical gas identification in the US.
When installing copper tubing for medical gas pipelines, which grade must be used?
Answer: Type ACR copper cleaned for oxygen service
Medical gas piping requires Type K or Type L hard-drawn copper tubing that has been cleaned for oxygen service and labeled accordingly.
In a medical gas pipeline system, what is the purpose of a zone valve box?
Answer: To isolate a section of the distribution system for maintenance or emergency
Zone valve boxes allow specific areas to be isolated from the pipeline system without shutting down the entire facility supply.
After brazing joints in a medical gas pipeline, what must be done before pressure testing?
Answer: Blown down with oil-free nitrogen to remove residual flux
After brazing, lines must be blown down with dry nitrogen to remove flux residue, which is corrosive and could contaminate the gas supply.
What is the maximum allowable pressure drop across a medical gas pipeline system during peak demand as specified by NFPA 99?
Answer: 5 psi
NFPA 99 requires that the pipeline system maintain pressures within ±5 psi (or ±10%) of the operating pressure under peak demand conditions.
How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. We have not yet matched this exam's facts to an official source, so confirm fees, dates and rules with the exam body. Not reviewed by a subject-matter expert. Editorial policy