Cardiac Implantable Electronic Devices (CIEDs) and MRI Safety Flashcards
7 cards from real MRSO practice questions. Tap to flip, then mark Knew It or Still Learning โ missed cards come back until you master them.
Read the first 7 Cardiac Implantable Electronic Devices (CIEDs) and MRI Safety flashcards as text
Which ASTM/ISO designation applies to a pacemaker that may be used in MRI only when all specified conditions of field strength, SAR, and gradient slew rate are met?
Answer: MR Conditional
MR Conditional indicates the device poses no known hazards only within the specific MRI conditions defined in its labeling, unlike MR Safe, which implies safety in all MRI environments.
Which mechanism is the primary source of cardiac injury risk from MRI in patients with implanted pacemaker leads?
Answer: RF-induced resistive heating concentrated at the electrode-tissue interface
RF energy is absorbed along the conductor and concentrates at the lead tip, producing resistive heating that can thermally damage myocardial tissue at the electrode-tissue interface.
What was the central finding of the MagnaSafe Registry regarding MRI performed in patients with non-MR-conditional CIEDs?
Answer: Clinically significant adverse events were rare when a standardized monitoring protocol was followed
The MagnaSafe Registry demonstrated that MRI in patients with legacy (non-MR-conditional) CIEDs had a very low rate of device-related adverse events when performed under a rigorous, standardized protocol with continuous monitoring.
Which whole-body average SAR limit is most commonly specified in MR-conditional pacemaker labeling to reduce the risk of RF-induced lead tip heating?
Answer: 2.0 W/kg
Most MR-conditional pacemaker labels specify a whole-body average SAR limit of 2.0 W/kg, which aligns with the IEC normal operating mode threshold and limits RF deposition in implanted leads.
Why are abandoned (non-functional) pacemaker leads considered particularly hazardous during MRI compared to connected leads?
Answer: Without a connected generator, RF-induced current concentrates entirely at the lead tip, maximizing heating
Abandoned leads without a proximal connection act as unloaded antennas; with no circuit path to dissipate current, all RF-induced energy concentrates at the electrode tip, producing greater heating than a connected lead.
What occurs in most conventional pacemakers upon entering the static magnetic field of an MRI scanner?
Answer: The reed switch closes, switching the device to asynchronous (fixed-rate) pacing mode
The static magnetic field closes the internal reed switch in most conventional pacemakers, activating asynchronous pacing mode โ the same response produced by placing an external magnet over the device.
Which CIED type carries the greatest overall MRI risk due to its high-voltage capacitors and defibrillation coil design?
Answer: Implantable cardioverter-defibrillator (ICD) with defibrillation coil
ICDs contain high-voltage capacitors and defibrillation coils that act as larger RF-absorbing antennas, increasing lead heating risk and creating potential for inappropriate high-energy shock delivery during MRI.