Cardiac Implantable Electronic Devices (CIEDs) and MRI Safety Flashcards
7 cards from real MRSO practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 7 Cardiac Implantable Electronic Devices (CIEDs) and MRI Safety flashcards as text
Prior to MRI, MR-conditional pacemakers are typically reprogrammed to which pacing mode to prevent inappropriate pacing inhibition?
Answer: Asynchronous (AOO/VOO/DOO) pacing mode
Asynchronous mode causes the pacemaker to pace at a fixed rate regardless of sensed signals, preventing the device from misinterpreting MRI-generated electromagnetic noise as intrinsic cardiac activity and inhibiting pacing.
What minimum time after CIED implantation do most MR-conditional device protocols require before performing MRI?
Answer: 4–6 weeks to allow fibrous lead fixation at the electrode-tissue interface
Newly implanted leads require 4–6 weeks for fibrous tissue ingrowth to stabilize the electrode-tissue interface; scanning before this increases the risk of lead dislodgement and amplifies heating injury at the unhealed tip.
What is the mandatory action required AFTER MRI in a patient with an MR-conditional pacemaker?
Answer: Device interrogation to verify parameters and restore original programmed settings
Post-scan interrogation confirms that pacing thresholds, sensing amplitudes, lead impedances, and stored therapies are unchanged and restores any programming alterations made before the scan.
Why is continuous ECG and pulse oximetry monitoring required throughout MRI in patients with CIEDs?
Answer: To detect pacing inhibition, inappropriate device therapy, or arrhythmias in real time during the examination
Continuous cardiac monitoring allows immediate detection of adverse device–MRI interactions such as pacing inhibition, inappropriate ICD shocks, or hemodynamically significant arrhythmias that require immediate intervention.
Why do epicardial pacemaker leads present a unique MRI heating challenge compared to standard transvenous endocardial leads?
Answer: Their extravascular routing creates larger, less predictable conductor loops that couple more RF energy
Epicardial leads traverse extravascular paths and form larger current loops of variable geometry, increasing RF coupling and concentrating greater heating at the electrode tip compared to standard intravascular leads.
Most MR-conditional pacemaker labels impose the strictest SAR restrictions or require additional precautions when imaging which body region?
Answer: Thorax/chest, where device and lead placement places maximum RF deposition over the hardware
Thoracic MRI positions the RF transmit coil directly over the pulse generator and leads, dramatically increasing local RF deposition and heating risk compared to imaging remote body regions.
For MRI of a pacemaker-dependent patient with an MR-conditional device, which professional must be immediately available throughout the examination?
Answer: An electrophysiologist or cardiologist capable of managing acute cardiac emergencies
Pacemaker-dependent patients risk life-threatening bradycardia or asystole if pacing is inhibited; an electrophysiologist or cardiologist must be immediately available to manage such emergencies and reprogram the device if needed.