  1. During a magnet quench, what is the main danger to people in the scan room if the quench pipe fails?

    Answer: Asphyxiation and cryogenic burns from helium gas

    Boiling helium can displace oxygen and cause frostbite, so the room should be evacuated at once.

  2. Rapidly switching gradient fields (dB/dt) most directly cause which effect?

    Answer: Peripheral nerve stimulation

    Fast-changing gradient fields induce currents in tissue that can stimulate peripheral nerves and cause tingling or twitching.

  3. Which patient has the highest risk of nephrogenic systemic fibrosis after receiving gadolinium?

    Answer: eGFR below 30 mL/min/1.73 m² or on dialysis

    NSF risk is mainly in severe chronic kidney disease or acute kidney injury, especially with higher-risk (group I) agents.

  4. What is the FDA guideline limit for peak unweighted sound pressure level in MRI?

    Answer: 140 dB

    FDA guidance sets the peak unweighted sound pressure limit at 140 dB, and hearing protection is required for all patients.

  5. A patient has a medication patch with a metallic foil backing. What is the main safety concern?

    Answer: RF heating that can burn the skin

    Metallic backing can heat during RF transmission, so these patches are usually removed before scanning with the prescriber's approval.

  6. Why does an ECG tracing inside the magnet often show an elevated T wave?

    Answer: The magnetohydrodynamic effect of blood flowing in the static field

    Ionic blood moving through the static field creates a voltage that adds to the ECG, mostly during the T wave.

  7. An MR Conditional implant's labeling specifies a B1+rms limit. What does this parameter describe?

    Answer: The time-averaged RF transmit field magnitude

    B1+rms is the root-mean-square RF transmit field, a more accurate measure of RF exposure for implants than SAR.

How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. Exam facts are checked against the official sources listed below. Not reviewed by a subject-matter expert. Editorial policy

Sources & references

Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy

  1. arrt.org — Passing score, score release, fees, exam information · verified Oct 2026
  2. assets-us-01.kc-usercontent.com — Renewal, score validity, test delivery, eligibility, exam format, test dates, time limit · verified Oct 2026