ARRT MRI Exam — Questions and Answers
Question 1: A technologist wants to reduce scan time while maintaining similar SNR. Which strategy is most effective?
- Increase the repetition time (TR)
- Increase slice thickness
- Decrease the matrix size
- Use parallel imaging (SENSE/GRAPPA) (Correct answer)
Correct answer: Use parallel imaging (SENSE/GRAPPA)
Parallel imaging techniques use multiple coil elements to under-sample k-space, reducing scan time while maintaining acceptable SNR through reconstruction algorithms.
Question 2: Which MRI finding is characteristic of a benign rotator cuff tear involving a full-thickness supraspinatus tendon rupture?
- Fluid signal gap (T2 bright) extending through the full thickness of the tendon (Correct answer)
- T1 bright signal in the tendon
- Low T2 signal with thickening of the tendon
- Low signal on all sequences through the tendon
Correct answer: Fluid signal gap (T2 bright) extending through the full thickness of the tendon
A full-thickness tear shows fluid (T2 hyperintense) extending completely through the tendon from bursal to articular surface.
Question 3: SSFP (steady-state free precession) sequences such as TrueFISP/FIESTA produce contrast that is proportional to which tissue property ratio?
- TE/T2*
- T2/T1 (Correct answer)
- T1/T2
- TR/T1
Correct answer: T2/T1
SSFP signal is proportional to T2/T1 ratio; tissues with high T2/T1 (like blood and fluids) appear very bright, making SSFP ideal for cardiac cine and vascular imaging.
Question 4: Wraparound (aliasing) artifact in MRI is caused by:
- T2* dephasing from metallic objects
- Phase errors from pulsatile flow
- Structures outside the FOV contributing signal that is folded back into the image (Correct answer)
- Gradient non-linearity at the edges of the bore
Correct answer: Structures outside the FOV contributing signal that is folded back into the image
When anatomy extends beyond the selected FOV in the phase-encode direction, it is aliased and appears superimposed on the opposite side of the image.
Question 5: A technologist who holds an RT(MR)(ARRT) credential and allows it to lapse must complete which process to reinstate it?
- Retake the full MRI registry examination
- Resubmit all original application materials and retake the exam
- Complete only a practical competency evaluation
- Apply for reinstatement and fulfill CE requirements (Correct answer)
Correct answer: Apply for reinstatement and fulfill CE requirements
ARRT allows lapsed credential holders to reinstate by applying for reinstatement and meeting outstanding CE obligations.
Question 6: Which coil type generally provides the highest SNR for imaging a small superficial structure?
- Volume coil
- Body coil
- Surface coil (Correct answer)
- Quadrature head coil
Correct answer: Surface coil
Surface coils placed close to the target anatomy have superior SNR for small or superficial structures because they detect signal from a limited volume with less noise.
Question 7: A metalworker reports possibly getting metal in his eye years ago but has no documentation. What is the most appropriate step before MRI?
- Proceed, since old fragments are always encapsulated
- Scan with a lower SAR setting
- Ask him to keep his eyes closed during the scan
- Get orbital radiographs to screen for a foreign body (Correct answer)
Correct answer: Get orbital radiographs to screen for a foreign body
Orbit radiographs are the standard way to screen for intraocular metal, which could move and cause blindness.
Question 8: Which of the following best explains why a phased-array coil improves SNR compared to a body coil?
- They apply stronger RF pulses to excite more protons
- Phased-array coils use higher magnetic field strength
- Multiple smaller coil elements are positioned close to the anatomy, maximizing signal pickup while limiting noise volume (Correct answer)
- They increase the repetition time automatically
Correct answer: Multiple smaller coil elements are positioned close to the anatomy, maximizing signal pickup while limiting noise volume
Each element of a phased-array coil is small and close to the anatomy, providing high sensitivity to local signal with limited noise, and their combined data significantly improves overall SNR.
Question 9: What is the effect of increasing receiver bandwidth on MRI image quality?
- Image contrast improves significantly
- SNR decreases but chemical shift artifact is reduced (Correct answer)
- SNR increases and scan time decreases
- Spatial resolution decreases
Correct answer: SNR decreases but chemical shift artifact is reduced
Increasing bandwidth reduces chemical shift artifact and allows shorter TE/TR, but more noise is accepted per unit time, lowering SNR.
Question 10: What is the primary advantage of using a higher-relaxivity GBCA (e.g., gadobenate dimeglumine) compared to a standard-relaxivity agent?
- It can be used at a lower dose to achieve comparable enhancement (Correct answer)
- It crosses the blood-brain barrier more readily
- It eliminates the need for dynamic imaging
- It is completely renally safe at any dose
Correct answer: It can be used at a lower dose to achieve comparable enhancement
Higher relaxivity means greater T1 shortening per unit concentration, so a lower molar dose can produce equivalent or superior enhancement compared to standard agents.
Question 11: Pulsatile flow artifact from a vessel typically appears as ghost images displaced in which direction?
- Phase-encoding direction (Correct answer)
- Slice-selection direction
- Frequency-encoding direction
- Both frequency and phase directions equally
Correct answer: Phase-encoding direction
Pulsatile flow ghosts appear in the phase-encoding direction because pulsatile variations in spin phase occur between TR intervals during phase-encode data collection.
Question 12: Iron oxide nanoparticles (e.g., USPIO) used as MRI contrast agents primarily shorten which relaxation time?
- T2 only with no T2* effect
- Neither T1 nor T2
- T2 and T2* (susceptibility effect) (Correct answer)
- T1 only
Correct answer: T2 and T2* (susceptibility effect)
Iron oxide particles create local magnetic field inhomogeneities that predominantly shorten T2 and T2*, causing signal loss (darkening) on T2-weighted and gradient echo images.
Question 13: What is contrast-to-noise ratio (CNR) in MRI and why is it important?
- The difference in SNR between two adjacent tissues; it determines the ability to distinguish lesions from normal tissue (Correct answer)
- The ratio of TR to TE for optimal contrast
- The number of phase encodes used to improve tissue differentiation
- The ratio of gadolinium dose to body weight
Correct answer: The difference in SNR between two adjacent tissues; it determines the ability to distinguish lesions from normal tissue
CNR is the SNR difference between adjacent tissues; high CNR is essential for detecting lesions because it reflects how well two tissues can be discriminated from each other.
Question 14: Why does an ECG tracing inside the magnet often show an elevated T wave?
- The magnetohydrodynamic effect of blood flowing in the static field (Correct answer)
- RF heating of the myocardium
- Gradient-induced arrhythmia
- Electrode polarization from contrast
Correct answer: The magnetohydrodynamic effect of blood flowing in the static field
Ionic blood moving through the static field creates a voltage that adds to the ECG, mostly during the T wave.
Question 15: This magnet has a strength of up to 0.3 T, can be switched off, and can be used in both horizontal and vertical field systems.
- Permanent magnet
- Selenoid magnet
- Superconducting
- Resistive magnet (Correct answer)
Correct answer: Resistive magnet
Resistive magnets consist of coils of wire through which a large electric current is continuously passed to generate a magnetic field. They can be switched off, offer field strengths typically up to 0.3 Tesla, and can be designed for both horizontal and vertical field orientations. However, they require significant continuous power and generate substantial heat, limiting their use in high-field MRI.
Question 16: A double inversion recovery preparation is commonly used in cardiac MRI to:
- Produce black-blood images by nulling the signal from flowing blood (Correct answer)
- Null fat and water at the same time to show the coronary arteries
- Measure myocardial T2* for iron overload
- Increase signal from gadolinium in the blood pool
Correct answer: Produce black-blood images by nulling the signal from flowing blood
A non-selective inversion followed by slice-selective re-inversion nulls blood that flows into the slice, giving black-blood images of the myocardium.
Question 17: Magnets that use liquid helium or cryogen to cool the wires after applying direct current to a coil of wire.
- Permanent magnet
- Selenoid magnet
- Superconducting (Correct answer)
- Resistive magnet
Correct answer: Superconducting
Superconducting magnets utilize special alloys that, when cooled to extremely low temperatures (near absolute zero) using cryogens like liquid helium, lose all electrical resistance. This allows a large current to flow indefinitely without power, creating very strong and stable magnetic fields. These magnets are essential for high-field MRI systems, typically operating at 1.5 Tesla and above.
Question 18: Magnetization of hydrogen using MRI
- In the primary magnetic field path
- perpendicular to the primary magnetic field (Correct answer)
Correct answer: perpendicular to the primary magnetic field
In MRI, the detectable signal is generated from transverse magnetization, which is created when an RF pulse tips the net magnetization of hydrogen protons away from the primary magnetic field (B0). This transverse magnetization precesses perpendicular to B0, inducing a current in the receiver coil. Therefore, the magnetization used for signal detection is perpendicular to the primary magnetic field.
Question 19: When performing MRI on a patient with an insulin pump, which step is essential before beginning the exam?
- Leave the pump in place since insulin is not ferromagnetic
- Increase the insulin dose before scanning
- Remove the pump (if removable) or verify the pump is MR-conditional with the manufacturer before scanning (Correct answer)
- Tape the pump to the patient's leg during the scan
Correct answer: Remove the pump (if removable) or verify the pump is MR-conditional with the manufacturer before scanning
Insulin pumps may malfunction, heat, or be damaged in the MRI environment; removal or manufacturer verification of MR-conditional status is mandatory.
Question 20: Require the employment of solenoid coils because of the B0's orientation; these coils are more effective than linear coils and can be electrically combined with other coils to enhance signal homogeneity across an area of interest.
- Phased array
- Quadrature
- Helmholtz pair
- Vertical (Correct answer)
Correct answer: Vertical
Vertical coils are specifically designed for MRI systems where the main magnetic field (B0) is oriented vertically, often found in open or permanent magnet systems. These coils are more effective than linear coils in this orientation and can be electrically combined with other coils to enhance signal homogeneity across the area of interest, optimizing signal reception for vertical field systems.
Question 21: For MRI of the shoulder, the patient should be positioned with the arm in which orientation to best center the glenohumeral joint?
- Arm crossed over the chest
- Arm in full internal rotation only
- Arm elevated above the head
- Arm at the side in neutral or slight external rotation, shoulder centered in the coil (Correct answer)
Correct answer: Arm at the side in neutral or slight external rotation, shoulder centered in the coil
Positioning the shoulder with the arm at the side in neutral or slight external rotation and centering the glenohumeral joint within the dedicated shoulder coil optimizes image quality.
Question 22: MR spectroscopy (MRS) of the brain identifies metabolites by their distinct chemical shifts. Which metabolite is elevated in tumors with high cell density and membrane turnover?
- Lactate
- N-acetylaspartate (NAA)
- Creatine (Cr)
- Choline (Cho) (Correct answer)
Correct answer: Choline (Cho)
Choline is a marker of membrane synthesis and cellular turnover; elevated choline is characteristic of tumors, demyelination, and other high-proliferation states.
Question 23: Under ARRT policy, which scenario would trigger an automatic ethics review before an MRI registry application is processed?
- Candidate has exceeded the CE deadline by one week
- Candidate previously worked as a radiography aide
- Candidate holds a second radiography credential
- Candidate has a prior misdemeanor conviction on record (Correct answer)
Correct answer: Candidate has a prior misdemeanor conviction on record
ARRT requires candidates with prior criminal history to undergo an ethics review to determine eligibility before the application proceeds.
Question 24: Which technique is most effective for reducing susceptibility artifact from a metallic implant in MRI?
- Increase the slice thickness
- Use a spin-echo or fast spin-echo (FSE) sequence with narrow bandwidth (Correct answer)
- Use a gradient-echo sequence
- Increase TE
Correct answer: Use a spin-echo or fast spin-echo (FSE) sequence with narrow bandwidth
Spin-echo and FSE sequences use 180° refocusing pulses that compensate for susceptibility-induced dephasing, and a narrow bandwidth reduces chemical shift but a wide bandwidth reduces susceptibility — MARS protocols use high bandwidth FSE.
Question 25: Which vital sign parameter is most critical to monitor during MRI in a sedated pediatric patient?
- Temperature with a mercury thermometer
- Oxygen saturation via MR-compatible pulse oximetry (Correct answer)
- Blood pressure only
- Heart rate via standard ECG only
Correct answer: Oxygen saturation via MR-compatible pulse oximetry
Continuous oxygen saturation monitoring via MR-compatible pulse oximetry is essential in sedated pediatric patients to detect respiratory compromise caused by sedation.
Question 26: Which of the following is a mandatory competency category that must be documented before a student can sit for the ARRT MRI registry?
- Cardiac MRI stress testing
- Head MRI without contrast (Correct answer)
- MRI-guided biopsy procedures
- MRI-compatible pacemaker implantation
Correct answer: Head MRI without contrast
Head MRI without contrast is among the required clinical competencies students must document as part of ARRT's clinical requirements.
Question 27: Which of the following is the most effective method to reduce motion artifact from bowel peristalsis in abdominal MRI?
- Increase TR
- Administer an antiperistaltic agent such as glucagon (Correct answer)
- Use a larger FOV
- Apply saturation bands inferior to the FOV
Correct answer: Administer an antiperistaltic agent such as glucagon
IV or IM glucagon inhibits smooth muscle contraction (antiperistaltic effect), reducing bowel motion artifact during abdominal MRI acquisitions.
Question 28: The range of frequencies sampled during the process is represented by the receiver bandwidth.
- gradient coils
- phase difference
- gradient of frequency encoding (Correct answer)
- gradient for slice selection
Correct answer: gradient of frequency encoding
The receiver bandwidth in MRI represents the range of frequencies sampled during the signal acquisition process. This range of frequencies is directly linked to the frequency encoding gradient, which spatially encodes information along one dimension of the image. A wider bandwidth allows for faster acquisition but can lead to a reduction in signal-to-noise ratio (SNR).
Question 29: Which MRI artifact appears as a ghosting or smearing of bright structures in the phase-encoding direction?
- Motion (ghosting) artifact (Correct answer)
- Susceptibility artifact
- Gibbs (truncation) artifact
- Chemical shift artifact
Correct answer: Motion (ghosting) artifact
Motion artifact manifests as ghost images displaced in the phase-encoding direction because phase-encoding samples data over many TR intervals while frequency encoding is instantaneous.
Question 30: Cardiac MRI uses which technique to synchronize data acquisition with the cardiac cycle and minimize cardiac motion artifact?
- ECG gating or cardiac triggering (Correct answer)
- Real-time imaging without synchronization
- Breath-hold sequences only
- Respiratory gating only
Correct answer: ECG gating or cardiac triggering
ECG gating (prospective triggering or retrospective gating) synchronizes k-space acquisition to specific phases of the cardiac cycle, minimizing motion artifact from myocardial contraction.
Question 31: Which coil type provides the highest signal-to-noise ratio (SNR) for imaging small superficial structures such as the wrist?
- Head coil
- Body coil
- Phased-array spine coil
- Dedicated extremity or surface coil (Correct answer)
Correct answer: Dedicated extremity or surface coil
Dedicated extremity or surface coils placed close to the anatomy maximize SNR by minimizing the receive volume to the region of interest.
Question 32: In a patient with an acute ischemic stroke, the infarcted region typically appears:
- Dark on DWI and bright on the ADC map
- Dark on both DWI and the ADC map
- Bright on DWI and dark on the ADC map (Correct answer)
- Bright on both DWI and the ADC map
Correct answer: Bright on DWI and dark on the ADC map
Cytotoxic edema restricts water diffusion, causing high DWI signal and a low apparent diffusion coefficient.
Question 33: When positioning a patient for a lumbar spine MRI, the technologist should place the patient in which position to reduce lumbar lordosis and improve image quality?
- Prone with legs flat
- Lateral decubitus position
- Seated upright
- Supine with knees flexed and supported by a foam wedge (Correct answer)
Correct answer: Supine with knees flexed and supported by a foam wedge
Flexing the knees with a foam wedge support flattens lumbar lordosis, reducing the curvature and aligning disc spaces parallel to axial slices for optimal visualization.
Question 34: In a gradient echo sequence, what replaces the 180° refocusing pulse used in spin echo?
- A slice-select gradient reversal
- A phase-encoding rewinder gradient
- A second 90° RF pulse
- A reversed readout gradient (Correct answer)
Correct answer: A reversed readout gradient
Gradient echo sequences use a reversed readout gradient to rephase spins, making them sensitive to T2* rather than T2.
Question 35: Parallel imaging techniques such as SENSE or GRAPPA shorten scan time mainly by:
- Raising the main magnetic field strength
- Reducing the number of frequency-encoding samples
- Skipping some phase-encoding lines and using multi-element coil sensitivity data to fill them in (Correct answer)
- Increasing the TR
Correct answer: Skipping some phase-encoding lines and using multi-element coil sensitivity data to fill them in
Parallel imaging undersamples k-space in the phase-encoding direction by an acceleration factor (R). It then uses the spatial sensitivity profiles of a multi-channel phased-array coil to unfold the aliased image (SENSE) or rebuild the missing lines (GRAPPA). The trade-off is lower SNR, roughly by the square root of R multiplied by the coil geometry (g) factor.
Question 36: Gadolinium contrast extravasation at the IV injection site is classified as which type of adverse event?
- Local injection site reaction (extravasation) (Correct answer)
- Non-allergic physiologic reaction
- Vascular reaction
- Allergic-like reaction
Correct answer: Local injection site reaction (extravasation)
Extravasation occurs when contrast leaks from the vein into surrounding tissue; it is a local injection site reaction, not a systemic allergic event.
Question 37: Provide more coverage area without sacrificing signal-to-noise ratio by using numerous coils and receivers.
- Vertical
- Phased array (Correct answer)
- Quadrature
- Helmholtz pair
Correct answer: Phased array
Phased array coils consist of multiple small, independent receiver coils, each with its own receiver channel. This innovative design allows for a larger field of view (coverage area) without sacrificing the high signal-to-noise ratio (SNR) typically associated with smaller coils. The signals from individual coils are then combined to form a comprehensive and high-quality image.
Question 38: The b-value in diffusion-weighted imaging controls the degree of diffusion weighting. Increasing the b-value:
- Increases overall image SNR
- Decreases T2 contribution
- Increases diffusion contrast but decreases SNR (Correct answer)
- Has no effect on image quality
Correct answer: Increases diffusion contrast but decreases SNR
Higher b-values provide greater diffusion contrast by increasing the dephasing of mobile water protons, but at the cost of reduced SNR due to greater signal attenuation from all spins.
Question 39: Which standard governs the format and network transfer of MR images between the scanner, workstations, and PACS?
- HIPAA
- DICOM (Correct answer)
- HL7
- JPEG 2000
Correct answer: DICOM
DICOM (Digital Imaging and Communications in Medicine) is the standard for storing, formatting, and transmitting medical images and their metadata between modalities, PACS, and viewing stations. HL7 handles text-based clinical and administrative messages. HIPAA is a privacy law, not an image format.
Question 40: Diffusion tensor imaging (DTI) and tractography are primarily used to evaluate which brain structure?
- Metabolite concentrations
- Cerebral blood flow
- Gray matter cortical thickness
- White matter fiber tracts and connectivity (Correct answer)
Correct answer: White matter fiber tracts and connectivity
DTI measures the directional diffusion of water in white matter to map axonal fiber tracts, enabling tractography of major pathways such as the corticospinal tract.
Question 41: Which MRI zone is defined as the area containing the MRI scanner itself where the static magnetic field is always present?
- Zone I
- Zone III
- Zone II
- Zone IV (Correct answer)
Correct answer: Zone IV
Zone IV is the MRI scanner room where the magnet is always on and unrestricted ferromagnetic items pose immediate danger.
Question 42: Positive susceptibility is present in these materials, however when exposed to an external magnetic field, they will stay magnetized even after the field is removed.
- Paramagnetic
- Ferromagnetic (Correct answer)
- Diamagnetic
- Superpapmagnetic
Correct answer: Ferromagnetic
Ferromagnetic materials are characterized by strong positive susceptibility and the ability to retain their magnetization even after the external magnetic field is removed. This property makes them suitable for creating permanent magnets. Examples include iron, nickel, and cobalt, which are strongly attracted to magnets and can become permanently magnetized.
Question 43: Partial Fourier (half-NEX) imaging cuts scan time by acquiring a little more than half of k-space. What property of k-space makes this possible?
- Gradient linearity
- T2* decay
- Chemical shift
- Conjugate (Hermitian) symmetry (Correct answer)
Correct answer: Conjugate (Hermitian) symmetry
In theory k-space is conjugate-symmetric: data in one half mirrors the data in the opposite half. So a little more than half can be acquired and the rest calculated. This saves time but lowers SNR, because less data is actually sampled.
Question 44: MR Conditional implants are safe to scan only when:
- The patient gives verbal consent
- Specific conditions of field strength, SAR, and gradient limits are met (Correct answer)
- The radiologist approves verbally
- The scan time is under 30 minutes
Correct answer: Specific conditions of field strength, SAR, and gradient limits are met
MR Conditional devices are safe only when the defined conditions of use (field strength, SAR, gradient dB/dt, scan duration) are strictly followed.
Question 45: In contrast to the spin echo sequence (SE), the gradient echo sequence (GRE)
- has a T2 dephasing pattern and no 180-degree pulse.
- possesses a 180-degree pulse and T2* dephasing.
- has T2* dephasing and lacks a 180-degree pulse. (Correct answer)
- possesses a 180-degree pulse and T2 dephasing
Correct answer: has T2* dephasing and lacks a 180-degree pulse.
Gradient Echo (GRE) sequences fundamentally differ from Spin Echo (SE) sequences because they do not use a 180-degree radiofrequency refocusing pulse. Instead, GRE sequences employ a gradient reversal to rephase the spins and create an echo. Since there is no 180-degree pulse to correct for static magnetic field inhomogeneities, GRE sequences are sensitive to T2* dephasing, making them more prone to susceptibility artifacts than SE sequences.
Question 46: Which artifact appears as parallel bright and dark bands in the phase-encoding direction when the FOV is too small and the patient extends beyond it?
- Aliasing (wraparound) (Correct answer)
- Chemical shift type 2
- Susceptibility artifact
- Gibbs ringing
Correct answer: Aliasing (wraparound)
When anatomy outside the FOV is phase-encoded, it aliases (wraps around) and appears as a superimposed copy on the opposite edge of the image, sometimes appearing as bands.
Question 47: Which parameter directly controls the degree of T2 weighting in a spin echo sequence?
- TR (repetition time)
- TE (echo time) (Correct answer)
- Bandwidth
- Flip angle
Correct answer: TE (echo time)
TE determines how long transverse magnetization decays before sampling; a longer TE allows more T2 decay, increasing T2 contrast.
Question 48: Which of the following precautions is required before administering a GBCA to a breastfeeding mother, according to ACR recommendations?
- Breastfeeding should be interrupted for 24–48 hours after GBCA administration
- Breastfeeding must be permanently stopped
- No precautions are required as only trace amounts enter breast milk (Correct answer)
- The infant must receive a blood test before the mother can resume breastfeeding
Correct answer: No precautions are required as only trace amounts enter breast milk
The ACR states that GBCAs enter breast milk in very small amounts (<0.04%) and oral bioavailability is negligible; no interruption is required, though the mother may choose to withhold for 12–24 hours if concerned.
Question 49: What document must candidates agree to when submitting an ARRT MRI registry application?
- ASRT Scope of Practice
- JRCERT Accreditation Standards
- ARRT Standards of Ethics (Correct answer)
- ACR Practice Parameter for MRI
Correct answer: ARRT Standards of Ethics
All ARRT applicants must attest to adhering to the ARRT Standards of Ethics as a condition of application.
Question 50: Echo planar imaging (EPI) is used in which MRI techniques due to its extremely fast acquisition speed?
- DWI, fMRI, and perfusion imaging (Correct answer)
- STIR and inversion recovery sequences
- MR arthrography and spectroscopy
- Conventional T1 spin echo and MRCP
Correct answer: DWI, fMRI, and perfusion imaging
EPI acquires an entire k-space in a single excitation, enabling the speed needed for diffusion-weighted imaging, fMRI BOLD imaging, and dynamic perfusion studies.
Question 51: The placement of the slice is established by the
- phase difference
- Radiofrequency coil's transmit frequency
- frequency of the RF pulse received
- The RF pulse's transmit frequency (Correct answer)
Correct answer: The RF pulse's transmit frequency
The placement of the slice in MRI is precisely established by the transmit frequency of the RF pulse. When a magnetic field gradient is applied, protons at different locations precess at different frequencies. By selecting a specific RF transmit frequency, only protons resonating at that exact frequency within the gradient field will be excited, thereby defining the slice's position.
Question 52: On a sagittal MRI of the cervical spine, the posterior longitudinal ligament appears as a thin structure along which surface of the vertebral bodies?
- Lateral
- Anterior
- Superior
- Posterior (Correct answer)
Correct answer: Posterior
The posterior longitudinal ligament runs along the posterior surfaces of the vertebral bodies inside the spinal canal, anterior to the thecal sac.
Question 53: Fast spin echo sequences, in contrast to spin echo, have
- Diminished susceptibility and T2 weighting
- T2 weighting decreased but susceptibility rose
- T2 weighting increased but susceptibility decreased (Correct answer)
- elevated susceptibility and elevated T2 weighting
Correct answer: T2 weighting increased but susceptibility decreased
Fast Spin Echo (FSE) sequences utilize multiple 180-degree refocusing pulses, which effectively reduce the impact of magnetic susceptibility artifacts by repeatedly rephasing spins. This leads to decreased susceptibility artifacts compared to conventional spin echo. Additionally, FSE sequences can achieve stronger T2 weighting due to the extended echo train and multiple echoes contributing to the signal, often resulting in increased T2 weighting.
Question 54: Which MR labeling symbol is a red circle with a diagonal line through it?
- MR Compatible
- MR Safe
- MR Conditional
- MR Unsafe (Correct answer)
Correct answer: MR Unsafe
Under ASTM F2503, MR Unsafe items are labeled with a red circle and a diagonal slash.
Question 55: The anterior and posterior horns of the lateral ventricles are best evaluated for periventricular leukomalacia (PVL) on which MRI plane?
- Coronal only
- Oblique
- Axial and coronal (Correct answer)
- Sagittal only
Correct answer: Axial and coronal
Axial and coronal planes together best demonstrate periventricular T2 signal changes and volume loss around the lateral ventricular horns in PVL.
Question 56: By adding more loops and circuitry, these coils increase the efficiency with which the MR signal is induced in the coil and raise SNR by about 40% when compared to coils of the same size.
- Quadrature (Correct answer)
- Helmholtz pair
- Vertical
- Phased array
Correct answer: Quadrature
Quadrature coils (or circularly polarized coils) utilize two receiver coils oriented 90 degrees to each other. By combining the signals from both coils, they can detect the rotating component of the MR signal more efficiently. This design leads to a theoretical 41% increase in signal-to-noise ratio (SNR) compared to a single linear coil of the same size, improving image quality.
Question 57: To form a magnet field, wires are arranged side by side.
- Selenoid magnet (Correct answer)
- Resistive magnet
- Superconducting
- Permanent magnet
Correct answer: Selenoid magnet
A solenoid magnet is formed by coiling a wire into a cylindrical shape. When an electric current passes through this coiled wire, it generates a uniform magnetic field along the axis of the cylinder. This design is fundamental to many types of MRI magnets, particularly superconducting systems, where the solenoid configuration efficiently produces a strong, homogeneous field.
Question 58: Which of the following changes will increase spatial resolution without changing the FOV?
- Decreasing the matrix size
- Increasing NEX
- Increasing the matrix size (Correct answer)
- Increasing slice thickness
Correct answer: Increasing the matrix size
Increasing matrix size means more phase and frequency encoding steps, reducing voxel size and improving spatial resolution within the same FOV.
Question 59: The corpus callosum is best evaluated on which MRI plane for assessing callosal agenesis?
- Oblique
- Coronal
- Sagittal (Correct answer)
- Axial
Correct answer: Sagittal
The sagittal midline plane provides the best single view of the entire corpus callosum from genu to splenium for assessing agenesis or dysgenesis.
Question 60: The magnetic field's intensity is varied from head to foot by this gradient coil.
- X
- Z (Correct answer)
- Y
Correct answer: Z
In most MRI systems, the main magnetic field (B0) is oriented along the Z-axis, which typically runs from the patient's head to foot. The Z-gradient coil is specifically designed to create a linear variation in the magnetic field strength along this Z-axis. This allows for precise slice selection and spatial encoding in the head-to-foot dimension.
Question 61: Dynamic susceptibility contrast (DSC) perfusion imaging of the brain relies on which effect during the first pass of a gadolinium bolus?
- Labeling of arterial water with an inversion pulse
- Magnetization transfer between bound and free water
- T1 shortening that increases signal on spin echo images
- Transient T2* signal loss on rapid gradient echo EPI images (Correct answer)
Correct answer: Transient T2* signal loss on rapid gradient echo EPI images
Concentrated gadolinium in the capillaries creates local susceptibility gradients, causing a temporary T2* signal drop that is tracked over time.
Question 62: Which parameter in MRI controls the field of view (FOV) and therefore directly affects spatial resolution when matrix size is held constant?
- TR
- Receiver bandwidth
- TE
- FOV size relative to matrix (voxel size = FOV/matrix) (Correct answer)
Correct answer: FOV size relative to matrix (voxel size = FOV/matrix)
Voxel size = FOV divided by matrix; decreasing FOV while maintaining matrix size decreases voxel size, increasing spatial resolution but reducing SNR.
Question 63: For a contrast-enhanced MRA of the aorta, the bolus tracking technique monitors contrast arrival in which vessel to trigger the 3D acquisition?
- Right atrium
- Abdominal aorta or descending thoracic aorta at the level of interest (Correct answer)
- Superior mesenteric artery
- Femoral artery
Correct answer: Abdominal aorta or descending thoracic aorta at the level of interest
A monitoring slice is placed at the aorta at the target level; when the contrast bolus arrives (signal threshold), the 3D acquisition is automatically triggered.
Question 64: The standard dose of a non-specific extracellular GBCA for most MRI examinations is:
- 0.01 mmol/kg
- 0.3 mmol/kg
- 0.05 mmol/kg
- 0.1 mmol/kg (Correct answer)
Correct answer: 0.1 mmol/kg
The standard dose for most extracellular GBCAs is 0.1 mmol/kg body weight, which provides adequate contrast enhancement for most clinical applications.
Question 65: On MRI, cortical bone appears as which signal intensity on both T1 and T2 weighted images?
- Intermediate signal
- Low signal (signal void) (Correct answer)
- Variable signal
- High signal
Correct answer: Low signal (signal void)
Cortical bone has very few mobile protons and produces a signal void (low signal) on both T1 and T2 weighted MRI sequences.
Question 66: The normal intervertebral disc nucleus pulposus appears as which signal on T2-weighted MRI in a healthy young adult?
- Signal void
- Intermediate signal
- Low signal
- High signal (bright) (Correct answer)
Correct answer: High signal (bright)
The healthy nucleus pulposus is rich in water-containing proteoglycans and appears bright (high signal) on T2-weighted sequences.
Question 67: Susceptibility-weighted imaging (SWI) is particularly sensitive to which types of tissue or substances?
- Gadolinium-enhanced tissue
- Myelinated white matter
- Blood products (hemosiderin, deoxyhemoglobin), calcium, and iron-containing structures (Correct answer)
- Fat and water interfaces
Correct answer: Blood products (hemosiderin, deoxyhemoglobin), calcium, and iron-containing structures
SWI exploits magnetic susceptibility differences between tissues; hemosiderin, deoxyhemoglobin, calcium, and iron create local field distortions that appear as marked hypointensity on SWI images.
Question 68: Parallel imaging techniques such as GRAPPA and SENSE reduce scan time by:
- Applying stronger gradient pulses
- Increasing the magnetic field strength
- Using multiple receiver coil elements to partially fill k-space and reduce phase encoding steps (Correct answer)
- Eliminating the need for radiofrequency pulses
Correct answer: Using multiple receiver coil elements to partially fill k-space and reduce phase encoding steps
Parallel imaging exploits spatial sensitivity information from multiple coil elements to skip phase encoding lines, reconstructing the full image and reducing acquisition time.
Question 69: Which MRI sequence is the standard for evaluating the internal structure of bone marrow and is most sensitive to marrow infiltration or edema?
- T1-weighted spin-echo
- T2*-weighted gradient echo
- Proton density without fat suppression
- STIR or fat-suppressed T2-weighted FSE (Correct answer)
Correct answer: STIR or fat-suppressed T2-weighted FSE
STIR and fat-suppressed T2 sequences null normal fatty marrow signal, making pathological edema, infiltration, or tumor stand out as high signal against the nulled fat background.
Question 70: IV line tubing and monitoring cables used inside the MRI suite must be:
- MR-conditional equipment specifically rated for use in the magnetic field (Correct answer)
- Standard clinical-grade equipment from any hospital supply
- Removed prior to scanning and replaced after
- Made of stainless steel only
Correct answer: MR-conditional equipment specifically rated for use in the magnetic field
All equipment used within the MRI environment must be labeled MR-conditional or MR Safe to prevent projectile events, burns, or device malfunction.
Question 71: Which of the following best describes the concept of signal-to-noise ratio (SNR) in MRI?
- The ratio of mean signal intensity in a region of interest to the standard deviation of background noise (Correct answer)
- The number of signal averages divided by scan time
- The difference in signal between two adjacent tissues
- The ratio of TE to TR
Correct answer: The ratio of mean signal intensity in a region of interest to the standard deviation of background noise
SNR is quantitatively defined as the mean signal from a region of interest divided by the standard deviation of noise measured in a background air region.
Question 72: On an axial MRI of the abdomen at the level of L1, which structure is seen as a round low-signal structure anterior to the aorta?
- Superior mesenteric artery (Correct answer)
- Portal vein
- Inferior vena cava
- Celiac axis
Correct answer: Superior mesenteric artery
The superior mesenteric artery (SMA) arises from the anterior aorta at approximately L1 and appears as a round flow-void structure surrounded by fat on axial images.
Question 73: Which artifact arises when signal from outside the FOV wraps around to the opposite side of the image?
- Chemical shift artifact
- Gibbs ringing artifact
- Aliasing (wrap-around) artifact (Correct answer)
- Susceptibility artifact
Correct answer: Aliasing (wrap-around) artifact
Aliasing occurs in the phase-encode direction when anatomy outside the FOV is undersampled, causing it to appear on the opposite side.
Question 74: Chemical shift artifact of the first kind occurs at interfaces between fat and water and manifests as:
- Image wrap-around from the opposite side
- Zebra-stripe banding artifact
- Signal voids around metallic implants
- A bright band and a dark band at the fat-water interface in the frequency-encoding direction (Correct answer)
Correct answer: A bright band and a dark band at the fat-water interface in the frequency-encoding direction
First-kind chemical shift artifact creates a bright rim on one side and a dark void on the other side at fat-water interfaces in the frequency-encoding direction due to the 3.5 ppm resonance frequency difference.
Question 75: A patient reports a prior allergic-like reaction to a GBCA. Which of the following is the recommended premedication protocol?
- Corticosteroid + antihistamine, with the first steroid dose 13 hours before injection (Correct answer)
- No premedication is effective for GBCA reactions
- Aspirin 24 hours before
- Antihistamine only, 1 hour before
Correct answer: Corticosteroid + antihistamine, with the first steroid dose 13 hours before injection
The ACR-recommended premedication protocol for prior GBCA reactions includes corticosteroids given 13, 7, and 1 hour before contrast, combined with an antihistamine 1 hour before.
Question 76: Which section of the ARRT MRI content specifications covers pulse sequence selection and acquisition parameter optimization?
- Procedures
- Image Production (Correct answer)
- Physics and Instrumentation
- Patient Care and Education
Correct answer: Image Production
Image Production covers acquisition parameters, pulse sequences, and image quality optimization.
Question 77: A patient reports having a cardiac pacemaker. What is the standard MRI safety classification for most older pacemakers?
- MR Safe
- MR Conditional
- MR Compatible
- MR Unsafe (Correct answer)
Correct answer: MR Unsafe
Most older pacemakers are classified MR Unsafe because the magnetic field can interfere with device function or cause heating.
Question 78: A patient with a GFR of 45 mL/min/1.73m² requires a contrast-enhanced MRI. According to ACR guidelines, which GBCA group is preferred?
- Group III macrocyclic agents (Correct answer)
- Group II linear non-ionic agents
- Group I linear ionic agents
- Contrast should be completely avoided
Correct answer: Group III macrocyclic agents
The ACR recommends macrocyclic GBCAs (Group III) for patients with reduced renal function (GFR 30–60) due to their superior thermodynamic and kinetic stability.
Question 79: In the ACR zone system, who is responsible for controlling access to Zone III?
- Any hospital staff with ID badge
- Level 2 MR personnel (Correct answer)
- Radiology technologists only
- MR-trained non-MR personnel
Correct answer: Level 2 MR personnel
Zone III access must be controlled by Level 2 MR personnel (those with full MR safety training) to prevent non-screened individuals from entering.
Question 80: Which coil type provides the best signal-to-noise ratio (SNR) for a specific small anatomy because it is placed closest to the structure?
- Body coil
- Surface coil (Correct answer)
- Head coil
- Volume coil
Correct answer: Surface coil
Surface coils are placed directly over the anatomy of interest; their proximity maximizes SNR for superficial or small structures compared to larger volume coils.
Question 81: A T1 hyperintense (bright) lesion in the brain is most consistent with which finding?
- Pure fluid-filled cyst
- Acute infarct
- Subacute blood or fat-containing lesion (Correct answer)
- Vasogenic edema
Correct answer: Subacute blood or fat-containing lesion
T1 hyperintensity in the brain is classically associated with subacute hemorrhage (methemoglobin), fat, protein-rich fluid, or melanin.
Question 82: What is the purpose of applying a saturation band in MRI?
- To increase SNR from the target anatomy
- To suppress signal from specified tissue or flowing blood to reduce artifacts (Correct answer)
- To shorten the overall scan time
- To improve spatial resolution in the phase direction
Correct answer: To suppress signal from specified tissue or flowing blood to reduce artifacts
Saturation bands apply an RF pulse to a defined region, nulling its magnetization so it contributes no signal, which suppresses motion or flow-related artifacts.
Question 83: A 'DC offset' or 'center dot' artifact seen as a bright spot in the center of the image is related to:
- B1 field inhomogeneity
- Patient positioning off-isocenter
- A small DC current offset in the receiver causing a spike in the center of k-space (Correct answer)
- Excessive TR causing saturation effects
Correct answer: A small DC current offset in the receiver causing a spike in the center of k-space
A DC offset artifact occurs when a constant voltage offset in the MR receiver chain introduces a spike at the origin of k-space (DC component), appearing as a bright dot at image center.
Question 84: If all other parameters stay the same, how does SAR change when a patient moves from a 1.5T to a 3T scanner?
- It is cut in half
- It doubles
- It roughly quadruples (Correct answer)
- It stays the same
Correct answer: It roughly quadruples
SAR is proportional to the square of the static field strength, so doubling B0 roughly quadruples SAR.
Question 85: Inversion recovery sequences null a specific tissue when the readout pulse is applied at time TI equal to:
- 0.693 × T1 of the tissue (T1 × ln2) (Correct answer)
- 2 × T1 of the tissue
- T1/2 of the tissue
- 0.5 × TE
Correct answer: 0.693 × T1 of the tissue (T1 × ln2)
The longitudinal magnetization of a tissue crosses zero at TI = T1 × ln(2) ≈ 0.693 × T1; applying the readout at this moment nulls that tissue's signal.
Question 86: What is the effect of reducing the number of phase-encoding steps while keeping FOV and frequency steps constant?
- Through-plane resolution is affected
- SNR improves and scan time increases
- Scan time decreases and phase-direction spatial resolution decreases (Correct answer)
- Frequency-direction resolution decreases
Correct answer: Scan time decreases and phase-direction spatial resolution decreases
Fewer phase-encoding steps means less time to fill k-space (shorter scan) but larger voxels in the phase direction, reducing spatial resolution in that dimension.
Question 87: MR spectroscopy (MRS) of the brain measures metabolite concentrations. Which metabolite, reduced in neuronal loss and gliosis, is considered a marker of neuronal integrity?
- Lactate (Lac)
- Choline (Cho)
- N-acetylaspartate (NAA) (Correct answer)
- Creatine (Cr)
Correct answer: N-acetylaspartate (NAA)
NAA (N-acetylaspartate) is located predominantly in neurons; its reduction indicates neuronal loss or dysfunction, making it a key spectroscopic marker of neuronal integrity.
Question 88: Which MRI sequence is most commonly used for evaluating liver lesions using T2 contrast with heavy T2 weighting for the bile ducts?
- DWI b1000
- T1 in-phase/opposed-phase
- STIR of the abdomen
- MRCP (heavily T2-weighted 3D or thick-slab) (Correct answer)
Correct answer: MRCP (heavily T2-weighted 3D or thick-slab)
MRCP uses heavily T2-weighted sequences to make bile and pancreatic fluid appear bright, providing a non-invasive map of the biliary and pancreatic ductal system.
Question 89: Substances in the presence of an applied magnetic field that show a modest rise in magnetic field. They exhibit both positive and low susceptibility.
- Ferromagnetic
- Superpapmagnetic
- Diamagnetic
- Paramagnetic (Correct answer)
Correct answer: Paramagnetic
Paramagnetic materials exhibit a weak, positive susceptibility to an external magnetic field. Their unpaired electrons align with the field, causing a modest increase in the local magnetic field, but this effect is lost once the external field is removed. Gadolinium, a common MRI contrast agent, is a well-known paramagnetic substance.
Question 90: A 'blooming' artifact on gradient echo (GRE) MRI is caused by which substance?
- Gadolinium contrast
- Hemosiderin or calcification (Correct answer)
- Fat
- Cerebrospinal fluid
Correct answer: Hemosiderin or calcification
Hemosiderin, calcium, and other susceptibility-inducing substances cause local field inhomogeneities that appear as signal loss blooming on GRE sequences.
Question 91: For imaging the brachial plexus, which specialized MRI sequence is most useful for nerve visualization?
- T1 axial post-contrast
- GRE in-phase
- Diffusion-weighted imaging
- STIR or fat-suppressed T2 coronal (Correct answer)
Correct answer: STIR or fat-suppressed T2 coronal
Fat-suppressed T2 or STIR coronal images suppress the surrounding fat signal and highlight the high-water-content neural tissue of the brachial plexus.
Question 92: Which of the following maneuvers most directly reduces chemical shift artifact of the first kind?
- Increase field of view
- Apply fat suppression
- Increase receiver bandwidth (Correct answer)
- Switch from spin echo to gradient echo
Correct answer: Increase receiver bandwidth
Increasing the receiver bandwidth reduces the chemical shift displacement (measured in pixels) between fat and water signals along the frequency-encode direction.
Question 93: What is the primary trade-off when reducing the field of view (FOV) in MRI?
- Improved SNR with no change in resolution
- Decreased scan time and improved SNR
- Improved spatial resolution but increased risk of aliasing artifacts (Correct answer)
- Increased slice coverage with same resolution
Correct answer: Improved spatial resolution but increased risk of aliasing artifacts
Reducing FOV increases spatial resolution by making voxels smaller, but if anatomy extends beyond the reduced FOV, aliasing (wrap-around) artifacts occur.
Question 94: Magnets formed of ferrous material found naturally or from blocks or slabs.
- Selenoid
- Reistive
- Superconducting
- Permanent (Correct answer)
Correct answer: Permanent
Permanent magnets are created from naturally occurring ferromagnetic materials or specially treated blocks of ferrous material. These magnets retain their magnetism without requiring an external power source. They produce a constant magnetic field and are typically utilized in low-field MRI systems, offering a simpler and less costly design.
Question 95: Most superconducting magnets have a (horizontal or vertical) magnetiic field and are of the design type __.
- Selenoid magnet (Correct answer)
- Permanent magnet
- Superconducting
- Resistive magnet
Correct answer: Selenoid magnet
Most high-field superconducting MRI magnets are designed as solenoid magnets. In this configuration, the superconducting wire coils are arranged to create a strong, uniform magnetic field that runs horizontally through the bore of the scanner. This solenoid design is highly efficient for generating the powerful and homogeneous fields required for advanced MRI imaging.
Question 96: Which classification of GBCAs carries the LOWEST risk for NSF based on ACR guidance?
- Group II (linear non-ionic)
- Group I (linear ionic)
- Group III (macrocyclic) (Correct answer)
- All GBCAs carry equal NSF risk
Correct answer: Group III (macrocyclic)
Macrocyclic GBCAs (Group III) have a more stable cage-like structure that resists transmetallation and are associated with the lowest NSF risk.
Question 97: Which property of hydrogen makes it the most commonly used nucleus in clinical MRI?
- Low atomic mass and zero spin
- High natural abundance and large magnetic moment (Correct answer)
- High atomic number and radioactivity
- Absence of electron cloud interference
Correct answer: High natural abundance and large magnetic moment
Hydrogen (¹H) is abundant in the body and has a large gyromagnetic ratio, producing strong MRI signals.
Question 98: The echo time (TE) in gradient-echo sequences affects which type of contrast?
- TR governs all contrast in GRE sequences
- T1 contrast only
- Proton density contrast only
- T2* contrast — longer TE increases T2* weighting and susceptibility effects (Correct answer)
Correct answer: T2* contrast — longer TE increases T2* weighting and susceptibility effects
In gradient-echo sequences, TE controls T2* contrast; longer TE allows greater T2* dephasing, increasing susceptibility effects from blood products, iron, and air-tissue interfaces.
Question 99: What does the acronym STIR stand for, and what tissue does it primarily null?
- Standard Tissue Inversion Recovery — nulls all tissue equally
- Short-TE Inversion Recovery — nulls fat signal
- Steady-state T1 Inversion Recovery — nulls water signal
- Short-T1 Inversion Recovery — nulls fat signal (Correct answer)
Correct answer: Short-T1 Inversion Recovery — nulls fat signal
STIR (Short-T1 Inversion Recovery) applies an inversion pulse with a TI timed to null fat signal (TI ≈ 150 ms at 1.5T), providing robust fat suppression independent of B0 homogeneity.
Question 100: Which GBCA is approved by the FDA as a liver-specific agent taken up by hepatocytes?
- Gadobenate dimeglumine (MultiHance)
- Gadopentetate dimeglumine (Magnevist)
- Gadobutrol (Gadavist)
- Gadoxetate disodium (Eovist/Primovist) (Correct answer)
Correct answer: Gadoxetate disodium (Eovist/Primovist)
Gadoxetate disodium (Eovist in the US, Primovist in Europe) is taken up by functioning hepatocytes via OATP transporters, providing hepatobiliary phase imaging for liver lesion characterization.
Question 101: Spin echo sequence is modified by rapid spin echo sequence by
- Adding repeated pulses of 90 degrees
- Gradient additions for spoilers
- progressively adding gradients encoded with frequencies
- Adding 180-degree pulses in succession (Correct answer)
Correct answer: Adding 180-degree pulses in succession
The rapid spin echo (also known as Fast Spin Echo or Turbo Spin Echo) sequence modifies the conventional spin echo by applying a train of multiple 180-degree refocusing pulses after a single 90-degree excitation pulse. Each 180-degree pulse generates a separate echo, allowing multiple lines of k-space to be acquired within a single TR period. This significantly shortens scan times compared to conventional spin echo sequences.
Question 102: Which patient has the highest risk of nephrogenic systemic fibrosis after receiving gadolinium?
- eGFR of 65 mL/min/1.73 m²
- eGFR of 95 mL/min/1.73 m²
- eGFR below 30 mL/min/1.73 m² or on dialysis (Correct answer)
- A patient with a mild contrast allergy history
Correct answer: eGFR below 30 mL/min/1.73 m² or on dialysis
NSF risk is mainly in severe chronic kidney disease or acute kidney injury, especially with higher-risk (group I) agents.
Question 103: Which of the following parameters most directly controls T1 contrast in a spin echo sequence?
- Bandwidth
- Echo time (TE)
- Flip angle in gradient echo
- Repetition time (TR) (Correct answer)
Correct answer: Repetition time (TR)
TR determines how much longitudinal relaxation occurs between excitations; a short TR emphasizes T1 differences because tissues with shorter T1 recover faster.
Question 104: On diffusion-weighted MRI (DWI), an acute ischemic infarct typically shows:
- Low signal on both DWI and ADC maps
- High signal on DWI and high ADC values
- High signal on DWI and low ADC values (Correct answer)
- Low signal on DWI and high ADC values
Correct answer: High signal on DWI and low ADC values
Cytotoxic edema in acute infarction restricts water diffusion, producing high signal on DWI and low apparent diffusion coefficient (ADC) values.
Question 105: Gadolinium retention in the brain has been observed particularly in which structure on unenhanced T1-weighted images in patients receiving multiple GBCA doses?
- Caudate nucleus
- Thalamus
- Hippocampus
- Dentate nucleus of the cerebellum (Correct answer)
Correct answer: Dentate nucleus of the cerebellum
Gadolinium deposition in the dentate nucleus and globus pallidus manifests as T1 hyperintensity on unenhanced brain MRI after multiple GBCA administrations.
Question 106: Multiple sclerosis (MS) plaques in the brain are classically found in which location on MRI?
- Periventricular white matter, corpus callosum, and juxtacortical regions (Correct answer)
- Basal ganglia bilaterally
- Cerebellum exclusively
- Gray matter of the cortex only
Correct answer: Periventricular white matter, corpus callosum, and juxtacortical regions
MS demyelinating plaques characteristically occur in periventricular white matter, corpus callosum (Dawson fingers), juxtacortical, and infratentorial regions.
Question 107: Which type of aneurysm clip is considered MR Unsafe regardless of age or appearance?
- Platinum clips
- Titanium clips
- Cobalt-chromium clips
- Ferromagnetic stainless steel clips (Correct answer)
Correct answer: Ferromagnetic stainless steel clips
Ferromagnetic stainless steel aneurysm clips are MR Unsafe because the static field can exert significant torque on the clip, risking rupture of the aneurysm.
Question 108: In cardiac MRI, which sequence is used to assess myocardial viability and detect scar tissue?
- True FISP cine imaging
- Black-blood T2 STIR
- Late gadolinium enhancement (LGE) with inversion recovery (Correct answer)
- First-pass perfusion with adenosine stress
Correct answer: Late gadolinium enhancement (LGE) with inversion recovery
Late gadolinium enhancement (LGE) uses an inversion recovery sequence 10–15 minutes after GBCA injection; scarred myocardium retains gadolinium and appears bright against nulled normal myocardium.
Question 109: What happens to image contrast when the echo time (TE) is increased in a T2-weighted sequence?
- T1 contrast increases
- Tissues appear uniformly bright
- T2 contrast increases as tissues with short T2 decay faster (Correct answer)
- SNR improves due to longer signal collection
Correct answer: T2 contrast increases as tissues with short T2 decay faster
Longer TE allows more T2 decay to occur, so tissues with short T2 lose signal faster and appear darker, enhancing T2 contrast between tissues.
Question 110: In MR angiography, time-of-flight (TOF) technique produces vascular signal by exploiting which physical principle?
- T2 prolongation of blood
- Inflow enhancement: unsaturated blood entering a saturated imaging slab (Correct answer)
- T1 shortening by gadolinium contrast
- Chemical shift between blood and surrounding fat
Correct answer: Inflow enhancement: unsaturated blood entering a saturated imaging slab
TOF MRA relies on flow-related enhancement: flowing unsaturated blood entering a slice that has saturated stationary tissue appears bright relative to background.
Question 111: In contrast to SE, GRE sequences employ
- shorter TE and longer TR
- reduced TR and TE lengths (Correct answer)
- longer TE and shorter TR
- greater TE and TR lengths
Correct answer: reduced TR and TE lengths
Gradient Echo (GRE) sequences typically employ significantly shorter repetition times (TR) and echo times (TE) compared to Spin Echo (SE) sequences. This is feasible because GRE sequences use smaller flip angles for excitation and do not require the longer timing associated with the 180-degree refocusing pulses. These reduced TR and TE values enable much faster image acquisition, which is advantageous for dynamic imaging or when scan speed is crucial.
Question 112: A pregnant MRI technologist should follow which guideline regarding working in the MRI suite?
- Take a leave of absence immediately upon learning of pregnancy
- Limit time in the scanner room primarily during the first trimester due to acoustic and gradient exposure (Correct answer)
- Wear a lead apron to reduce MRI exposure
- Avoid all scanner rooms during pregnancy
Correct answer: Limit time in the scanner room primarily during the first trimester due to acoustic and gradient exposure
Current guidelines suggest limiting time within the bore and active scanner room primarily in the first trimester; there is no requirement to cease work entirely.
Question 113: What does k-space represent in MRI?
- The physical space inside the magnet bore
- The gradient amplitude map
- The raw data space containing spatial frequency information from which the MR image is reconstructed by Fourier transform (Correct answer)
- The frequency range of the RF pulse
Correct answer: The raw data space containing spatial frequency information from which the MR image is reconstructed by Fourier transform
K-space is the mathematical data domain where raw MRI signal is stored as spatial frequencies; a Fourier transform converts k-space data into the final diagnostic image.
Question 114: What is the mechanism of action of gadolinium-based contrast agents (GBCAs) in MRI?
- They shorten T1 relaxation time of nearby water protons, increasing signal on T1-weighted images (Correct answer)
- They increase T2 relaxation time, increasing signal on T2-weighted images
- They block radiofrequency pulses to create contrast
- They directly emit radio waves that increase signal
Correct answer: They shorten T1 relaxation time of nearby water protons, increasing signal on T1-weighted images
Gadolinium shortens the T1 relaxation time of surrounding water protons through dipole-dipole interactions, causing increased signal intensity on T1-weighted sequences.
Question 115: What is the relationship between voxel size and SNR in MRI?
- Voxel size has no effect on SNR
- SNR decreases as voxel volume doubles
- Smaller voxels produce higher SNR
- Larger voxels produce higher SNR (Correct answer)
Correct answer: Larger voxels produce higher SNR
Larger voxels contain more protons contributing signal, increasing SNR, though at the cost of spatial resolution.
Question 116: Which anatomical landmark is used to prescribe the standard axial plane for brain MRI?
- Anterior commissure–posterior commissure (AC-PC) line (Correct answer)
- Orbitomeatal baseline
- Frankfurt plane
- Hard palate line
Correct answer: Anterior commissure–posterior commissure (AC-PC) line
Axial brain MRI slices are typically aligned parallel to the AC-PC line for standardized anatomical coverage and comparison.
Question 117: On MRI of the liver, a simple hepatic cyst is expected to show which signal characteristics?
- T1 high, T2 high, with peripheral enhancement
- T1 high, T2 low
- T1 low, T2 low
- T1 low, T2 high, no enhancement (Correct answer)
Correct answer: T1 low, T2 high, no enhancement
Simple hepatic cysts have low T1 signal, very high T2 signal due to pure fluid content, and show no contrast enhancement.
Question 118: Following an RF burst, protons will
- As per T2*, recover along B0, and thereafter dephase in accordance with T1.
- As per T2*, dephase in accordance with T1 and then recover along B0.
- Dephase in accordance with T2*, then recover in accordance with T1 along B0. (Correct answer)
- As instructed by T1, recover along B0, and then dephase as instructed by T2*.
Correct answer: Dephase in accordance with T2*, then recover in accordance with T1 along B0.
Immediately following an RF burst, the excited protons begin to lose their phase coherence, causing the transverse magnetization to decay. This dephasing is governed by T2* relaxation, which includes both intrinsic molecular interactions and static magnetic field inhomogeneities. Simultaneously, the longitudinal magnetization, which was tipped away from B0, begins to recover along the B0 field, a process governed by T1 relaxation.
Question 119: In the context of the ARRT MRI registry, 'structured education' refers to which requirement?
- Attending MRI vendor training only
- Online CE modules totaling 50 hours
- Completion of an accredited primary radiologic technology program (Correct answer)
- Hospital orientation programs
Correct answer: Completion of an accredited primary radiologic technology program
Structured education means graduating from a formal, ARRT-recognized educational program as the foundation for registry eligibility.
Question 120: Rapidly switching gradient fields (dB/dt) most directly cause which effect?
- Tissue heating from RF
- Ionization of tissue
- Peripheral nerve stimulation (Correct answer)
- Projectile accidents
Correct answer: Peripheral nerve stimulation
Fast-changing gradient fields induce currents in tissue that can stimulate peripheral nerves and cause tingling or twitching.
Question 121: When should a technologist notify the supervising radiologist or radiologist immediately during an MRI exam?
- When the patient requests a blanket
- When any adverse reaction to contrast, change in patient condition, or safety concern arises (Correct answer)
- When the patient completes the screening form
- When the scan takes longer than expected
Correct answer: When any adverse reaction to contrast, change in patient condition, or safety concern arises
Any adverse reaction to contrast agent, unexpected change in patient condition (loss of consciousness, seizure, cardiac event), or safety concern requires immediate radiologist notification.
Question 122: The 'magic angle' artifact in MRI tendons occurs at approximately what angle relative to the main magnetic field?
- 30°
- 55° (Correct answer)
- 45°
- 15°
Correct answer: 55°
Magic angle artifact occurs when a tendon is oriented at approximately 55° (the magic angle) to B0, causing artificial T1/PD signal increase that can mimic pathology.
Question 123: Which MRI screening form section is most critical for identifying implanted devices that may be MR Unsafe?
- Implants and devices section (Correct answer)
- Medication list section
- Surgical history section
- Allergies section
Correct answer: Implants and devices section
The implants and devices section identifies items such as pacemakers, cochlear implants, and aneurysm clips that may be MR Unsafe.
Question 124: The signal-to-noise ratio (SNR) in MRI is proportional to which of the following?
- 1 / (TR × TE)
- Voxel volume × √(NEX) × √(scan time factors) (Correct answer)
- Flip angle / repetition time
- Field of view / matrix size only
Correct answer: Voxel volume × √(NEX) × √(scan time factors)
SNR increases with larger voxel volume, more signal averages (NEX), and longer acquisition times that allow more k-space sampling.
Question 125: In MRI of the breast, a finding described as a 'rim-enhancing mass' on dynamic contrast imaging is most suspicious for:
- Malignant lesion (Correct answer)
- Fibroadenoma
- Simple cyst
- Fat necrosis
Correct answer: Malignant lesion
Rim enhancement of a breast mass, indicating peripheral enhancement with central necrosis, is a suspicious finding associated with malignancy.
Question 126: These materials are employed as T2 contrast agents because they show positive susceptibility, are stronger than paramagnetic materials but weaker than ferromagnetic materials.
- Ferromagnetic
- Superpapmagnetic (Correct answer)
- Paramagnetic
- Diamagnetic
Correct answer: Superpapmagnetic
Superparamagnetic materials possess magnetic properties stronger than paramagnetic substances but, unlike ferromagnetic materials, they do not retain magnetism once the external field is removed. They are often used as T2 contrast agents in MRI because their large magnetic moments create significant local field inhomogeneities, leading to rapid T2 decay and signal loss, which enhances contrast in certain tissues.
Question 127: A candidate applying for the ARRT MRI registry via the post-primary pathway must hold a primary ARRT certification in which of the following disciplines?
- Medical dosimetry
- Medical assisting
- Dental radiology
- Radiography, nuclear medicine, radiation therapy, or sonography (Correct answer)
Correct answer: Radiography, nuclear medicine, radiation therapy, or sonography
The post-primary pathway requires a current, active ARRT primary credential in an approved imaging or therapy discipline.
Question 128: On a phase contrast flow study, velocities above the selected VENC will cause:
- Signal void from intravoxel dephasing only
- Aliasing, where high velocities are shown as flow in the opposite direction (Correct answer)
- Zipper artifact across the phase direction
- Complete fat suppression failure
Correct answer: Aliasing, where high velocities are shown as flow in the opposite direction
Phase shifts beyond ±180° wrap around, so velocities exceeding the VENC are encoded as flow in the opposite direction.
Question 129: Which sequence uses both magnitude and phase data from a high-resolution 3D gradient echo acquisition to highlight hemosiderin, deoxyhemoglobin, and calcification?
- Double inversion recovery
- Susceptibility-weighted imaging (SWI) (Correct answer)
- Arterial spin labeling
- Diffusion tensor imaging
Correct answer: Susceptibility-weighted imaging (SWI)
SWI combines magnitude and filtered phase images to greatly increase sensitivity to susceptibility differences such as blood products and calcium.
Question 130: What is the preferred patient position and coil for imaging the rotator cuff of the shoulder?
- Supine, arm abducted 90°, flex coil
- Supine, arm at side in neutral rotation, dedicated shoulder coil (Correct answer)
- Prone, body coil
- Sitting upright, surface coil
Correct answer: Supine, arm at side in neutral rotation, dedicated shoulder coil
Supine positioning with the arm at the side in neutral rotation and a dedicated shoulder coil provides optimal SNR and comfort for rotator cuff imaging.
Question 131: Which force on a ferromagnetic object depends most directly on the spatial field gradient, not just field strength?
- Lorentz force on blood
- Translational (projectile) force (Correct answer)
- Torque
- RF heating
Correct answer: Translational (projectile) force
Translational attraction depends on the spatial gradient of the static field, which is often greatest near the bore entrance.
Question 132: For a brain MRI, which standard plane best demonstrates the corpus callosum in its entirety?
- Oblique
- Axial
- Coronal
- Sagittal (Correct answer)
Correct answer: Sagittal
The midsagittal plane shows the entire corpus callosum from the genu to the splenium as a continuous C-shaped structure.
Question 133: Which patient positioning accessory is routinely used to reduce hand and wrist movement during an MRI wrist examination?
- Foam head cushion
- Knee support wedge under the wrist
- Dedicated wrist coil with the patient prone and arm extended overhead ('Superman position') (Correct answer)
- Compression wrap around the shoulder
Correct answer: Dedicated wrist coil with the patient prone and arm extended overhead ('Superman position')
The 'Superman position' with the arm extended overhead minimizes the distance from wrist to isocenter and uses a dedicated wrist coil, optimizing SNR and reducing motion artifact.
Question 134: FLAIR (Fluid-Attenuated Inversion Recovery) suppresses which signal and is most useful for detecting:
- CSF signal; periventricular and cortical lesions (Correct answer)
- Muscle signal; soft tissue tumors
- Fat signal; bone marrow edema
- Blood signal; hemorrhage
Correct answer: CSF signal; periventricular and cortical lesions
FLAIR uses an inversion pulse timed to null CSF, making periventricular lesions and cortical pathology (MS plaques, subarachnoid hemorrhage) visible against suppressed fluid.
Question 135: What differentiates T2 from T2* is
- The combination of changing molecular magnetic field inhomogeneity and static magnetic field inhomogeneity is represented by T2*, which equals T2. (Correct answer)
- T2* is the product of changing molecular magnetic field inhomogeneity and static magnetic field inhomogeneity, which is represented by T2.
- The combination of changing molecular magnetic field inhomogeneity and static magnetic field inhomogeneity, or T2*, is represented by T2.
- The combination of changing molecular magnetic field inhomogeneity and static magnetic field inhomogeneity is represented by T2*, or T2*.
Correct answer: The combination of changing molecular magnetic field inhomogeneity and static magnetic field inhomogeneity is represented by T2*, which equals T2.
T2 relaxation describes the decay of transverse magnetization due to intrinsic molecular interactions and microscopic magnetic field inhomogeneities. T2* relaxation, however, encompasses both these intrinsic T2 effects and additional macroscopic static magnetic field inhomogeneities inherent to the scanner and patient. Therefore, T2* is always shorter than or equal to T2, as it accounts for more dephasing factors.
Question 136: This kind of magnetic shielding confines the fringe field by using metal—typically steel—in the scan room walls.
- Active
- Passive (Correct answer)
Correct answer: Passive
Passive magnetic shielding involves lining the walls, floor, and ceiling of the MRI scan room with ferromagnetic materials, typically steel. This material absorbs and redirects the stray magnetic fields, known as fringe fields, produced by the main magnet. This effectively contains the magnetic field within the room, preventing interference with external devices or personnel.
Question 137: Which mathematical process converts the raw data stored in k-space into the MR image?
- Bloch equation integration
- Maximum intensity projection
- Fourier transform (Correct answer)
- Filtered back projection
Correct answer: Fourier transform
The MR signal is recorded as frequency and phase information in k-space. A 2D or 3D Fourier transform breaks that data into its component frequencies and maps them to pixel locations and intensities. Filtered back projection is mainly a CT method, and MIP is a post-processing display technique.
Question 138: On MRI of the prostate using PI-RADS scoring, which sequence is the most dominant sequence for peripheral zone lesions?
- MR spectroscopy
- Diffusion-weighted imaging (DWI) (Correct answer)
- T1-weighted
- Dynamic contrast-enhanced (DCE)
Correct answer: Diffusion-weighted imaging (DWI)
DWI is the dominant sequence for peripheral zone lesion characterization in the PI-RADS v2.1 scoring system for prostate MRI.
Question 139: Which MRI sequence is most sensitive for detecting early bone marrow edema in stress fractures?
- STIR or fat-saturated T2 (Correct answer)
- T1-weighted SE
- GRE in-phase
- Proton density without fat saturation
Correct answer: STIR or fat-saturated T2
STIR and fat-saturated T2 sequences suppress the bright fat signal of normal marrow, making the high-signal edema of stress fractures conspicuous.
Question 140: A small bandwidth makes _slices possible.
- Slimmer (Correct answer)
- Dicker
Correct answer: Slimmer
In MRI, the slice selection gradient, in conjunction with the radiofrequency (RF) pulse, determines the slice thickness. A smaller receiver bandwidth means that a narrower range of frequencies is sampled. This corresponds to a smaller range of magnetic field strengths along the slice selection gradient, resulting in the selection of a thinner, or 'slimmer,' slice.
Question 141: Which patient condition requires consultation with the ordering physician regarding MRI safety due to potential risks during the first trimester?
- Mild hypertension controlled with oral medication
- Known diabetes mellitus
- Prior abdominal surgery more than 1 year ago
- Known pregnancy, especially first trimester (Correct answer)
Correct answer: Known pregnancy, especially first trimester
MRI in the first trimester is performed with caution; although no confirmed harm is established, ACR guidelines recommend physician consultation and clinical justification for first-trimester MRI.
Question 142: The 'late gadolinium enhancement' (LGE) pattern of subendocardial distribution in cardiac MRI is most consistent with which diagnosis?
- Myocarditis
- Cardiac sarcoidosis
- Ischemic scar (prior myocardial infarction) (Correct answer)
- Hypertrophic cardiomyopathy
Correct answer: Ischemic scar (prior myocardial infarction)
Subendocardial LGE following a coronary territory distribution is the classic pattern of ischemic scar from prior myocardial infarction.
Question 143: For a spoiled gradient echo sequence, the flip angle that gives maximum signal for a tissue with a given T1 and TR is called the:
- Ernst angle (Correct answer)
- Magic angle
- Larmor angle
- Nyquist angle
Correct answer: Ernst angle
The Ernst angle, where cos θ = e^(-TR/T1), produces the highest signal for a given TR and T1.
Question 144: When a constant magnetic field is present, hydrogen protons
- do not experience any magnetic field effects
- Align yourself along the field and revolve around its center.
- Stay basically orientated randomly while rotating around the field axis. (Correct answer)
- Ascend the field and rotate around its center.
Correct answer: Stay basically orientated randomly while rotating around the field axis.
When hydrogen protons are placed in a strong constant magnetic field (B0), they align either parallel or anti-parallel to the field, creating a slight net magnetization. However, individual protons do not simply align; they also precess (rotate) around the axis of the main magnetic field at their characteristic Larmor frequency. While there's a net alignment, their individual spin orientations remain somewhat random relative to each other as they precess.
Question 145: What is the purpose of the radiofrequency (RF) pulse in an MRI sequence?
- To tip net magnetization away from B0 (Correct answer)
- To suppress gradient coil noise
- To increase static magnetic field strength
- To align hydrogen nuclei with each other
Correct answer: To tip net magnetization away from B0
The RF pulse at the Larmor frequency tips the net magnetization vector away from the longitudinal axis to induce signal.
Question 146: Which portion of k-space primarily determines image contrast?
- The center (low spatial frequencies) (Correct answer)
- The corners of k-space only
- The edges equally with the center
- The periphery (high spatial frequencies)
Correct answer: The center (low spatial frequencies)
The center of k-space contains low spatial frequency data, which encodes the bulk signal intensity and contrast of the image.
Question 147: The ARRT MRI exam content category 'Procedures' would most likely include a question about which topic?
- Brain MRI protocol selection for stroke evaluation (Correct answer)
- Cryogen quench valve operation
- Larmor frequency calculations
- Patient screening form design standards
Correct answer: Brain MRI protocol selection for stroke evaluation
Procedures covers clinical protocols such as selecting the appropriate MRI protocol for specific pathologies.
Question 148: Which organization accredits MRI education programs whose graduates may be eligible to sit for the ARRT MRI registry?
- The Joint Commission (TJC)
- American Society of Radiologic Technologists (ASRT)
- American College of Radiology (ACR)
- Joint Review Committee on Education in Radiologic Technology (JRCERT) (Correct answer)
Correct answer: Joint Review Committee on Education in Radiologic Technology (JRCERT)
JRCERT accredits radiologic technology education programs, and graduation from an accredited program satisfies ARRT educational requirements.
Question 149: In MRI, what does the term 'k-space' represent?
- The final reconstructed image space
- The physical space inside the MRI bore
- A map of gradient field inhomogeneities
- The raw frequency-domain data before Fourier transformation into an image (Correct answer)
Correct answer: The raw frequency-domain data before Fourier transformation into an image
K-space is the mathematical frequency domain where raw MRI data is stored; a Fourier transform converts k-space data into the final spatial image.
Question 150: What is the primary advantage of parallel imaging techniques (e.g., GRAPPA, SENSE) in MRI?
- They improve T1 contrast
- They increase SNR at the same scan time
- They eliminate the need for gradient coils
- They reduce scan time or allow higher spatial resolution by using multiple coil elements to under-sample k-space (Correct answer)
Correct answer: They reduce scan time or allow higher spatial resolution by using multiple coil elements to under-sample k-space
Parallel imaging exploits spatial sensitivity differences among phased-array coil elements to reconstruct fully sampled images from under-sampled k-space, reducing scan time by the acceleration factor (R).
Question 151: Which breast MRI positioning protocol is standard for screening and diagnostic evaluation?
- Supine with arms overhead
- Upright in a dedicated open MRI
- Prone with the breasts in a dedicated bilateral breast coil (Correct answer)
- Lateral decubitus with the ipsilateral breast dependent
Correct answer: Prone with the breasts in a dedicated bilateral breast coil
Prone positioning with bilateral dedicated breast coils provides optimal tissue separation, bilateral coverage, and reproducible enhancement kinetics.
Question 152: Which structure in the knee joint is best evaluated with MRI and appears as a low-signal-intensity triangular structure?
- Medial meniscus (Correct answer)
- Anterior cruciate ligament
- Articular cartilage
- Patellar tendon
Correct answer: Medial meniscus
The medial meniscus is best visualized on MRI as a low-signal-intensity fibrocartilaginous triangular structure on coronal and sagittal images.
Question 153: Bilateral symmetric T2 hyperintensity in the basal ganglia of a child is most associated with which condition?
- Wilson's disease or metabolic disorder (Correct answer)
- Glioblastoma multiforme
- Multiple sclerosis
- Acute ischemic stroke
Correct answer: Wilson's disease or metabolic disorder
Bilateral symmetric basal ganglia T2 signal abnormality in a child raises concern for metabolic disorders such as Wilson's disease, mitochondrial disease, or toxic encephalopathy.
Question 154: What determines the slice thickness in a standard 2D MRI acquisition?
- The field of view and matrix size
- The number of phase-encoding steps
- The bandwidth of the RF pulse and the slope of the slice-select gradient (Correct answer)
- The repetition time and echo time
Correct answer: The bandwidth of the RF pulse and the slope of the slice-select gradient
Slice thickness = RF bandwidth / (gyromagnetic ratio × gradient strength); a wider RF bandwidth or shallower gradient yields a thicker slice.
Question 155: In dynamic contrast-enhanced (DCE) MRI of the breast, which kinetic enhancement curve pattern is most concerning for malignancy?
- No enhancement
- Persistent (type I)
- Plateau (type II)
- Washout (type III) (Correct answer)
Correct answer: Washout (type III)
A washout (type III) kinetic curve — rapid initial enhancement followed by signal decrease — is most suspicious for malignancy due to arteriovenous shunting in tumor neovasculature.
Question 156: Which MRI-compatible monitoring device is used to detect heart rate and trigger cardiac-gated sequences?
- Blood pressure cuff inflation timing
- Finger-clip optical pulse oximeter (used for cardiac gating)
- Standard hospital ECG with adhesive electrodes
- MR-conditional ECG with carbon or fiber-optic leads free of conductive loops (Correct answer)
Correct answer: MR-conditional ECG with carbon or fiber-optic leads free of conductive loops
MR-conditional ECG systems with non-conductive or carbon fiber leads detect the R-wave for cardiac gating without creating conductive loops that could cause burns.
Question 157: One forms a spin echo through
- changing the path of a proton using a radiofrequency pulse (Correct answer)
- Using a gradient of a negative magnetic field
- Turning B0 in the opposite direction
Correct answer: changing the path of a proton using a radiofrequency pulse
A spin echo is formed by applying a 180-degree radiofrequency (RF) pulse after an initial 90-degree RF excitation pulse. This 180-degree pulse effectively 'flips' the spins, causing them to rephase and produce a detectable echo signal. By reversing the effects of static magnetic field inhomogeneities, the 180-degree pulse allows for the recovery of signal that would otherwise be lost due to T2* dephasing.
Question 158: Which MRI technique encodes functional brain activity by detecting changes in blood oxygenation level?
- Diffusion tensor imaging (DTI)
- Arterial spin labeling (ASL)
- MR spectroscopy (MRS)
- Blood oxygen level-dependent (BOLD) fMRI (Correct answer)
Correct answer: Blood oxygen level-dependent (BOLD) fMRI
BOLD fMRI detects regional neural activity via the magnetic susceptibility difference between oxyhemoglobin and deoxyhemoglobin, providing maps of brain activation.
Question 159: The difference between the hydrogen's periodic frequency in fat and water is known as
- CHEMICAL SHIFT (Correct answer)
- FID
- FOURIER TRANSFORM
- SPIN DENSITY
Correct answer: CHEMICAL SHIFT
Chemical shift refers to the slight difference in the resonant frequency of protons when they are in different chemical environments, such as those found in fat versus water molecules. This frequency difference arises because electron shielding around protons varies with their molecular bonds, causing them to experience slightly different effective magnetic fields. This phenomenon is exploited in MRI to differentiate tissues but can also lead to artifacts if not properly accounted for.
Question 160: For temporomandibular joint (TMJ) MRI, images are obtained with the jaw in which two positions?
- Right lateral and left lateral
- Supine and prone
- Open and closed mouth (Correct answer)
- Clenched and relaxed
Correct answer: Open and closed mouth
TMJ MRI is performed with the mouth both open and closed to evaluate articular disc position and movement through the full range of condylar motion.
Question 161: What term describes an object becoming a high-speed projectile when brought near a strong MRI magnetic field?
- Missile effect (Correct answer)
- Susceptibility effect
- Quenching effect
- Gradient artifact
Correct answer: Missile effect
The missile effect (or projectile effect) occurs when ferromagnetic objects are rapidly accelerated toward the bore by the strong static magnetic field.
Question 162: Which labeling term indicates that an item has been tested and poses no known hazards in ALL MRI environments?
- MR Compatible
- MR Approved
- MR Safe (Correct answer)
- MR Conditional
Correct answer: MR Safe
The ASTM 'MR Safe' label means the item poses no known hazards in all MRI environments and is made of non-conductive, non-magnetic, non-RF-reactive materials.
Question 163: A post-contrast T1 fat-suppressed image is requested. Why is STIR a poor choice for fat suppression here?
- STIR also nulls gadolinium-enhancing tissue, since gadolinium shortens T1 toward that of fat (Correct answer)
- STIR cannot be used at field strengths above 1.0 T
- STIR adds too much chemical shift artifact
- STIR needs a TR shorter than 100 ms, which reduces SNR
Correct answer: STIR also nulls gadolinium-enhancing tissue, since gadolinium shortens T1 toward that of fat
STIR nulls signal by T1 rather than frequency, so enhancing tissue with T1 shortened toward fat's may be suppressed too.
Question 164: A 'corduroy' or 'herringbone' artifact covering the entire image is most often caused by:
- A spike (impulse noise) in k-space (Correct answer)
- Gradient eddy currents
- Patient motion
- RF transmitter miscalibration
Correct answer: A spike (impulse noise) in k-space
A single corrupted point (spike) in k-space Fourier-transforms into a sinusoidal wave overlaid across the entire image, producing a corduroy or herringbone pattern.
Question 165: Which sequence is a 3D T1-weighted gradient echo with an inversion recovery preparation pulse, widely used for volumetric brain imaging?
- MPRAGE (Correct answer)
- PROPELLER
- HASTE
- FIESTA
Correct answer: MPRAGE
MPRAGE (magnetization-prepared rapid gradient echo) uses an inversion pulse before a fast 3D GRE readout to give strong gray-white T1 contrast.
Question 166: Under IEC normal operating mode, what is the whole-body averaged SAR limit?
- 8 W/kg
- 4 W/kg
- 1 W/kg
- 2 W/kg (Correct answer)
Correct answer: 2 W/kg
Normal operating mode limits whole-body SAR to 2 W/kg, while first level controlled mode allows up to 4 W/kg.
Question 167: Traditional GRE order
- Reduce the impact of residual transverse magnetization by using spoiler gradients. (Correct answer)
- To lessen the impact of residual transverse magnetization, use long TR.
- Maintain transverse magnetization for additional attempts
Correct answer: Reduce the impact of residual transverse magnetization by using spoiler gradients.
In traditional (spoiled) Gradient Echo (GRE) sequences, spoiler gradients are applied at the end of each TR period. The primary purpose of these gradients is to intentionally dephase any residual transverse magnetization. This prevents the magnetization from contributing to the signal in subsequent TRs, ensuring that the signal primarily reflects T1 weighting and maintains consistent image contrast by avoiding contamination from previous excitations.
Question 168: For MRI of the breast, the patient should be positioned in which orientation?
- Lateral decubitus on the affected side
- Prone with breasts hanging freely in a dedicated bilateral breast coil (Correct answer)
- Upright sitting position
- Supine with arms at sides
Correct answer: Prone with breasts hanging freely in a dedicated bilateral breast coil
Prone positioning with the breasts hanging freely in dedicated bilateral surface coils minimizes motion and maximizes coil coupling for optimal breast MRI.
Question 169: Which sequence parameter directly controls the slice thickness in a 2D spin-echo acquisition?
- Bandwidth of the readout gradient
- TE duration
- TR duration
- Bandwidth of the RF pulse divided by the slice-select gradient amplitude (Correct answer)
Correct answer: Bandwidth of the RF pulse divided by the slice-select gradient amplitude
Slice thickness = RF pulse bandwidth ÷ slice-select gradient amplitude; thicker slices can be produced by widening the RF bandwidth or reducing the gradient strength.
Question 170: Which type of question format is exclusively used on the ARRT MRI registry computer-based examination?
- Matching
- True/False
- Multiple choice with one best answer (Correct answer)
- Fill-in-the-blank
Correct answer: Multiple choice with one best answer
The ARRT MRI exam uses only multiple-choice questions with four options and one best answer.
Question 171: What sound pressure level (dB) can MRI gradient noise reach, necessitating hearing protection for patients?
- 130–140 dB
- 99–115 dB (Correct answer)
- 60–70 dB
- 75–85 dB
Correct answer: 99–115 dB
MRI gradient switching can generate noise levels of 99–115 dB SPL, exceeding safe hearing thresholds and requiring earplugs or headphones for patients.
Question 172: What score does ARRT use to determine passing on the MRI registry examination?
- A criterion-referenced passing standard set by content experts (Correct answer)
- A scaled score of 75 out of 99
- A norm-referenced score based on the current testing cohort
- A raw score of 75%
Correct answer: A criterion-referenced passing standard set by content experts
ARRT uses a criterion-referenced passing standard established by a panel of content experts, not a simple percentage or norm-based cut score.
Question 173: A patient develops nausea and vomiting during IV gadolinium infusion. The most appropriate immediate action is to:
- Stop the infusion, assess the patient, and administer antiemetic if needed while monitoring for escalation (Correct answer)
- Increase infusion rate to complete dosing faster
- Administer epinephrine immediately
- Continue the infusion as nausea is expected
Correct answer: Stop the infusion, assess the patient, and administer antiemetic if needed while monitoring for escalation
Nausea and vomiting may be physiologic reactions to GBCA; the injection should be stopped, the patient assessed, symptoms treated, and the reaction monitored for potential escalation to a more serious event.
Question 174: Balanced steady-state free precession (SSFP) sequences have an advantage over conventional GRE sequences.
- Employ gradient spoilers to lessen transverse magnetization
- To create spin echoes, use 180-degree pulses.
- To create spin echoes, hold onto your transverse magnetization. (Correct answer)
- To navigate k-space, use gradients of alternating magnetic fields.
Correct answer: To create spin echoes, hold onto your transverse magnetization.
Balanced steady-state free precession (bSSFP) sequences have a key advantage over conventional spoiled GRE sequences by actively preserving transverse magnetization from previous excitations. Unlike spoiled GREs that use spoiler gradients to eliminate residual transverse magnetization, bSSFP sequences are designed to maintain both longitudinal and transverse magnetization in a steady state. This allows for the formation of both FID and echo signals, leading to high signal-to-noise ratio and characteristic T2/T1 contrast.
Question 175: On a coronal MRI of the knee, which structure appears as a low-signal C-shaped band along the tibial plateau?
- Medial meniscus (Correct answer)
- Posterior cruciate ligament
- Anterior cruciate ligament
- Iliotibial band
Correct answer: Medial meniscus
The medial meniscus appears as a low-signal C-shaped (bow-tie) structure on coronal images along the medial tibial plateau.
Question 176: Time-of-flight MRA makes flowing blood bright mainly because:
- Fresh unsaturated spins enter the slice while stationary tissue is saturated by repeated short-TR pulses (Correct answer)
- Gadolinium shortens the T1 of blood
- Blood has a much longer T2 than surrounding tissue
- Phase shifts are proportional to flow velocity
Correct answer: Fresh unsaturated spins enter the slice while stationary tissue is saturated by repeated short-TR pulses
TOF uses flow-related enhancement: inflowing fully magnetized blood gives strong signal compared with saturated stationary tissue.
Question 177: Which SAR limit (Specific Absorption Rate) is the FDA guideline for whole-body RF energy deposition in a 15-minute period for non-controlled conditions?
- 2 W/kg
- 1.5 W/kg
- 3 W/kg
- 4 W/kg (Correct answer)
Correct answer: 4 W/kg
The FDA guideline for whole-body average SAR under normal operating mode is 4 W/kg over any 15-minute period.
Question 178: Magnetization transfer contrast (MTC) in MRI is based on the exchange of magnetization between which two proton pools?
- Intracellular and extracellular water
- Arterial and venous blood
- Fat protons and water protons
- Free water protons and bound macromolecular protons (Correct answer)
Correct answer: Free water protons and bound macromolecular protons
MTC involves saturation of bound macromolecular protons, which transfer their saturation to free water protons, reducing signal in tissues with high macromolecular content.
Question 179: A 2D spin-echo scan uses TR = 500 ms, a 256 (phase) x 256 (frequency) matrix, and 2 NEX. What is the approximate scan time?
- 2 minutes 8 seconds
- 8 minutes 32 seconds
- 4 minutes 16 seconds (Correct answer)
- 1 minute 4 seconds
Correct answer: 4 minutes 16 seconds
2D scan time = TR x number of phase-encoding steps x NEX. Here that is 0.5 s x 256 x 2 = 256 s, or about 4 minutes 16 seconds. The frequency matrix does not affect scan time.
Question 180: This kind of magnetic shielding confines the main magnet's field by using other magnets and their corresponding magnetic fields.
- Active (Correct answer)
- Passive
Correct answer: Active
Active magnetic shielding uses additional superconducting coils, placed around the main magnet, that generate opposing magnetic fields. These opposing fields effectively cancel out the main magnet's fringe field at a certain distance. This method significantly reduces the extent of the stray magnetic field outside the scanner, making it a more complex but often more effective and lighter shielding solution than passive methods.
Question 181: Which positioning aid is used to maintain the ankle in a 90° neutral position during MRI to optimally visualize the Achilles tendon?
- Ankle positioning block or foam wedge (Correct answer)
- Sand bag
- Vacuum immobilization device
- Compression bandage
Correct answer: Ankle positioning block or foam wedge
A foam wedge or ankle block holds the foot perpendicular to the leg, minimizing magic angle artifact and providing consistent Achilles tendon orientation.
Question 182: Which of the following best describes the 'magic angle' artifact and at what angle it occurs?
- T2 darkening of tendons at 45° to B0
- Artifactual T1/PD signal increase in ordered collagen structures at ~55° to B0 (Correct answer)
- Chemical shift banding in fat at 55°
- Susceptibility dropout at 45° to B0
Correct answer: Artifactual T1/PD signal increase in ordered collagen structures at ~55° to B0
When ordered collagen fibers (tendons, ligaments) are oriented at ~55° to B0, dipole-dipole interactions average to zero, artifactually increasing signal on short TE sequences.
Question 183: Which of the following best describes the purpose of the ARRT MRI 'didactic' requirement in the post-primary pathway?
- Attending at least one MRI conference per year
- Passing a vendor-specific MRI operations test
- Shadowing a radiologist for 40 hours
- Completion of a formal coursework curriculum covering MRI physics, procedures, and patient care (Correct answer)
Correct answer: Completion of a formal coursework curriculum covering MRI physics, procedures, and patient care
The didactic requirement ensures candidates have formal academic instruction in MRI-specific knowledge areas before sitting for the registry.
Question 184: In which direction does the chemical shift type 2 (India ink) artifact appear at fat-water interfaces, and on which sequence type is it seen?
- In-phase direction on STIR sequences
- Phase direction on gradient-echo sequences (Correct answer)
- Frequency direction on spin-echo sequences
- Both directions on EPI
Correct answer: Phase direction on gradient-echo sequences
Chemical shift type 2 (India ink or black boundary artifact) occurs in gradient-echo sequences at fat-water voxel boundaries and appears as a dark rim at these interfaces.
Question 185: Diffusion-weighted imaging (DWI) measures the random Brownian motion of water molecules. Restricted diffusion (e.g., in acute stroke) appears as:
- High signal on DWI, low ADC (Correct answer)
- Low signal on DWI, high ADC
- High signal on DWI, high ADC
- Low signal on DWI, low ADC
Correct answer: High signal on DWI, low ADC
In acute ischemia, cytotoxic edema restricts water movement; restricted diffusion shows high signal on DWI (b=1000) and low signal on ADC map (low diffusivity).
Question 186: What is the primary reason gadolinium-based contrast agents should not be used in patients undergoing hemodialysis without a specific clinical indication?
- Gadolinium is not effectively removed by standard hemodialysis
- Risk of NSF remains even after hemodialysis removes gadolinium (Correct answer)
- Dialysis filters destroy the contrast agent before it reaches the target tissue
- Risk of anaphylaxis is higher in dialysis patients
Correct answer: Risk of NSF remains even after hemodialysis removes gadolinium
While hemodialysis can remove most gadolinium, some residual free gadolinium remains after each session, and repeated exposure can still cause NSF in high-risk patients.
Question 187: Which parameter, when increased, will most directly improve the signal-to-noise ratio (SNR) in an MRI image?
- Increasing the number of excitations (NEX/NSA) (Correct answer)
- Decreasing the field of view (FOV)
- Increasing the bandwidth
- Increasing the echo train length (ETL)
Correct answer: Increasing the number of excitations (NEX/NSA)
Increasing NEX/NSA averages more signal acquisitions, which reduces noise and improves SNR proportional to the square root of the number of averages.
Question 188: The magnetic field's intensity is changed by this gradient coil from right to left.
- Y
- X (Correct answer)
- Z
Correct answer: X
In a standard MRI coordinate system, the X-axis typically corresponds to the right-to-left (or left-to-right) direction across the patient. The X-gradient coil is engineered to produce a linear change in the magnetic field strength along this X-axis. This gradient is primarily used for spatial encoding, particularly for frequency encoding, in the right-to-left dimension.
Question 189: A method of expressing the gradient performance in terms of T/m/s, which is the gradient magnetic field's acceleration to its maximum amplitude.
- Slew rate (Correct answer)
- Coil configuration
- Rise time
Correct answer: Slew rate
Slew rate is a critical performance metric for MRI gradient coils, measuring how quickly a gradient can reach its maximum strength. It is expressed in Tesla per meter per second (T/m/s), representing the acceleration of the gradient magnetic field. A higher slew rate allows for faster switching of gradients, which translates to quicker image acquisition and shorter echo times, improving overall scan efficiency.
Question 190: Which of the following best describes the purpose of placing the frequency-encoding gradient in the anterior-posterior direction for spine MRI?
- To reduce phase wrap (aliasing) from anterior abdominal structures (Correct answer)
- To shorten scan time
- To increase SNR
- To reduce susceptibility artifact
Correct answer: To reduce phase wrap (aliasing) from anterior abdominal structures
Placing frequency encoding AP moves phase wrap artifact (aliasing) to the AP direction, preventing anterior abdominal fat from wrapping onto the spine.
Question 191: Which ACR safety zone is the MR scanner room itself?
- Zone IV (Correct answer)
- Zone I
- Zone III
- Zone II
Correct answer: Zone IV
Zone IV is the magnet room and must be clearly marked and accessible only under direct MR personnel supervision.
Question 192: Gradient echo (GRE) sequences differ from spin echo sequences primarily because GRE:
- Always requires longer scan times
- Cannot be used for dynamic contrast imaging
- Uses a 180° refocusing pulse to eliminate field inhomogeneity effects
- Uses a gradient reversal to form an echo, making it sensitive to magnetic susceptibility (T2*) (Correct answer)
Correct answer: Uses a gradient reversal to form an echo, making it sensitive to magnetic susceptibility (T2*)
GRE uses gradient reversal without a 180° refocusing pulse, making it sensitive to susceptibility effects (T2*) and allowing faster acquisition.
Question 193: Increasing the echo train length (ETL) in a fast spin echo sequence while other parameters stay the same will:
- Shorten scan time but increase image blurring and decrease the number of slices per TR (Correct answer)
- Increase scan time and reduce blurring
- Increase the number of slices per TR and improve T1 weighting
- Have no effect on scan time
Correct answer: Shorten scan time but increase image blurring and decrease the number of slices per TR
A longer ETL fills more k-space lines per TR, reducing scan time but adding T2 blurring and taking more time per slice in each TR.
Question 194: When performing MR arthrography of the hip, the patient is typically positioned with the hip in which orientation?
- Frog-leg lateral position
- Maximum external rotation
- Neutral or slight internal rotation (Correct answer)
- 30° internal rotation
Correct answer: Neutral or slight internal rotation
Neutral to slight internal rotation during hip MR arthrography places the femoral neck parallel to the imaging plane and optimizes labral visualization.
Question 195: Which of the following is the correct management for a patient who experiences an acute severe allergic-like reaction (anaphylaxis) following GBCA injection?
- Administer oral Benadryl and observe for 15 minutes
- Administer IV saline bolus only
- Administer 50 mg IV diphenhydramine and continue scan
- Stop contrast injection, call code team, administer epinephrine 0.3 mg IM (1:1000) (Correct answer)
Correct answer: Stop contrast injection, call code team, administer epinephrine 0.3 mg IM (1:1000)
Anaphylaxis requires immediate cessation of injection, emergency team activation, and epinephrine 0.3 mg IM (1:1000 concentration) in the lateral thigh as first-line treatment.
Question 196: Which ligament of the knee is best seen on sagittal MRI images and connects the posterior femur to the anterior tibia?
- Posterior cruciate ligament
- Anterior cruciate ligament (Correct answer)
- Medial collateral ligament
- Arcuate ligament
Correct answer: Anterior cruciate ligament
The anterior cruciate ligament (ACL) runs from the posterior lateral femoral condyle to the anterior tibial spine and is best evaluated on sagittal MRI.
Question 197: On brain MRI, the hippocampus is best evaluated for atrophy (as in Alzheimer's disease) on which plane?
- Coronal, especially perpendicular to the long axis of the hippocampus (Correct answer)
- Axial only
- Sagittal only
- Oblique axial
Correct answer: Coronal, especially perpendicular to the long axis of the hippocampus
Coronal T1-weighted images perpendicular to the long axis of the hippocampus provide the best visualization of hippocampal volume and atrophy.
Question 198: Which MRI technique uses the difference in signal between in-phase and opposed-phase images to detect intravoxel fat within a lesion?
- Chemical shift imaging (Dixon technique) (Correct answer)
- MR spectroscopy
- Diffusion-weighted imaging
- STIR fat suppression
Correct answer: Chemical shift imaging (Dixon technique)
Chemical shift imaging exploits the precession frequency difference between fat and water; signal drop on opposed-phase images indicates intravoxel fat content.
Question 199: What is the primary consequence of using a high acceleration factor in parallel imaging?
- Reduced SNR due to the g-factor penalty (Correct answer)
- Improved T1 contrast
- Increased spatial resolution
- Reduced scan time with no image quality loss
Correct answer: Reduced SNR due to the g-factor penalty
High acceleration factors in parallel imaging introduce a geometry (g-factor) penalty that reduces SNR, especially in the center of the image where coil sensitivity overlap is less distinct.
Question 200: What is the main purpose of having a patient avoid crossed legs or hands clasped during scanning?
- To improve coil sensitivity
- To reduce motion artifact
- To prevent conductive loops that can cause RF burns (Correct answer)
- To lower acoustic noise
Correct answer: To prevent conductive loops that can cause RF burns
Skin-to-skin contact can form a closed conductive loop in which RF energy induces current, causing burns.
Question 201: Fat and water are imaged with a two-point Dixon technique at 1.5 T. Which pair of TEs is most appropriate for opposed-phase and in-phase images?
- 10 ms and 20 ms
- 4.4 ms and 8.8 ms
- 1.1 ms and 2.2 ms
- 2.2 ms and 4.4 ms (Correct answer)
Correct answer: 2.2 ms and 4.4 ms
At 1.5 T, fat and water are about 220 Hz apart, so they are opposed-phase near 2.2 ms and in-phase near 4.4 ms.
Question 202: Which inversion recovery technique is used to suppress fat signal in musculoskeletal MRI?
- STIR (Short TI Inversion Recovery) (Correct answer)
- MPRAGE
- FLAIR
- VIBE
Correct answer: STIR (Short TI Inversion Recovery)
STIR uses a short inversion time timed to null the longitudinal magnetization of fat, suppressing fat signal uniformly regardless of field inhomogeneity.
Question 203: A technologist increases slice thickness from 3 mm to 6 mm. What is the expected change in image quality?
- SNR decreases and spatial resolution improves
- Both SNR and spatial resolution improve
- No change occurs because FOV is unchanged
- SNR improves but through-plane resolution decreases (Correct answer)
Correct answer: SNR improves but through-plane resolution decreases
Thicker slices contain more protons and generate more signal, improving SNR, but fine anatomical detail in the through-plane direction is reduced.
Question 204: Which of the following is a documented risk of MRI during the first trimester of pregnancy?
- Acoustic noise damage to fetal hearing
- Ionizing radiation exposure
- All of the above except ionizing radiation (Correct answer)
- Gadolinium retention in fetal tissue
Correct answer: All of the above except ionizing radiation
MRI does not use ionizing radiation, but first-trimester risks include acoustic noise and potential gadolinium retention; non-contrast scanning is preferred.
Question 205: A technologist wants to reduce chemical shift artifact at the fat-water interface. Which parameter change is most appropriate?
- Increase the FOV
- Decrease the number of slices
- Increase the receiver bandwidth (Correct answer)
- Decrease the receiver bandwidth
Correct answer: Increase the receiver bandwidth
Increasing receiver bandwidth reduces the frequency difference between fat and water in pixel units, shrinking the spatial shift of fat signal relative to water signal.
Question 206: In a fast spin echo sequence, the effective TE is determined by which factor?
- The echo acquired when the central lines of k-space are filled (Correct answer)
- The echo spacing multiplied by the echo train length
- The time of the first echo in the echo train
- The average TE of all echoes in the train
Correct answer: The echo acquired when the central lines of k-space are filled
Image contrast comes mostly from the center of k-space, so the echo that fills the central lines sets the effective TE.
Question 207: Which ARRT resource provides the official list of examination content categories, subcategories, and approximate percentage weighting for the MRI registry exam?
- ASRT curriculum guide
- ACR MRI practice parameters
- JRCERT program outcomes report
- ARRT Content Specifications document (Correct answer)
Correct answer: ARRT Content Specifications document
The ARRT Content Specifications document is the authoritative blueprint that outlines all tested domains and their weightings for the MRI registry exam.
Question 208: The 'ABER' position used in shoulder MRI arthrography stands for:
- Abduction and Endorotation
- Abduction and Elbow Rest
- Abduction and External Rotation (Correct answer)
- Adduction and External Rotation
Correct answer: Abduction and External Rotation
ABER (Abduction and External Rotation) positions the arm to tension the inferior glenohumeral ligament and improve detection of partial-thickness undersurface rotator cuff tears.
Question 209: During a magnet quench, what is the main danger to people in the scan room if the quench pipe fails?
- Sudden increase in magnetic field strength
- Ionizing radiation exposure
- Asphyxiation and cryogenic burns from helium gas (Correct answer)
- Electrical shock from the gradient coils
Correct answer: Asphyxiation and cryogenic burns from helium gas
Boiling helium can displace oxygen and cause frostbite, so the room should be evacuated at once.
Question 210: Which coil configuration is used to image the entire spine in a single acquisition with uniform SNR?
- Single surface coil placed at the lumbar spine
- Phased-array spine coil spanning cervical, thoracic, and lumbar regions (Correct answer)
- Head and neck coil alone
- Body transmit/receive coil only
Correct answer: Phased-array spine coil spanning cervical, thoracic, and lumbar regions
A multi-element phased-array spine coil allows large FOV imaging with high SNR throughout the cervical, thoracic, and lumbar spine in a single or stitched acquisition.
Question 211: T2* relaxation differs from T2 relaxation primarily because T2* includes:
- Magnetic field inhomogeneities in addition to spin-spin interactions (Correct answer)
- Chemical shift effects only
- Gradient-induced dephasing only
- Only spin-lattice interactions
Correct answer: Magnetic field inhomogeneities in addition to spin-spin interactions
T2* accounts for both intrinsic spin-spin relaxation (T2) and additional dephasing caused by local magnetic field inhomogeneities.
Question 212: The quantity of determines the receiver bandwidth.
- Frequency samples in the matrix (Correct answer)
- Phase changes within the matrix
Correct answer: Frequency samples in the matrix
The receiver bandwidth in MRI is directly determined by the number of frequency samples (pixels) in the frequency encoding direction. A higher number of samples requires a wider bandwidth to accurately capture the range of frequencies present in the signal within the acquisition time. This ensures proper spatial resolution and minimizes aliasing artifacts in the resulting image.
Question 213: Which MRI protocol modification is used to evaluate the acetabular labrum of the hip with maximum sensitivity for tears?
- Diffusion-weighted imaging of the hip
- MR arthrography with intra-articular gadolinium dilution (Correct answer)
- Standard T2 FSE without contrast
- T1 fat-saturated without contrast
Correct answer: MR arthrography with intra-articular gadolinium dilution
MR arthrography with intra-articular dilute gadolinium distends the joint capsule, separating the labrum from the femoral head and maximizing sensitivity for labral tears.
Question 214: What is the main biological risk associated with rapidly switching gradient magnetic fields during an MRI scan?
- Acoustic damage
- Radiofrequency burns
- Peripheral nerve stimulation (Correct answer)
- Hypothermia
Correct answer: Peripheral nerve stimulation
Rapidly switching gradients can induce electrical currents in body tissue, causing peripheral nerve stimulation (PNS).
Question 215: A patient with a metallic foreign body in the eye should be evaluated with which imaging modality BEFORE entering the MRI suite?
- PET scan
- Bone scan
- CT scan (Correct answer)
- Ultrasound
Correct answer: CT scan
A CT scan of the orbits is used to detect metallic ocular foreign bodies before MRI because ferromagnetic metal near the eye is a contraindication.
Question 216: What is the recommended action when a patient with an unknown metallic implant cannot be identified before a clinically urgent MRI?
- Perform the scan with a 0.5T open system only
- Consult the radiologist and ordering physician to assess risk vs. benefit (Correct answer)
- Proceed with the scan at lowest SAR
- Cancel the exam without exception
Correct answer: Consult the radiologist and ordering physician to assess risk vs. benefit
The radiologist and referring physician must jointly assess whether clinical urgency outweighs the unquantified risk of scanning with an unknown implant.
Question 217: When is a candidate's ARRT MRI registry exam score report typically available?
- Immediately on screen at the testing center
- After 30 days by postal mail only
- Within 3–5 business days via the ARRT website (Correct answer)
- Six weeks after the exam window closes
Correct answer: Within 3–5 business days via the ARRT website
ARRT posts official score reports to the candidate's online account within approximately 3–5 business days after testing.
Question 218: What is the FDA guideline limit for peak unweighted sound pressure level in MRI?
- 120 dB
- 85 dB
- 99 dB
- 140 dB (Correct answer)
Correct answer: 140 dB
FDA guidance sets the peak unweighted sound pressure limit at 140 dB, and hearing protection is required for all patients.
Question 219: In a spin echo (SE) sequence, increasing the repetition time (TR) while keeping TE short primarily results in:
- Decreased T1 weighting (more proton density/T2-like contrast) (Correct answer)
- Increased T2 weighting
- Increased SNR only with no contrast change
- Increased T1 weighting
Correct answer: Decreased T1 weighting (more proton density/T2-like contrast)
A long TR allows full T1 recovery of all tissues, reducing T1 contrast differences; combined with short TE, this produces proton density weighting.
Question 220: For MRI of the internal auditory canals (IACs), which imaging plane is most commonly used as the primary plane?
- Axial (Correct answer)
- Sagittal oblique
- Oblique coronal
- Coronal
Correct answer: Axial
Axial high-resolution images are the primary plane for IAC MRI because they show the full length of CN VII and CN VIII within the canal.
Question 221: An MR Conditional implant's labeling specifies a B1+rms limit. What does this parameter describe?
- The static field strength allowed
- The gradient slew rate
- The maximum spatial field gradient
- The time-averaged RF transmit field magnitude (Correct answer)
Correct answer: The time-averaged RF transmit field magnitude
B1+rms is the root-mean-square RF transmit field, a more accurate measure of RF exposure for implants than SAR.
Question 222: Which MRI sequence is used to visualize flowing blood as bright signal ('bright blood') without contrast injection by exploiting the inflow of unsaturated spins into the imaging slice?
- Time-of-flight (TOF) MRA (Correct answer)
- Contrast-enhanced MRA
- VIBE
- Phase contrast MRA
Correct answer: Time-of-flight (TOF) MRA
TOF MRA uses gradient echo with short TR to saturate stationary tissue; fresh unsaturated blood flowing into the slice appears bright due to the inflow enhancement effect.
Question 223: For pituitary gland MRI, which plane is considered the primary imaging plane?
- Coronal (Correct answer)
- Sagittal
- Axial
- Oblique axial
Correct answer: Coronal
Coronal thin-slice images are the primary plane for pituitary imaging because they best demonstrate lateral extension, cavernous sinus invasion, and microadenomas.
Question 224: Which MRI technique measures cerebral blood flow and volume without contrast injection, using arterial water protons as an endogenous tracer?
- Dynamic susceptibility contrast (DSC) MRI
- Arterial spin labeling (ASL) (Correct answer)
- Phase contrast MRI
- BOLD fMRI
Correct answer: Arterial spin labeling (ASL)
ASL uses RF pulses to magnetically label (invert) arterial blood water protons proximal to the imaging slab; labeled images are subtracted from control images to quantify cerebral perfusion.
Question 225: A technologist adds a -90° 'flip-back' pulse at the end of each echo train in a fast spin echo sequence. What is the main benefit?
- It suppresses the fat signal more uniformly
- It removes chemical shift artifact in the frequency direction
- It returns residual transverse magnetization to the longitudinal axis, allowing bright fluid with a shorter TR (Correct answer)
- It eliminates the need for phase encoding
Correct answer: It returns residual transverse magnetization to the longitudinal axis, allowing bright fluid with a shorter TR
Driven equilibrium (DRIVE/FRFSE/RESTORE) flips leftover transverse magnetization back to +Z so fluid stays bright on T2 images with a reduced TR.
Question 226: How does increasing the echo train length (ETL) in a fast spin echo (FSE) sequence affect image quality?
- T1 contrast increases significantly
- Scan time decreases but T2 blurring may increase (Correct answer)
- SNR improves proportional to ETL
- Spatial resolution improves in all directions
Correct answer: Scan time decreases but T2 blurring may increase
Longer ETL fills k-space faster (reducing scan time) but each echo has progressively more T2 decay, causing blurring particularly of edges and fine detail.
Question 227: Which gradient-echo sequence uses a very short TR and a spoiler gradient to eliminate residual transverse magnetization before the next RF pulse?
- Spoiled gradient echo (SPGR/FLASH) (Correct answer)
- GRASS
- Steady-state free precession (SSFP/TrueFISP)
- FIESTA
Correct answer: Spoiled gradient echo (SPGR/FLASH)
Spoiled GRE sequences (SPGR, FLASH) use RF or gradient spoiling to destroy residual transverse magnetization between pulses, producing T1-weighted images with high SNR efficiency.
Question 228: Gibbs (truncation) artifact is caused by:
- Field inhomogeneity near air-tissue interfaces
- Undersampling of k-space — insufficient high-frequency data to represent sharp tissue interfaces (Correct answer)
- Aliasing from a FOV smaller than the patient
- Patient motion during scanning
Correct answer: Undersampling of k-space — insufficient high-frequency data to represent sharp tissue interfaces
Gibbs artifact arises from the Fourier series representation of sharp edges when high spatial-frequency k-space data is truncated, producing ringing lines parallel to the interface.
Question 229: Fast spin-echo (FSE/TSE) sequences acquire multiple lines of k-space per TR using which approach?
- Multiple 90° pulses with a single 180° refocusing pulse
- Rapidly toggling the frequency-encoding gradient
- A single 90° pulse followed by multiple 180° refocusing pulses (echo train) (Correct answer)
- Using a series of partial flip-angle pulses
Correct answer: A single 90° pulse followed by multiple 180° refocusing pulses (echo train)
FSE uses an echo train (RARE principle): after a 90° excitation, a series of 180° pulses generates multiple spin echoes, each phase-encoded differently, filling multiple k-space lines per TR.
Question 230: Which k-space trajectory fills the center of k-space first, providing low-spatial-frequency (contrast) information early in the acquisition and reducing motion sensitivity?
- Linear (sequential) k-space filling
- Centric (elliptic centric) k-space ordering (Correct answer)
- EPI single-shot trajectory
- Reverse linear ordering
Correct answer: Centric (elliptic centric) k-space ordering
Centric k-space ordering acquires the low-frequency center of k-space at the beginning of the scan, capturing image contrast during peak enhancement and minimizing motion effects on image quality.
ARRT MRI Exam
The ARRT Magnetic Resonance Imaging examination certifies technologists in MRI, assessing competency in patient care, MRI safety, image production physics and sequences, and clinical imaging procedures.
Exam Rules
- You can skip questions and return to them later
- Flag questions for review before submitting
- No feedback shown until you submit the entire exam
- Unanswered questions count as wrong — answer everything
- 10 pretest questions are mixed in and don't affect your score
- Timer auto-submits when time runs out
- Your progress is auto-saved every 30 seconds
Practice questions written with AI assistance and published by Nathan Beja, founder of Practice Test Geeks. They are not real exam questions. Exam facts are checked against the official sources linked below. Editorial policy
Sources & references