← All IHS Flashcard Decks

Troubleshooting Hearing Instruments Flashcards

7 cards from real IHS practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 7 Troubleshooting Hearing Instruments flashcards as text
  1. A patient reports that voices sound natural but background noise is excessively amplified after a fitting adjustment. Which parameter should the clinician review?

    Answer: Directional microphone settings or noise reduction algorithm strength

    Directional microphone mode and noise reduction algorithms directly control the balance between speech and background noise amplification.

  2. After cleaning an ITE hearing aid with a brush and wax loop, the patient still reports no sound. What is the next troubleshooting step?

    Answer: Replace the wax guard and retest

    If brushing and wax loop cleaning did not restore function, the wax guard itself is likely blocked and needs replacement before further electrical troubleshooting.

  3. A patient whose hearing aid previously worked well now reports it sounds distorted specifically during phone calls. In-person speech is clear. What is the likely cause?

    Answer: Telecoil or Bluetooth streaming pathway issue, not the main microphone or receiver

    If distortion occurs only during phone use (telecoil or Bluetooth) but not with the microphone, the fault lies within the telecoil, wireless module, or streaming pathway.

  4. A dispensing professional uses a hearing aid test box (coupler measurements) and finds output is 15 dB below the manufacturer's specification. The battery is new. What is most likely?

    Answer: Internal receiver or amplifier failure requiring manufacturer repair

    A 15 dB output deficit compared to manufacturer specifications in a calibrated test box indicates an internal component failure—most commonly the receiver or amplifier circuit.

  5. A hearing aid with feedback management activated still produces constant whistling when properly inserted. What is the recommended next step?

    Answer: Have the patient return for a new earmold impression to improve acoustic seal

    Persistent feedback despite active feedback management usually means the physical acoustic seal is inadequate; a new impression and earmold will correct the venting leak.

  6. A patient reports their rechargeable hearing aid's battery life has decreased from 18 hours to 6 hours over several months. The charging contacts are clean. What is most likely?

    Answer: Battery cell degradation from charge cycles reducing capacity

    Rechargeable lithium-ion cells degrade over hundreds of charge cycles, permanently losing capacity and resulting in shorter daily runtime.

  7. Which of the following is the correct sequence for troubleshooting a hearing aid reported as 'completely dead'?

    Answer: Check battery → inspect wax guard/tubing → listen with stethoscope → send for repair if needed

    The standard troubleshooting hierarchy starts with the most common, reversible causes (battery, wax guard, tubing) before committing to repair or replacement.

How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. Exam facts are checked against the official sources listed below. Not reviewed by a subject-matter expert. Editorial policy

Sources & references

Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy

  1. ihsinfo.org — Time limit, test delivery, eligibility, fees, retake policy · verified Oct 2026
  2. myhome.ihsinfo.org — Passing score, retake fee · verified Oct 2026