  1. A patient reports their BTE hearing aid produces a loud, continuous squeal only when their hand is placed near the ear. What is the most likely cause?

    Answer: Acoustic feedback due to a loose earmold fit

    A loose or improperly fitted earmold allows amplified sound to escape and re-enter the microphone, causing feedback squealing.

  2. A patient's ITE hearing aid produces intermittent sound that cuts in and out. After ruling out battery and wax blockage, what is the next most likely cause?

    Answer: Defective volume control or loose internal connection

    Intermittent sound after ruling out battery and wax typically indicates a faulty potentiometer, cracked solder joint, or loose internal connection.

  3. Which tool is used to verify that a hearing aid's output matches its prescribed target in a clinical troubleshooting session?

    Answer: Real-ear measurement (REM) system

    Real-ear measurement systems use a probe microphone to verify hearing aid output in the ear canal against prescriptive targets.

  4. A receiver-in-canal (RIC) hearing aid suddenly sounds very weak. The battery is new and the dome is clear of wax. What should be checked next?

    Answer: Inspect and replace the wax guard on the receiver

    A clogged wax guard on the RIC receiver is a very common cause of sudden reduced output even when the battery is fresh and the dome appears clear.

  5. A patient complains that their hearing aid sounds 'tinny' or too sharp after a programming adjustment. Which parameter was most likely increased excessively?

    Answer: High-frequency gain

    Excessive high-frequency gain produces a tinny or harsh sound quality, as the treble components are over-amplified relative to the bass.

  6. During troubleshooting, you notice corrosion on the battery contacts of a hearing aid. What is the recommended corrective action?

    Answer: Clean contacts with a dry cotton swab or special contact cleaner, then test

    Corroded battery contacts should be cleaned with a dry swab or approved contact cleaner to restore proper electrical connection before replacing the device.

  7. A patient's hearing aid works normally on a listening stethoscope but the patient still reports no sound in their ear. What is the most likely problem?

    Answer: The earmold or dome tubing is blocked or disconnected

    If the aid functions on a stethoscope, the aid itself is working, so the issue lies in the acoustic pathway to the ear—typically a blocked or disconnected earmold/tubing.

How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. Exam facts are checked against the official sources listed below. Not reviewed by a subject-matter expert. Editorial policy

Sources & references

Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy

  1. ihsinfo.org — Time limit, test delivery, eligibility, fees, retake policy · verified Oct 2026
  2. myhome.ihsinfo.org — Passing score, retake fee · verified Oct 2026