  1. Which real-ear measurement (REM) target is most commonly used when fitting hearing aids to adults using evidence-based best practices?

    Answer: NAL-NL2

    NAL-NL2 (National Acoustic Laboratories Non-Linear version 2) is a widely adopted prescriptive target for fitting hearing aids in adults.

  2. When performing a real-ear aided response (REAR) measurement, the probe microphone is placed:

    Answer: Approximately 5–6 mm from the eardrum

    The probe tube is inserted to approximately 5–6 mm from the tympanic membrane to accurately capture the sound at the eardrum.

  3. A patient reports that soft speech sounds are inaudible even at maximum volume. The MOST likely fitting issue is:

    Answer: Insufficient gain for soft inputs

    Insufficient gain for soft-level inputs means quiet sounds are not amplified enough to reach audibility.

  4. The DSL m[i/o] prescriptive method was specifically developed for:

    Answer: Pediatric populations and infants

    DSL m[i/o] (Desired Sensation Level) was designed to optimize audibility of speech for children, including infants.

  5. Which term describes the maximum output of a hearing aid regardless of input level?

    Answer: OSPL90

    OSPL90 (Output Sound Pressure Level at 90 dB input) is the standardized measure of a hearing aid's maximum output.

  6. A first-time hearing aid user complains that their own voice sounds hollow or like they are talking in a barrel. This is most likely caused by:

    Answer: The occlusion effect

    The occlusion effect occurs when a closed earmold blocks the ear canal, amplifying bone-conducted low-frequency energy of one's own voice.

  7. To reduce the occlusion effect without sacrificing low-frequency amplification for a patient with a sloping high-frequency loss, the best earmold modification is:

    Answer: Increase vent diameter

    Increasing the vent diameter allows low-frequency bone-conducted sound to escape, reducing the hollow sensation without affecting amplification for patients who do not need low-frequency gain.

How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. Exam facts are checked against the official sources listed below. Not reviewed by a subject-matter expert. Editorial policy

Sources & references

Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy

  1. ihsinfo.org — Time limit, test delivery, eligibility, fees, retake policy · verified Oct 2026
  2. myhome.ihsinfo.org — Passing score, retake fee · verified Oct 2026