โ† All Histology Certification Flashcard Decks

Quality Control and Quality Assurance Flashcards

7 cards from real Histology Certification practice questions. Tap to flip, then mark Knew It or Still Learning โ€” missed cards come back until you master them.

Read the first 7 Quality Control and Quality Assurance flashcards as text
  1. When selecting tissue for use as an IHC positive control, the most important criterion is that the tissue:

    Answer: Must contain a known, confirmed concentration of the target antigen

    The positive control tissue must contain a verified, known quantity of target antigen to reliably confirm that the antibody is detecting the correct antigen at the expected intensity.

  2. In H&E quality control assessment, 'eosin intensity' is evaluated by examining:

    Answer: The pink-to-red coloration of cytoplasm and extracellular matrix components

    Eosin is an acidophilic dye staining cytoplasm, collagen, and extracellular matrix pink to red; evaluating this coloration is a standard H&E QC parameter for stain quality.

  3. Which instrument must be regularly calibrated and verified in histology QC to ensure consistent tissue section thickness?

    Answer: Microtome used for cutting paraffin-embedded tissue

    The microtome must be regularly calibrated to ensure it cuts at the correct and consistent thickness, as section thickness directly affects staining quality and diagnostic accuracy.

  4. A 'delta check' in laboratory quality control is used to:

    Answer: Compare a patient's current result with their previous result to detect unexpected discrepancies

    A delta check flags unexpected changes between a patient's current and prior results, helping identify pre-analytical errors such as specimen mix-ups or mislabeling.

  5. According to best QC practices, internal control slides in a histology laboratory should be stained:

    Answer: With each staining run or batch of patient slides processed

    QC slides must accompany every staining run so that any run-specific failure is detected before patient results are reported, preventing diagnostic errors.

  6. What is the purpose of 'method validation' when a new staining procedure is introduced in a histology laboratory?

    Answer: To confirm the procedure performs as expected and meets defined criteria before use on patient specimens

    Method validation establishes that a new procedure reliably produces accurate, reproducible results within acceptable performance limits before it is used for patient diagnosis.

  7. When an 'out-of-control' QC result is identified in histology, the laboratory must:

    Answer: Hold or repeat affected patient results until the problem is identified and corrected

    Patient results from an out-of-control run cannot be reported until root cause is identified and corrective action verified, as the results may be unreliable and could harm patient care.

Quality Control and Quality Assurance Flashcards โ€” Histology Certification Study Cards with Answers