← All GXMO Flashcard Decks

GXMO Quality Control and Maintenance Flashcards

6 cards from real GXMO practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 6 GXMO Quality Control and Maintenance flashcards as text
  1. Which quality control test is used to verify that the light field of the collimator accurately corresponds to the X-ray field?

    Answer: Beam alignment (light/radiation field congruence) test

    The beam alignment test verifies that the visible light field from the collimator matches the actual radiation field within accepted tolerances (typically ±2% of SID).

  2. The half-value layer (HVL) of an X-ray beam is a measure of:

    Answer: Beam quality (penetrating ability)

    HVL is the thickness of a specified material (usually aluminum) required to reduce beam intensity by 50%, indicating the beam's penetrating quality.

  3. A sensitometer is used in radiographic quality control to:

    Answer: Expose film/CR plates to a standardized series of light steps for processing evaluation

    A sensitometer exposes imaging media to controlled light increments, and the resulting density steps are measured with a densitometer to assess processor consistency.

  4. Acceptable tolerance for kVp accuracy in a quality control program is generally:

    Answer: ±5% of the indicated kVp

    Quality control standards (NCRP, ACR) generally accept kVp accuracy within ±5% of the set value to ensure consistent image contrast and patient dose.

  5. The purpose of collimator shutters in an X-ray unit is to:

    Answer: Restrict the primary beam to the area of clinical interest

    Collimator shutters are adjustable lead blades that restrict the X-ray beam to the area being examined, reducing patient dose and scatter.

  6. Which QC test evaluates whether the X-ray tube's focal spot has enlarged beyond acceptable limits?

    Answer: Star resolution pattern or pinhole camera test

    The star resolution pattern or pinhole camera test measures actual focal spot size; an enlarged focal spot reduces spatial resolution.

How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. Exam facts are checked against the official sources listed below. Not reviewed by a subject-matter expert. Editorial policy

Sources & references

Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy

  1. codes.ohio.gov — Score validity · verified Oct 2026
  2. hdmaster.com — Test delivery, eligibility, fees, exam format, passing score, registration deadline, retake fee, retake policy, score release, time limit · verified Oct 2026