GXMO Quality Control and Maintenance Flashcards
6 cards from real GXMO practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 6 GXMO Quality Control and Maintenance flashcards as text
Automatic exposure control (AEC) devices should be tested regularly. A common QC test for AEC verifies:
Answer: Consistency of mAs output across different kVp settings
AEC QC tests confirm that the system delivers consistent mAs (and therefore consistent receptor exposure) across a range of kVp values and patient thicknesses.
Artifacts caused by dirt or debris on a CR (computed radiography) imaging plate most commonly appear as:
Answer: White (unexposed) or black dots and lines on the image
Foreign material on a CR plate blocks or scatters the laser during readout, creating white (or occasionally dark) artifacts on the processed image.
The base-plus-fog (B+F) level of radiographic film processed without exposure should be:
Answer: 0.18 or less
A B+F density of 0.18 or less indicates that the film base and chemical fog are within acceptable limits, ensuring adequate image contrast.
Which component of a digital radiography system converts X-ray photons directly into an electrical signal without an intermediate step?
Answer: Direct DR flat panel detector (amorphous selenium)
Direct DR detectors use amorphous selenium to convert X-rays directly to electrical charge, while indirect detectors use a scintillator to first convert X-rays to light.
Repeat analysis in a quality control program is performed to:
Answer: Identify common causes of repeat exposures and reduce patient dose
Repeat analysis tracks the frequency and reasons for repeated exposures, allowing targeted corrective action to reduce unnecessary patient radiation dose.
The speed of an X-ray film-screen system refers to:
Answer: Its sensitivity to radiation (ability to produce diagnostic density with less exposure)
Screen-film speed (e.g., 100, 200, 400) indicates the system's sensitivity; higher speed requires less radiation to achieve the same image density, reducing patient dose.
How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. Exam facts are checked against the official sources listed below. Not reviewed by a subject-matter expert. Editorial policy
Sources & references
Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy
- codes.ohio.gov — Score validity · verified Oct 2026
- hdmaster.com — Test delivery, eligibility, fees, exam format, passing score, registration deadline, retake fee, retake policy, score release, time limit · verified Oct 2026