Electrode Placement and Montages Flashcards

7 cards from real EEG practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 7 Electrode Placement and Montages flashcards as text
  1. In the 10-20 system, which electrode is located at the vertex of the skull?

    Answer: Cz

    Cz is located at the vertex (top) of the skull at the intersection of the midline and a line connecting the two ears.

  2. What does a referential montage compare each electrode to?

    Answer: A common reference electrode

    In a referential montage, each active electrode is compared to a single common reference electrode, often placed at the ear or linked mastoids.

  3. Which electrode placement best captures activity from the temporal lobe?

    Answer: T3/T4 (T7/T8)

    T3/T4 (now called T7/T8 in the revised system) are positioned over the temporal lobes and best capture temporal lobe activity.

  4. What is the purpose of the ground electrode in an EEG recording?

    Answer: To reduce common-mode noise and electrical interference

    The ground electrode helps the amplifier reject common-mode noise and electrical interference by providing a reference point for the differential amplifier.

  5. In a bipolar longitudinal montage, what is the typical chain order for the left hemisphere?

    Answer: Fp1-F3-C3-P3-O1

    The standard longitudinal bipolar (double banana) chain runs from front to back: Fp1-F3-C3-P3-O1 for the left parasagittal chain.

  6. Which statement about electrode impedance is correct for clinical EEG?

    Answer: Impedance should be below 5 kΩ for optimal recording

    Electrode impedance should be below 5 kΩ (ideally 1-5 kΩ) to ensure adequate signal quality and minimize noise artifacts.

  7. What does the '10' and '20' refer to in the International 10-20 system?

    Answer: Percentage of total skull distance between electrode sites

    The 10 and 20 refer to the percentage of the total front-to-back or left-to-right skull distance used as the spacing between electrode positions.