Artifact Identification Flashcards
7 cards from real EEG practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 7 Artifact Identification flashcards as text
A rhythmic artifact at exactly 60 Hz appearing uniformly across all electrodes is most consistent with:
Answer: AC line noise
60 Hz AC line noise appears as a uniform, rhythmic interference at power line frequency across all or many electrodes.
Which artifact would most likely appear as a high-amplitude, low-frequency deflection time-locked to the R-wave on ECG?
Answer: Pulse artifact
Pulse artifact is caused by scalp vessel pulsations under electrodes and appears as slow deflections synchronous with the cardiac cycle.
A technologist notices intermittent square-wave transients isolated to a single electrode that disappear when the electrode impedance is rechecked and re-gelled. This is most consistent with:
Answer: Electrode pop
Electrode pops are caused by intermittent loss of contact or impedance instability at a single electrode and typically resolve with re-application.
Slow, sinusoidal, high-amplitude waves that wax and wane over frontal electrodes in a drowsy, diaphoretic patient are most characteristic of:
Answer: Sweat artifact
Sweat artifact produces slow (< 1 Hz), large-amplitude sinusoidal waves predominantly over frontal regions due to sweat gland activity.
An artifact appearing as a rhythmic, 1–3 Hz broad, high-amplitude slow-wave pattern that correlates with the patient's assisted ventilation rate is called:
Answer: Ventilator artifact
Ventilator artifact is caused by rhythmic chest and head movement from mechanical ventilation, producing slow waves at the ventilation rate.
In a referential montage, an artifact that appears identical in amplitude and polarity on ALL channels simultaneously most likely originates from:
Answer: The reference electrode
When artifact appears on all channels equally in a referential montage, it is localized to the common reference electrode, not the active electrodes.
High-frequency, irregular bursts of activity seen maximally over temporal and frontal electrodes during a patient interview are most consistent with:
Answer: Chewing/jaw muscle artifact
Chewing produces high-frequency myogenic artifact maximal over temporal regions due to masseter and temporalis muscle proximity.