EDT Quality Assurance & Documentation Flashcards
6 cards from real EDT practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 6 EDT Quality Assurance & Documentation flashcards as text
An EDX laboratory quality assurance program should include:
Answer: Establishing normal reference values for the lab's specific patient population, equipment, and technique
Each EDX lab should establish its own normal values because reference ranges vary with patient demographics, equipment type, electrode placement techniques, and lab temperature protocols.
In the EMG report, the section describing the clinical indication and relevant history is important because:
Answer: It provides context that is essential for accurately interpreting the EDX findings in relation to the clinical question
Clinical context allows the interpreting physician to correlate EDX abnormalities with the clinical presentation, ensuring findings are interpreted in the appropriate clinical framework.
Infection control protocols in the EDX laboratory require that reusable needle electrodes:
Answer: Be sterilized using approved sterilization methods or replaced with single-use disposable needles between patients
Single-use disposable needles are the standard of care in most modern labs; if reusable needles are used, rigorous sterilization per facility policy is mandatory to prevent infection transmission.
When an EDX study is technically limited due to peripheral edema or obesity, the report should:
Answer: Document the technical limitations and their potential impact on the study's reliability
Transparency about technical limitations (edema, obesity, inability to obtain supramaximal stimulation) is essential so the referring clinician can appropriately weight the EDX conclusions.
Needle gauge selection for EMG affects recordings because smaller gauge needles:
Answer: Record from a smaller territory, are less painful, but may have higher impedance
Smaller gauge needles have a smaller recording territory (reduced uptake area), causing less patient discomfort but potentially higher impedance and smaller recorded MUP amplitudes.
The primary purpose of the reference electrode in a monopolar needle EMG recording is:
Answer: To provide an electrically neutral reference point so the differential amplifier records only local muscle activity
The reference electrode provides a neutral potential for the differential amplifier; the active needle minus the reference captures local muscle electrical activity while rejecting common-mode noise.