EDT Quality Assurance & Documentation Flashcards
6 cards from real EDT practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 6 EDT Quality Assurance & Documentation flashcards as text
When a patient's CMAP or SNAP amplitude is unexpectedly low, the first technical factor to assess is:
Answer: Whether the recording electrodes have adequate skin contact and low impedance
High electrode impedance due to poor skin contact is one of the most common technical causes of reduced CMAP or SNAP amplitude and should be checked before concluding pathology.
Stimulus artifact in NCS is best minimized by:
Answer: Ensuring the ground electrode is placed between the stimulator and recording electrodes
Placing the ground electrode between the stimulator and recording electrodes provides a current path to ground that reduces stimulus artifact contaminating the recording.
60 Hz power line interference on the EMG tracing is best eliminated by:
Answer: Optimizing skin preparation, reducing electrode impedance, and ensuring proper grounding
60 Hz interference is primarily caused by high electrode impedance; good skin preparation (abrasion and cleaning) and proper ground electrode placement reduce this artifact.
The recommended stimulation rate for standard nerve conduction studies is:
Answer: 1–3 Hz
Stimulation at 1–3 Hz allows sufficient recovery time between stimuli to prevent rate-dependent changes in CMAP or SNAP amplitude.
Why is supramaximal stimulation required for accurate nerve conduction studies?
Answer: To ensure all axons in the nerve are activated, producing a maximal and reproducible evoked potential
Supramaximal stimulation (typically 20–30% above the level that produces maximal response) ensures all axons are activated, making amplitude and velocity measurements reliable.
Volume conduction artifacts on SNAP recordings can be reduced by:
Answer: Using near-nerve needle electrodes or ring electrodes with standardized technique
Near-nerve needle electrodes or ring electrodes with proper placement minimize volume-conducted signals from adjacent nerves and improve SNAP recording specificity.
How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. Exam facts are checked against the official sources listed below. Not reviewed by a subject-matter expert. Editorial policy
Sources & references
Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy