ECG Troubleshooting Flashcards
7 cards from real ECG practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 7 ECG Troubleshooting flashcards as text
An ECG technician observes a 60-cycle (60 Hz) interference pattern on the tracing. Which corrective measure is most appropriate?
Answer: Check electrode contact and ensure leads are not crossing electrical equipment
60 Hz AC interference is caused by poor electrode contact or proximity to electrical equipment; improving electrode contact and repositioning cables eliminates this artifact.
A patient's ECG shows PR interval of 320 ms that is constant across all beats. How is this classified?
Answer: First-degree AV block
First-degree AV block is defined as a PR interval greater than 200 ms that remains constant with every beat conducted through the AV node.
On an ECG, the QT interval measures 480 ms at a heart rate of 60 bpm. What is the corrected QT (QTc) and its significance?
Answer: QTc = 480 ms; prolonged, risk of Torsades de Pointes
At 60 bpm (RR = 1.0 sec), QTc by Bazett's formula equals QT/√RR = 480/1.0 = 480 ms, which is prolonged and associated with risk of Torsades de Pointes.
A patient with known COPD has an ECG showing right axis deviation, P pulmonale, and dominant R wave in V1. What is the most likely interpretation?
Answer: Right ventricular hypertrophy secondary to cor pulmonale
Chronic COPD causes pulmonary hypertension leading to right ventricular hypertrophy, which produces right axis deviation, tall P waves (P pulmonale), and a dominant R in V1.
Which ECG feature distinguishes Mobitz II second-degree AV block from Mobitz I (Wenckebach)?
Answer: Mobitz II has a fixed PR interval before the dropped beat; Mobitz I has a progressively lengthening PR
Mobitz II features a constant PR interval with sudden non-conducted P waves, whereas Mobitz I shows progressive PR prolongation before a dropped beat.
During ECG acquisition, the technician notices the calibration pulse measures 5 mm instead of the standard 10 mm. What does this indicate?
Answer: The gain is set to half standard (5 mm/mV)
Standard ECG calibration is 10 mm/mV; a 5 mm calibration pulse means the gain is halved (5 mm/mV), which will make all waveforms appear half their true amplitude.
A patient's ECG shows ST elevation in leads II, III, and aVF with reciprocal ST depression in I and aVL. Which coronary artery is most likely occluded?
Answer: Right coronary artery
The inferior leads (II, III, aVF) are supplied primarily by the right coronary artery (RCA) in most patients, so inferior STEMI usually reflects RCA occlusion.
How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. Exam facts are checked against the official sources listed below. Not reviewed by a subject-matter expert. Editorial policy
Sources & references
Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy
- nhanow.com — Test delivery, score release, score validity, eligibility, passing score, renewal, retake policy · verified Oct 2026
- info.nhanow.com — Time limit · verified Oct 2026