Patient Monitoring Techniques Flashcards

6 cards from real DAANCE practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 6 Patient Monitoring Techniques flashcards as text
  1. During a moderate sedation procedure, the capnograph alarm sounds and the digital EtCO2 reading drops to zero, with a corresponding flat waveform. The pulse oximeter, however, still reads 98%. What is the most likely cause of this finding?

    Answer: There is apnea or a disconnection in the sampling line.

    A flatline capnograph waveform indicates that no carbon dioxide is being detected. This occurs most commonly when the patient stops breathing (apnea) or the sampling line has become disconnected from the patient or the monitor. [6, 23] Oxygen saturation, as measured by pulse oximetry, will take time to decrease, making capnography a more immediate indicator of ventilatory changes. [3, 20]

  2. A dental anesthesia assistant is monitoring an obese patient and obtains a blood pressure reading of 175/105 mmHg using a standard adult cuff. Which of the following is the most critical and immediate next step?

    Answer: Re-measure the blood pressure using a large adult or thigh cuff.

    Using a blood pressure cuff that is too small for a patient's arm circumference can cause a falsely high reading. [1, 7, 11] Before concluding that the patient is severely hypertensive, the assistant must ensure an accurate measurement by using the appropriately sized cuff. [32] Studies have shown that a regular cuff on a patient requiring a large or extra-large cuff can overestimate systolic blood pressure by nearly 5 to 20 mmHg respectively. [14]

  3. While observing the ECG monitor during a sedation case, the assistant notes the sudden appearance of a single, wide, and bizarrely-shaped QRS complex that occurs earlier than the next expected beat. There is no P wave preceding it. This finding is best described as a:

    Answer: Premature ventricular contraction (PVC)

    A premature ventricular contraction (PVC) is an ectopic beat originating from the ventricles. Its key ECG characteristics include being premature, having a wide (≥ 0.12s) and abnormal QRS morphology, lacking a preceding P wave, and typically being followed by a compensatory pause. [9, 12, 34]

  4. Which of the following conditions is most likely to cause a pulse oximeter to display a reading that is falsely high or misleadingly normal?

    Answer: Carbon monoxide poisoning

    Carbon monoxide binds to hemoglobin with an affinity much higher than oxygen, forming carboxyhemoglobin. A standard pulse oximeter cannot differentiate between carboxyhemoglobin and oxyhemoglobin, and will therefore report a falsely high or normal oxygen saturation level, which can mask severe underlying hypoxia. Poor perfusion, hypothermia, and dark nail polish are more likely to cause falsely low or no reading. [26, 30]

  5. The primary purpose of using a precordial or pretracheal stethoscope during dental anesthesia and sedation is to provide continuous, real-time auditory monitoring of:

    Answer: Respiratory patterns and heart tones.

    A precordial or pretracheal stethoscope is a simple, non-invasive monitor that allows the anesthesia assistant to continuously listen to the patient's breath and heart sounds. [4, 10] This provides immediate detection of changes in respiratory rate, airway patency (e.g., obstruction, wheezing), and heart rate/rhythm. [18, 24]

  6. When comparing patient monitoring devices, which provides the most immediate, breath-by-breath feedback on the adequacy of a patient's ventilation?

    Answer: Capnograph (End-Tidal CO2)

    Capnography directly measures the concentration of carbon dioxide in a patient's exhaled breath, providing an instantaneous reflection of their ventilatory status. [3, 21] Pulse oximetry measures oxygenation, which is a late sign of ventilatory failure, as a patient can hypoventilate for several minutes before their oxygen saturation begins to drop, especially if receiving supplemental oxygen. [17, 20] ECG and NIBP monitor cardiovascular, not respiratory, function.