  1. What type of water is recommended for the final rinse after high-level disinfection?

    Answer: Sterile or bacteria-filtered (critical) water

    Using sterile or filtered water prevents recontamination with waterborne organisms such as Pseudomonas and nontuberculous mycobacteria.

  2. What is the purpose of flushing channels with 70–90% alcohol followed by forced air after HLD?

    Answer: To promote drying and reduce microbial growth in channels

    Alcohol speeds evaporation and forced air removes moisture that would otherwise support bacterial growth.

  3. How should flexible endoscopes be stored after reprocessing?

    Answer: Hanging vertically in a ventilated cabinet with valves and caps removed

    Vertical hanging in a ventilated or drying cabinet with detachable parts removed allows channels to drain and dry.

  4. Why should the manufacturer's shipping/carrying case NOT be used for routine storage of reprocessed scopes?

    Answer: It can trap moisture and become contaminated, recontaminating the scope

    Carrying cases are difficult to clean and retain moisture, so they are for shipping only.

  5. Which duodenoscope component has been linked to CRE outbreaks because it is difficult to clean?

    Answer: The elevator mechanism at the distal tip

    The complex elevator mechanism can harbor bioburden, so it requires meticulous manual cleaning per IFU.

  6. What is a borescope used for in endoscope reprocessing quality programs?

    Answer: Visually inspecting internal channels for residual debris, moisture, or damage

    A borescope is a small camera passed through channels to reveal problems not visible externally.

  7. What is 'fluid invasion' in a flexible endoscope?

    Answer: Liquid entering the scope's sealed internal components through a leak

    Fluid invasion occurs when a leaking scope is immersed, causing costly internal damage and infection risk.

Written and edited by , founder of Practice Test Geeks, with AI assistance. Exam facts are checked against the official sources linked below. Editorial policy

Sources & references

Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy

  1. myhspa.org — Fees, exam format, passing score, registration deadline, renewal, retake fee, retake policy, test dates, score validity, eligibility · verified Nov 2026
  2. prometric.com — Test delivery · verified Nov 2026