Flexible Endoscope Reprocessing and High-Level Disinfection Flashcards

7 cards from real CRCST practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 7 Flexible Endoscope Reprocessing and High-Level Disinfection flashcards as text
  1. When using an automated endoscope reprocessor (AER), which statement is generally true?

    Answer: Manual cleaning, including brushing, is still required unless the AER is FDA-cleared to replace it

    Most AERs only disinfect, so manual cleaning per the scope IFU must still be done beforehand.

  2. How should channel connectors for an AER be chosen?

    Answer: Model-specific connectors listed by the AER manufacturer for that endoscope

    Incorrect connectors can leave channels unexposed to disinfectant, so compatibility must follow both IFUs.

  3. Why should each endoscope reprocessing cycle be documented with the scope serial number, patient, operator, and AER used?

    Answer: For traceability during infection investigations or recalls

    Complete records link a scope to patients and processes, allowing tracing if an outbreak or failure occurs.

  4. Which gloves provide appropriate protection when handling glutaraldehyde or OPA?

    Answer: Nitrile or butyl rubber gloves

    Nitrile and butyl rubber resist aldehyde penetration better than latex or vinyl.

  5. If manual cleaning cannot begin within the time specified by the endoscope manufacturer, what should be done?

    Answer: Follow the manufacturer's delayed reprocessing protocol, such as an extended soak

    Delayed reprocessing allows soil to dry, so manufacturers specify extra steps before normal cleaning.

  6. What is the correct workflow design for an endoscope reprocessing area?

    Answer: Unidirectional flow from dirty to clean areas

    One-way flow from dirty to clean prevents recontamination of disinfected scopes.

  7. What should be used to manage a spill of glutaraldehyde or OPA?

    Answer: A spill kit with an appropriate neutralizer, following the SDS

    Aldehyde spills require PPE and a neutralizing spill kit per the Safety Data Sheet, not improvised methods.

Written and edited by , founder of Practice Test Geeks, with AI assistance. Exam facts are checked against the official sources linked below. Editorial policy

Sources & references

Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy

  1. myhspa.org — Fees, exam format, passing score, registration deadline, renewal, retake fee, retake policy, test dates, score validity, eligibility · verified Nov 2026
  2. prometric.com — Test delivery · verified Nov 2026