Flexible Endoscope Reprocessing and High-Level Disinfection Flashcards
7 cards from real CRCST practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 7 Flexible Endoscope Reprocessing and High-Level Disinfection flashcards as text
Under the Spaulding classification, a flexible GI endoscope that contacts intact mucous membranes is considered:
Answer: Semicritical
Devices contacting mucous membranes are semicritical and require at least high-level disinfection.
High-level disinfection is defined as a process that destroys:
Answer: All microorganisms except high numbers of bacterial spores
HLD kills all vegetative organisms, mycobacteria, fungi, and viruses but not necessarily large numbers of spores.
Why must thorough cleaning precede high-level disinfection with aldehyde-based disinfectants?
Answer: Aldehydes can fix proteins, making residual soil harder to remove and shielding microbes
Glutaraldehyde and OPA crosslink proteins, so any remaining soil can become fixed and protect microorganisms.
How often should the minimum effective concentration (MEC) of a reusable high-level disinfectant be verified?
Answer: Each day of use or more often as directed by the manufacturer
Reusable HLD dilutes with each cycle, so MEC must be tested with the product-specific strip per IFU, at least daily.
A reusable HLD solution fails its MEC test strip on day 5 of a 14-day reuse life. What is the correct action?
Answer: Discard the solution and replace it, regardless of remaining reuse days
A failed MEC test means the solution is no longer effective and must be discarded even if within its reuse life.
Inadequate rinsing of OPA from cystoscopes has been associated with what patient risk?
Answer: Anaphylaxis-like reactions in bladder cancer patients
Residual OPA on repeatedly used cystoscopes has been linked to anaphylaxis-like reactions, so thorough rinsing is essential.
Which statement about HLD contact time and temperature is correct?
Answer: Contact time and temperature must follow the disinfectant's FDA-cleared label
Efficacy claims are tied to the specific time and temperature on the product's FDA-cleared label.
Written and edited by Nathan Beja, founder of Practice Test Geeks, with AI assistance. Exam facts are checked against the official sources linked below. Editorial policy