CT Image Quality and Reconstruction Flashcards

6 cards from real Computed Tomography Test practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 6 CT Image Quality and Reconstruction flashcards as text
  1. What is the primary advantage of iterative reconstruction over filtered back projection (FBP) in CT?

    Answer: Significant noise reduction at equivalent or lower dose

    Iterative reconstruction repeatedly refines the image estimate against raw data, reducing noise substantially, which allows lower dose acquisition without sacrificing image quality.

  2. Which window level and window width settings are appropriate for evaluating brain parenchyma on CT?

    Answer: Level 35 HU, Width 80 HU

    Brain window settings (level ~35 HU, width ~80 HU) optimize contrast between gray matter, white matter, and CSF within the narrow HU range of brain tissue.

  3. Ring artifacts in CT are most commonly caused by which problem?

    Answer: Defective or mis-calibrated detector elements

    A single malfunctioning detector channel consistently produces incorrect readings at the same projection angle, creating a characteristic ring or arc artifact in the reconstructed image.

  4. What is the relationship between CT slice thickness and the partial volume artifact?

    Answer: Thinner slices reduce partial volume artifact

    Thinner CT slices reduce the volume of each voxel, decreasing the chance that multiple tissue types are averaged together within a single voxel.

  5. What does multiplanar reformation (MPR) allow clinicians to do with CT data?

    Answer: View reconstructed images in coronal, sagittal, and oblique planes from axial data

    MPR uses isotropic or near-isotropic volumetric CT data to generate reconstructions in any desired imaging plane without additional scanning.

  6. The concept of isotropic voxels in modern multi-detector CT (MDCT) means that voxels are:

    Answer: Equal in all three dimensions (x, y, z)

    Isotropic voxels (e.g., 0.5 × 0.5 × 0.5 mm) allow high-quality MPR and 3D reconstruction in any plane without staircase artifacts.