ARRT Computed Tomography (CT) Exam — Questions and Answers
Question 1: Non-contrast CT brain: anticoagulated patient with right hemiparesis and abrupt onset of disorientation. What does the picture depict?
- Calcification of the choroid plexus
- Subdural haemorrhage
- Epidural haemorrhage
- Enhancing mass
- Intraventricular haemorrhage (Correct answer)
Correct answer: Intraventricular haemorrhage
An anticoagulated patient experiencing abrupt neurological deficits like right hemiparesis and disorientation is at high risk for intracranial hemorrhage. On a non-contrast CT, acute blood appears hyperdense (bright white). Intraventricular hemorrhage (IVH) refers to bleeding into the ventricular system of the brain, which would appear as bright, dense material filling or layering within the ventricles on the CT scan.
Question 2: What does the term CTDI stand for in CT radiation dosimetry?
- CT Dose Interval
- CT Diagnostic Imaging
- Computed Tomography Density Index
- Computed Tomography Dose Index (Correct answer)
Correct answer: Computed Tomography Dose Index
CTDI (Computed Tomography Dose Index) is a standardized measure of radiation dose output from a CT scanner for a single rotation.
Question 3: Which CT image reconstruction algorithm produces the sharpest edges but introduces the most noise?
- Soft tissue kernel
- Standard kernel
- Lung kernel
- Bone (sharp) kernel (Correct answer)
Correct answer: Bone (sharp) kernel
Bone or high-frequency kernels enhance edge detail for structures like cortical bone but amplify image noise compared to smooth kernels.
Question 4: Dual-energy CT (DECT) can be used to create virtual non-contrast (VNC) images. What does this eliminate for the patient?
- The need for a separate pre-contrast CT acquisition (Correct answer)
- The need for IV access
- The use of a contrast injector
- Post-processing time
Correct answer: The need for a separate pre-contrast CT acquisition
VNC images derived from DECT data simulate non-contrast images by mathematically subtracting iodine signal, eliminating a dedicated unenhanced acquisition and reducing dose.
Question 5: Which regulatory body in the United States sets occupational radiation dose limits for CT technologists?
- FDA
- ACR
- Nuclear Regulatory Commission (NRC) (Correct answer)
- Joint Commission
Correct answer: Nuclear Regulatory Commission (NRC)
The NRC sets the annual occupational whole-body dose limit of 50 mSv (5 rem) for radiation workers in the US, including CT technologists.
Question 6: A patient experiences mild urticaria after CT contrast injection. What is the most appropriate immediate action?
- Immediately inject epinephrine
- Administer IV corticosteroids and discharge
- Administer diphenhydramine (Benadryl) and observe the patient (Correct answer)
- Continue the scan without intervention
Correct answer: Administer diphenhydramine (Benadryl) and observe the patient
Mild urticaria (hives only, no systemic symptoms) is treated with antihistamines and close observation while monitoring for progression to a more severe reaction.
Question 7: What type of CT imaging is typically used for visualizing blood vessels with high contrast resolution?
- Partial volume effect (Correct answer)
- Cupping artifact
- Beam-hardening artifact
- Streak artifact
Correct answer: Partial volume effect
The partial volume effect occurs when a structure, such as a small blood vessel, is smaller than the CT voxel size, causing its density to be averaged with surrounding tissues. While not a 'type of CT imaging,' this phenomenon directly impacts the perceived contrast and resolution of small structures. It can make blood vessels appear less distinct, challenging their visualization even when contrast agents are used to enhance them. Understanding and minimizing the partial volume effect is crucial for accurate vascular imaging.
Question 8: Which pre-medication regimen is typically used for patients with prior moderate contrast reactions undergoing non-emergent contrast CT?
- Oral prednisone 13 hours, 7 hours, and 1 hour before, plus diphenhydramine 1 hour before (Correct answer)
- IV methylprednisolone 30 minutes before only
- Single dose of epinephrine before the scan
- No premedication is effective
Correct answer: Oral prednisone 13 hours, 7 hours, and 1 hour before, plus diphenhydramine 1 hour before
The standard ACR-recommended steroid premedication protocol uses oral prednisone at three intervals before the scan combined with antihistamine to reduce breakthrough reaction risk.
Question 9: Which term refers to the ability of a CT scanner to acquire images in a very short time interval?
- Axial resolution
- Temporal resolution (Correct answer)
- Contrast resolution
- Spatial resolution
Correct answer: Temporal resolution
Temporal resolution in CT refers to the ability of the scanner to capture images of moving structures with minimal motion blurring, essentially how quickly the data for a single image can be acquired. High temporal resolution is critical for imaging rapidly moving organs like the heart, reducing artifacts caused by cardiac motion or patient breathing. It is often achieved with faster gantry rotation times and multi-detector arrays.
Question 10: Which parameter most directly controls CT image noise?
- mAs (Correct answer)
- Reconstruction interval
- FOV
- kVp
Correct answer: mAs
mAs (milliampere-seconds) determines the quantity of X-ray photons produced; higher mAs means more photons, less quantum noise, and better image quality.
Question 11: What is the standard iodine concentration range (in mg I/mL) for contrast agents commonly used in diagnostic CT?
- 100–150 mg I/mL
- 50–100 mg I/mL
- 500–600 mg I/mL
- 300–370 mg I/mL (Correct answer)
Correct answer: 300–370 mg I/mL
Most CT contrast agents contain 300–370 mg I/mL; higher concentrations improve vascular opacification while lower volumes can be used to reduce total iodine load.
Question 12: Which CT parameter, when increased, directly increases patient radiation dose?
- FOV
- Detector row width
- mAs (Correct answer)
- Pitch (above 1.0)
Correct answer: mAs
Increasing mAs increases the number of X-ray photons produced, directly increasing the radiation dose delivered to the patient.
Question 13: What artifact appears in cardiac CT when the heart rate is irregular or too fast for the gantry rotation time?
- Ring artifact
- Beam hardening from the spine
- Motion artifact causing blurring of cardiac structures (Correct answer)
- Cupping artifact
Correct answer: Motion artifact causing blurring of cardiac structures
Cardiac motion during acquisition causes blurring and misregistration of cardiac structures; ECG gating and fast gantry rotation times mitigate but cannot eliminate this when heart rate is high or irregular.
Question 14: X-rays of the body are captured from various angles during computed tomography (CT), which produces images that resemble 2- and 3-dimensional slices (cross-sections). All jewelry and metal objects, such as zippers and belt buckles close to the body portion being scanned, should be taken off by anyone getting a CT scan. Why are patients asked to remove metal before a CT scan?
- The goods could burn the individual receiving the scan since they heat up in the machine.
- The machine's rays could cause metallic items to release poisonous fumes.
- The items can damage the CT machine
- The items may block the x-rays and blur the images (Correct answer)
Correct answer: The items may block the x-rays and blur the images
Agile Methodology is characterized by its iterative and incremental approach, focusing on flexibility, collaboration, and rapid delivery. It breaks down projects into small, manageable iterations (sprints) to allow for quick progressions and faster releases of working software. This enables teams to adapt to changing requirements and deliver value continuously.
Question 15: On a non-contrast CT brain, a hyperdense (bright) lesion within the basal ganglia most likely represents:
- Acute hemorrhage (Correct answer)
- Metastatic lesion
- Demyelinating plaque
- Subacute infarction
Correct answer: Acute hemorrhage
Acute blood is hyperdense on non-contrast CT due to the high protein content of clotted hemoglobin.
Question 16: What is the typical Hounsfield Unit (HU) value for simple (non-hemorrhagic) cerebrospinal fluid?
- 80–100 HU
- 0 HU (Correct answer)
- −100 HU
- 35–45 HU
Correct answer: 0 HU
Simple CSF measures approximately 0–15 HU, close to the value of water, allowing differentiation from higher-density hemorrhage or proteinaceous fluid.
Question 17: Which component of a CT scanner is responsible for rotating the X-ray tube and detector around the patient?
- Collimator
- Gantry (Correct answer)
- Console
- Generato
Correct answer: Gantry
The gantry is the large, donut-shaped component of a CT scanner that houses the X-ray tube and the detectors. It is responsible for rotating these components around the patient to acquire X-ray data from multiple angles. The patient lies on a table that slides through the opening of the gantry during the scan.
Question 18: BEAM COLLIMATION and DETECTOR COLLIMATION settings determine the section width in the MSCT System.
- Collimation (msct)-detector (section) collimation
- SSCT Beam Collimation Consists of:
- MSCT Beam Collimation Consists of:
- MSCT Beam Collimation Consists of: Post-Patient Collimation (Correct answer)
Correct answer: MSCT Beam Collimation Consists of: Post-Patient Collimation
In Multi-Slice Computed Tomography (MSCT), both pre-patient and post-patient collimation are critical for defining the section width and managing radiation dose. Post-patient collimation, located just before the detectors, further refines the X-ray beam after it has passed through the patient. This ensures that only the desired X-rays reach the detectors, contributing to precise slice thickness and optimal image quality.
Question 19: What is the typical Hounsfield Unit (HU) range for normal brain tissue on a non-contrast CT scan?
- 200 to 300 HU
- 0 to 50 HU
- 100 to 200 HU
- -100 to 100 HU (Correct answer)
Correct answer: -100 to 100 HU
Normal brain tissue on a non-contrast CT scan typically has Hounsfield Unit (HU) values ranging from approximately 20-40 HU for gray matter and slightly lower for white matter. The option of -100 to 100 HU provides a broader range that encompasses these normal values, distinguishing brain parenchyma from much denser structures like bone (hundreds of HU) or less dense structures like CSF (around 0-10 HU) or fat (negative HU).
Question 20: In CT, which body region is most sensitive to radiation-induced cancer risk?
- Skin
- Bone marrow and gonads (Correct answer)
- Brain
- Extremities
Correct answer: Bone marrow and gonads
Bone marrow and gonads have the highest tissue weighting factors (wT), making them most sensitive to stochastic radiation-induced cancer risk.
Question 21: What is the purpose of the bow-tie (wedge) filter in CT?
- To remove high-energy photons from the beam
- To increase radiation dose to the center of the patient
- To compensate for the varying thickness of the patient's body cross-section (Correct answer)
- To reduce motion artifacts
Correct answer: To compensate for the varying thickness of the patient's body cross-section
The bow-tie filter attenuates the beam more at the periphery (thinner patient edges) than at the center, equalizing detector signal and reducing skin dose.
Question 22: What is the purpose of the CT table's indexing capability?
- To synchronize with the patient's breathing
- To adjust the X-ray tube voltage
- To increase the gantry rotation speed
- To move the patient through the gantry during scanning (Correct answer)
Correct answer: To move the patient through the gantry during scanning
The CT table's indexing capability refers to its ability to precisely move the patient incrementally through the gantry opening. This controlled movement allows for the acquisition of sequential slices or a continuous helical scan, ensuring that the entire anatomical region of interest is covered systematically. This precise positioning is critical for accurate imaging and reconstruction.
Question 23: The term 'Z-axis dose profile' in CT refers to dose distribution along which direction?
- Left-to-right axis
- Patient cranio-caudal (head-to-foot) axis (Correct answer)
- Anterior-to-posterior axis
- Rotational axis of the gantry
Correct answer: Patient cranio-caudal (head-to-foot) axis
The Z-axis is the cranio-caudal axis along which the patient moves through the gantry, and the dose profile describes radiation distribution along this direction.
Question 24: A 'free air' (pneumoperitoneum) sign on CT of the abdomen following bowel perforation is best identified on which type of CT window?
- Soft tissue window only
- Bone window
- Mediastinal window
- Lung window settings applied to abdominal images (Correct answer)
Correct answer: Lung window settings applied to abdominal images
Free intraperitoneal air has the same HU as intraluminal air (−1000 HU) and is most conspicuous on lung window settings (wide width, low level) that make small gas pockets visible.
Question 25: What is the term for the phenomenon where a single tissue appears with different densities in different CT images due to varying X-ray attenuation angles?
- Streak artifact
- Cupping artifact
- Partial volume effect (Correct answer)
- Beam-hardening artifact
Correct answer: Partial volume effect
The partial volume effect occurs when a single voxel (the smallest 3D unit of a CT image) contains more than one type of tissue. The CT number assigned to that voxel will be an average of the attenuation values of all the tissues within it. This can lead to misrepresentation of tissue densities and blurring of small structures, especially when the slice thickness is large relative to the size of the object.
Question 26: What is the purpose of the CT bowtie filter?
- To eliminate beam hardening
- To improve image contrast
- To reduce patient dose
- To shape the X-ray beam (Correct answer)
Correct answer: To shape the X-ray beam
A bowtie filter is a specialized filter used in CT scanners that is thicker at the edges and thinner in the center. Its primary purpose is to shape the X-ray beam, making it more intense at the center and less intense at the periphery. This design helps to reduce the dose to the patient's periphery while maintaining a more uniform signal reaching the detectors, which is crucial for image quality.
Question 27: Which CT reconstruction parameter controls the trade-off between image noise and spatial resolution?
- Scan field of view
- Reconstruction kernel (filter) (Correct answer)
- Matrix size
- Gantry rotation speed
Correct answer: Reconstruction kernel (filter)
The reconstruction kernel (convolution filter) determines whether the image is optimized for smoothness (low noise) or edge sharpness (high resolution), with a fundamental trade-off between the two.
Question 28: What is the significance of a 'split-bolus' technique in contrast-enhanced CT?
- It administers oral and IV contrast simultaneously
- It doubles the contrast volume to improve enhancement
- It combines two contrast injection phases into a single acquisition to reduce radiation dose (Correct answer)
- It pre-mixes contrast with saline to reduce osmolality
Correct answer: It combines two contrast injection phases into a single acquisition to reduce radiation dose
The split-bolus technique injects contrast in two separated boluses so that a single scan acquisition captures both nephrographic and excretory phase opacification, reducing the number of scan passes and radiation dose.
Question 29: In CT, the process of reconstructing an image from projection data using mathematical algorithms is called:
- Iterative refinement
- Back-projection
- Fourier synthesis
- Filtered back-projection (Correct answer)
Correct answer: Filtered back-projection
Filtered back-projection (FBP) applies a convolution filter to raw projection data before back-projecting it to form the image.
Question 30: What is the purpose of a CT scout image?
- To measure bone density (Correct answer)
- To estimate patient size
- To detect metallic implants
- To preview the anatomy
Correct answer: To measure bone density
Cardiac CT is a specialized type of CT scan specifically designed to image the coronary arteries and assess heart health. It uses advanced technology to capture detailed images of the heart's anatomy, identify blockages or narrowing in the coronary arteries (coronary artery disease), and evaluate cardiac function. This non-invasive procedure is crucial for diagnosing and managing heart conditions.
Question 31: Which medication must be withheld 48 hours before and after iodinated contrast administration in patients with renal impairment?
- Aspirin
- Atorvastatin
- Metformin (Correct answer)
- Lisinopril
Correct answer: Metformin
Metformin must be withheld because contrast-induced nephropathy can impair its renal clearance, leading to toxic accumulation and risk of lactic acidosis.
Question 32: On CT brain in a patient with suspected cerebral venous sinus thrombosis, the 'empty delta sign' is best seen:
- On non-contrast CT as hypodensity in the temporal lobe
- On CT angiography as MCA occlusion
- On post-contrast CT as a filling defect within the superior sagittal sinus (Correct answer)
- On bone windows as a skull base fracture
Correct answer: On post-contrast CT as a filling defect within the superior sagittal sinus
The empty delta sign appears after contrast as a triangular filling defect (clot) surrounded by enhancing collateral vessels within the superior sagittal sinus.
Question 33: Which brain imaging technique is most sensitive for detecting acute ischemic strokes?
- Magnetic Resonance Imaging (MRI) (Correct answer)
- Computed Tomography (CT)
- Positron Emission Tomography (PET)
- Single-Photon Emission Computed Tomography (SPECT)
Correct answer: Magnetic Resonance Imaging (MRI)
Magnetic Resonance Imaging (MRI), particularly with diffusion-weighted imaging (DWI) sequences, is the most sensitive technique for detecting acute ischemic strokes. DWI can detect changes in water movement within brain tissue that occur minutes after the onset of ischemia, often before any changes are visible on a standard CT scan. This early detection is crucial for timely intervention and treatment to minimize brain damage.
Question 34: Which type of brain herniation is characterized by downward displacement of the cerebellar tonsils through the foramen magnum on CT?
- Tonsillar herniation (Correct answer)
- Transtentorial herniation
- Subfalcine herniation
- Uncal herniation
Correct answer: Tonsillar herniation
Tonsillar herniation occurs when increased posterior fossa pressure drives the cerebellar tonsils through the foramen magnum, compressing the brainstem.
Question 35: Which type of CT scan is specifically designed for imaging the coronary arteries and heart structures?
- Cardiac CT (Correct answer)
- Pelvic CT
- Head CT
- Abdominal CT
Correct answer: Cardiac CT
Cardiac CT, often performed as Coronary CT Angiography (CCTA), is specifically designed for detailed imaging of the coronary arteries and heart structures. This technique uses rapid image acquisition and often involves synchronization with the patient's electrocardiogram (ECG) to minimize motion artifacts from the beating heart. It is crucial for detecting coronary artery disease, assessing stent patency, and evaluating cardiac anatomy.
Question 36: In CT, what is the 'matrix size' most commonly used in clinical scanners?
- 512 x 512 (Correct answer)
- 256 x 256
- 128 x 128
- 1024 x 1024
Correct answer: 512 x 512
Clinical CT scanners most commonly use a 512 x 512 matrix, balancing spatial resolution with acceptable processing time and storage requirements.
Question 37: A patient with a creatinine level of 2.8 mg/dL is scheduled for contrast-enhanced CT. What is the most appropriate next step?
- Consult with the ordering physician regarding contrast risk and alternatives (Correct answer)
- Automatically cancel the exam without notifying the physician
- Proceed with full contrast dose as ordered
- Administer contrast and monitor vitals closely
Correct answer: Consult with the ordering physician regarding contrast risk and alternatives
Elevated creatinine indicates reduced renal function and significant risk of contrast-induced nephropathy; the ordering physician must be consulted to weigh risk versus benefit.
Question 38: Colon cancer patient after brain contrast MRI Which diagnosis is more likely?
- Two enhancing metastases (Correct answer)
- Subarachnoid haemorrhage
- Two areas of normal calcification
- Two primary brain tumors
- Two benign meningiomas
Correct answer: Two enhancing metastases
In a patient with a known history of colon cancer, the presence of new brain lesions that enhance with contrast on an MRI is highly suspicious for metastatic disease. Colon cancer frequently metastasizes to the brain. Enhancing lesions typically indicate a breakdown of the blood-brain barrier, which is common in aggressive tumors like metastases, making this the most likely diagnosis.
Question 39: What kind of detector is used now that is both effective and popular?
- Stationary detectors
- Single detectors
- Solid state crystalline detectors (Correct answer)
- Xenon gas detectors
Correct answer: Solid state crystalline detectors
CT angiography is frequently preferred over traditional angiography because it is a significantly less invasive procedure. Traditional angiography involves inserting a catheter directly into an artery, which carries higher risks such as bleeding, infection, and arterial damage. CT angiography, on the other hand, only requires an intravenous injection of contrast material, making it safer and more comfortable for the patient.
Question 40: Which window setting is most appropriate for evaluating pulmonary nodules?
- Bone window (WW 2000, WL 500)
- Lung window (WW 1500, WL -600) (Correct answer)
- Brain window (WW 80, WL 40)
- Soft tissue window (WW 400, WL 40)
Correct answer: Lung window (WW 1500, WL -600)
The lung window uses a wide window width and low window level to optimize contrast between air-filled lung parenchyma and soft tissue nodules.
Question 41: Which type of contrast medium is often used in CT imaging to enhance visualization of blood vessels and other structures?
- Iron oxide
- Iodine-based (Correct answer)
- Gadolinium
- Barium sulfate
Correct answer: Iodine-based
A Hounsfield Unit (HU) is a quantitative scale used in CT imaging to measure the radiodensity of tissues. It is a standardized unit that relates the attenuation coefficient of a tissue to that of water, which is assigned a value of 0 HU. Different tissues have characteristic HU values, allowing radiologists to differentiate between structures like bone, soft tissue, fat, and air.
Question 42: What is the primary advantage of dual-energy CT (DECT) over single-energy CT?
- Material decomposition enabling differentiation of tissues with similar HU values (Correct answer)
- Reduced contrast volume requirements in all body regions
- Faster gantry rotation speed
- Higher spatial resolution in the z-axis
Correct answer: Material decomposition enabling differentiation of tissues with similar HU values
DECT acquires data at two kVp settings, enabling material decomposition algorithms that distinguish substances like iodine, uric acid, and calcium that may overlap in attenuation at single energy.
Question 43: What is the primary purpose of using automatic exposure control (AEC) in CT?
- Improve spatial resolution
- Adapt tube current to patient size and anatomy (Correct answer)
- Reduce contrast agent volume
- Increase scan speed
Correct answer: Adapt tube current to patient size and anatomy
AEC (also called tube current modulation) automatically adjusts mA based on patient attenuation to optimize image quality while minimizing dose.
Question 44: Which CT technique reduces dose by acquiring fewer projections and reconstructing with iterative algorithms?
- Dual-energy CT
- Sequential CT
- High-pitch spiral CT
- Low-dose CT with iterative reconstruction (Correct answer)
Correct answer: Low-dose CT with iterative reconstruction
Low-dose CT combined with iterative reconstruction algorithms can reduce dose by up to 80% compared to traditional filtered back projection while maintaining diagnostic quality.
Question 45: On CT brain, which structure normally appears hyperdense (bright) without contrast administration?
- CSF in ventricles
- White matter
- Cortical gray matter
- Choroid plexus (Correct answer)
Correct answer: Choroid plexus
The choroid plexus is normally hyperdense due to its highly vascular nature and calcification in adults.
Question 46: Which CT finding in the lung is most associated with malignancy in a solitary pulmonary nodule?
- Spiculated (irregular/stellate) margins with pleural tethering (Correct answer)
- Smooth, well-defined margins
- Satellite nodules
- Central calcification in a popcorn pattern
Correct answer: Spiculated (irregular/stellate) margins with pleural tethering
Spiculated margins reflect infiltrative tumor growth into adjacent lung parenchyma and are the strongest morphologic predictor of malignancy in a solitary pulmonary nodule.
Question 47: Brain non-contrast CT: Hypertension history and rapid start of left facial droop. What does the picture depict?
- Enhancing mass
- Intraventricular haemorrhage
- Subarachnoid haemorrhage
- Intracerebral haemorrhage (Correct answer)
- Subdural haemorrhage
Correct answer: Intracerebral haemorrhage
The clinical presentation of hypertension and rapid onset of neurological deficits like left facial droop strongly suggests an acute stroke. On a non-contrast CT scan, acute hemorrhage appears hyperdense (bright white) due to the high protein content of fresh blood. Intracerebral hemorrhage (ICH) is bleeding directly into the brain parenchyma, often associated with uncontrolled hypertension, and would appear as a focal hyperdensity within the brain tissue.
Question 48: Which artifact on CT brain appears as alternating bright and dark bands radiating from a high-density object such as a surgical clip?
- Partial volume artifact
- Motion artifact
- Streak artifact (Correct answer)
- Ring artifact
Correct answer: Streak artifact
Streak artifacts radiate from high-density objects like metal clips or dense bone due to photon starvation and polychromatic X-ray beam effects.
Question 49: What does the term 'partial volume effect' describe in CT imaging?
- Increased noise at the image periphery
- Averaging of attenuation from different tissues within a single voxel (Correct answer)
- Loss of signal at scan edges
- Artifact from metal implants
Correct answer: Averaging of attenuation from different tissues within a single voxel
Partial volume averaging occurs when a voxel contains multiple tissue types, and the assigned HU is an average of all tissues present, potentially obscuring small lesions.
Question 50: What is the purpose of CT perfusion imaging?
- To visualize blood vessels
- To evaluate tissue blood flow (Correct answer)
- To monitor contrast injection
- To assess bone density
Correct answer: To evaluate tissue blood flow
CT perfusion imaging is a functional imaging technique that assesses the dynamic distribution of an intravenously injected contrast agent within tissues over time. By analyzing the uptake and washout curves, it provides quantitative information about blood flow, blood volume, and mean transit time in various organs. This is crucial for evaluating conditions like stroke, tumors, and myocardial ischemia.
Question 51: Which artifact appears as dark streaks in a CT image, often caused by dense structures like bones?
- Metallic artifact
- Streak artifact (Correct answer)
- Motion artifact
- Beam-hardening artifact
Correct answer: Streak artifact
Streak artifacts are characterized by dark lines or bands that appear across a CT image, often originating from highly attenuating objects. While beam hardening is a common cause of streaks, especially from dense bone, the general term for these linear obscurations is "streak artifact." Metallic objects also cause severe streak artifacts due to their high attenuation and beam hardening effects.
Question 52: What is the most common type of iodinated contrast media used for CT examinations?
- Barium sulfate suspension
- Ionic high-osmolar contrast
- Non-ionic, water-soluble iodinated contrast (Correct answer)
- Gadolinium-based contrast
Correct answer: Non-ionic, water-soluble iodinated contrast
Non-ionic, water-soluble iodinated contrast agents are preferred for CT because they have lower osmolality and fewer adverse reactions compared to ionic agents.
Question 53: What is the recommended fasting time before IV contrast-enhanced CT in non-emergency patients?
- No fasting required
- 4 hours NPO for solids (Correct answer)
- 12-hour fast
- 24-hour fast
Correct answer: 4 hours NPO for solids
A 4-hour fast is recommended before contrast CT to reduce aspiration risk in the event of nausea or vomiting, while preventing dehydration.
Question 54: On non-contrast CT of the head, acute intraparenchymal hemorrhage typically appears as:
- Ring-enhancing
- Hypodense (dark) relative to brain
- Hyperdense (bright white) relative to adjacent brain parenchyma (Correct answer)
- Isodense to brain
Correct answer: Hyperdense (bright white) relative to adjacent brain parenchyma
Acute blood is hyperdense on NCCT due to the high protein content of clotted hemoglobin (55–75 HU), making it easily distinguishable from normal brain (~35 HU).
Question 55: Which type of intracranial hemorrhage appears hyperdense on a CT scan?
- Subarachnoid hemorrhage
- Epidural hemorrhage (Correct answer)
- Intraventricular hemorrhage
- Subdural hemorrhage
Correct answer: Epidural hemorrhage
All types of acute intracranial hemorrhage, including epidural, subdural, subarachnoid, and intraventricular, appear hyperdense (bright white) on a non-contrast CT scan. This is due to the high protein content of fresh blood, which causes it to attenuate X-rays more than surrounding brain tissue. An epidural hemorrhage, which collects between the dura mater and the inner surface of the skull, will clearly demonstrate this hyperdensity.
Question 56: Which intravenous catheter gauge is recommended for power injection during contrast-enhanced CT?
- 24–26 gauge
- 18–20 gauge (Correct answer)
- Any gauge is acceptable
- 14 gauge or larger
Correct answer: 18–20 gauge
An 18–20 gauge IV catheter is recommended for power injection; smaller gauges risk rupture at high flow rates, while larger gauges are unnecessary.
Question 57: A CT brain performed for trauma shows a lens-shaped (biconvex) hyperdense collection between the inner table of the skull and the brain. The most likely source of bleeding is:
- Middle meningeal artery (Correct answer)
- Anterior cerebral artery
- Vertebral artery
- Bridging cortical veins
Correct answer: Middle meningeal artery
Epidural hematomas are most commonly caused by tearing of the middle meningeal artery, producing the classic biconvex hyperdense collection.
Question 58: Which parameter is controlled by the milliampere-seconds (mAs) setting in CT scanning?
- Tube voltage
- Radiation exposure (Correct answer)
- Slice thickness
- Image resolution
Correct answer: Radiation exposure
Milliampere-seconds (mAs) is a key parameter in CT that directly controls the quantity of X-ray photons produced by the tube. A higher mAs setting means more X-ray photons are generated, leading to increased radiation exposure for the patient. Consequently, mAs directly influences the image noise level and the patient's radiation dose.
Question 59: What procedure results in a more monochromatic beam and lowers patient radiation exposure by reducing the longest wavelength photons?
- Bean hardening
- Detector collimation
- Filtration (Correct answer)
- Slip rings
Correct answer: Filtration
Detectors are the critical components within the CT gantry that measure the X-ray photons that have passed through the patient's body. These detectors convert the transmitted X-ray energy into electrical signals, which are then processed by a computer to reconstruct the cross-sectional images. The number and arrangement of detectors directly impact image quality and scan speed.
Question 60: Which modality provides ionizing radiation-free cross-sectional imaging and is preferred for repeated abdominal follow-up in young patients?
- CT with low-dose protocol
- Fluoroscopy
- Magnetic resonance imaging (MRI) (Correct answer)
- Positron emission tomography (PET)
Correct answer: Magnetic resonance imaging (MRI)
MRI uses magnetic fields and radiofrequency pulses with no ionizing radiation, making it safer for serial follow-up imaging in young or pediatric populations.
Question 61: What does a CT number of -1000 HU represent?
- Air (Correct answer)
- Water
- Fat
- Bone cortex
Correct answer: Air
Air has a CT attenuation value of approximately -1000 Hounsfield Units, the lowest end of the standard CT scale.
Question 62: What does 'multiplanar reformation' (MPR) allow in CT post-processing?
- Reconstruction of 3D surface-rendered models for printing
- Automated segmentation of lung nodules for volumetry
- Subtraction of pre-contrast from post-contrast images
- Display of axial raw data in coronal, sagittal, or oblique planes (Correct answer)
Correct answer: Display of axial raw data in coronal, sagittal, or oblique planes
MPR resamples the volumetric CT dataset to display images in any desired plane, allowing evaluation of anatomy that is difficult to assess on axial images alone.
Question 63: What type of CT post-processing technique is most useful for evaluating complex vascular anatomy by removing bone from CT angiography data?
- Surface rendering
- Minimum Intensity Projection (MinIP)
- Maximum Intensity Projection (MIP) after bone subtraction (Correct answer)
- Volume averaging
Correct answer: Maximum Intensity Projection (MIP) after bone subtraction
MIP displays the brightest voxels along each ray, ideal for showing enhanced vessels; subtracting bone first prevents it from obscuring vascular structures.
Question 64: A digital radiograph was produced while the patient bed was translated (moved) through the gantry, maintaining the X-ray tube and detector array in place.
- What Does a SSCT AND MSCT require?
- What is the purpose of Flying Focal Spot Technology ?
- What is the purpose of an scout image in CT ?
- What is a Scout Image in CT ? (Correct answer)
Correct answer: What is a Scout Image in CT ?
The description accurately defines the acquisition of a scout image in CT, also known as a topogram or surview. A scout image is a preliminary digital radiograph produced by translating the patient bed through the gantry while the X-ray tube and detector array remain stationary. This image is essential for planning the subsequent axial or helical CT scans, allowing the technologist to define the precise scan range and angle.
Question 65: Which of the following best describes the dose profile of a helical CT scan with a pitch greater than 1?
- Gaps exist between dose profiles from successive rotations (Correct answer)
- Patient dose is eliminated
- Overlapping dose profiles occur
- Dose is uniformly distributed
Correct answer: Gaps exist between dose profiles from successive rotations
A pitch >1 means the table moves faster than one detector width per rotation, creating gaps in the dose profile and generally reducing patient dose.
Question 66: Which factor influences the spatial resolution of a CT scan?
- Detector size
- Slice thickness (Correct answer)
- X-ray tube current
- Patient weight
Correct answer: Slice thickness
Spatial resolution refers to a CT scan's ability to distinguish between two closely spaced objects. Slice thickness directly influences the resolution along the z-axis (the patient's long axis). Thinner slices provide finer detail and better separation of structures in the axial plane, thereby improving the overall spatial resolution. This allows for clearer visualization of small anatomical features and subtle abnormalities.
Question 67: Which window level and window width settings are appropriate for evaluating brain parenchyma on CT?
- Level 400 HU, Width 2000 HU
- Level 35 HU, Width 80 HU (Correct answer)
- Level 60 HU, Width 350 HU
- Level −600 HU, Width 1500 HU
Correct answer: Level 35 HU, Width 80 HU
Brain window settings (level ~35 HU, width ~80 HU) optimize contrast between gray matter, white matter, and CSF within the narrow HU range of brain tissue.
Question 68: What does the term "slice thickness" refer to in CT imaging?
- The spacing between slices (Correct answer)
- The width of the gantry
- The strength of the X-ray beam
- The size of the patient
Correct answer: The spacing between slices
One significant advantage of CT over conventional X-ray imaging is its superior ability to differentiate between various soft tissues. While conventional X-rays provide a 2D projection, CT creates detailed cross-sectional images that clearly distinguish structures like organs, muscles, and fat based on their different X-ray absorption properties. This enhanced contrast is crucial for diagnosing a wide range of conditions.
Question 69: Which CT finding best characterizes a simple hepatic cyst?
- Solid lesion with arterial enhancement
- Well-defined, water-density (0–10 HU), non-enhancing lesion with imperceptible walls (Correct answer)
- Calcified lesion with mass effect
- Lesion with internal septations and thick walls
Correct answer: Well-defined, water-density (0–10 HU), non-enhancing lesion with imperceptible walls
Simple hepatic cysts are fluid-filled with near-water attenuation, sharply defined margins, no internal complexity, and no enhancement after contrast administration.
Question 70: Which reconstruction algorithm produces the sharpest bone detail in CT imaging?
- Soft tissue (smooth) kernel
- Standard kernel
- High-frequency (edge-enhancing) kernel (Correct answer)
- Low-frequency kernel
Correct answer: High-frequency (edge-enhancing) kernel
High-frequency or bone kernels accentuate edge transitions, producing sharper images ideal for cortical bone and temporal bone detail, though they increase image noise.
Question 71: In CT fluoroscopy-guided procedures, which group is at highest risk for radiation injury?
- The patient's eyes
- The interventionalist's hands if placed in the beam (Correct answer)
- Other patients in adjacent rooms
- Technologists monitoring from the control room
Correct answer: The interventionalist's hands if placed in the beam
CT fluoroscopy uses continuous X-ray exposure, and the interventionalist's hands can receive very high doses if placed near or in the primary beam.
Question 72: Which type of contrast agent is commonly used in CT scans to enhance blood vessels and soft tissues?
- Gadolinium
- Iron oxide
- Barium sulfate
- Iodine-based (Correct answer)
Correct answer: Iodine-based
Iodine-based contrast agents are the most commonly used type in CT scans to enhance blood vessels and soft tissues. Iodine has a high atomic number, which allows it to effectively absorb X-rays, making structures where it accumulates (like blood vessels or highly vascularized lesions) appear brighter on the CT images. This enhancement is vital for detecting and characterizing various pathologies.
Question 73: A patient's CT of the chest shows a focal area of ground-glass opacity (GGO). What does GGO characteristically represent?
- Calcified lymph nodes in the hilum
- Complete airspace consolidation with no air bronchograms
- Partial filling of alveoli or interstitial thickening with preserved underlying architecture (Correct answer)
- Pleural fluid accumulation in the lung base
Correct answer: Partial filling of alveoli or interstitial thickening with preserved underlying architecture
GGO appears as a hazy increased attenuation that does not obscure underlying vessels or bronchi, reflecting partial alveolar filling or thickened interstitium.
Question 74: What does the term "windowing" refer to in CT image analysis?
- Adjusting the field of view
- Applying a convolution kernel
- Enhancing image sharpness
- Changing the display brightness and contrast (Correct answer)
Correct answer: Changing the display brightness and contrast
Windowing in CT imaging refers to the process of adjusting the range of Hounsfield units (CT numbers) displayed on the monitor, which directly controls the brightness (window level) and contrast (window width) of the image. This allows radiologists to optimize the visualization of specific tissue types, such as bone, soft tissue, or lung, by highlighting their particular density ranges. It's a crucial post-processing technique for diagnostic interpretation.
Question 75: Who is credited with the invention of Computed Tomography (CT)?
- Marie Curie
- Sir Isaac Newton
- Albert Einstein
- Godfrey Hounsfield (Correct answer)
Correct answer: Godfrey Hounsfield
Sir Godfrey Hounsfield, an English electrical engineer, is widely recognized for his groundbreaking work in developing the first commercially viable Computed Tomography (CT) scanner. His invention in the early 1970s revolutionized medical diagnostics by providing detailed cross-sectional images of the body. For this achievement, he was awarded the Nobel Prize in Physiology or Medicine in 1979.
Question 76: What is the primary advantage of using saline flush after contrast injection in CT?
- Reduces scan time
- Increases image noise
- Prevents nephrotoxicity
- Pushes residual contrast into central circulation, improving vascular opacification and reducing dose (Correct answer)
Correct answer: Pushes residual contrast into central circulation, improving vascular opacification and reducing dose
A 30–50 mL saline chaser propels contrast remaining in the IV tubing and peripheral vein into the bloodstream, maximizing the usable contrast bolus.
Question 77: What is the Hounsfield Unit (HU) value of water on a CT scan?
- +1000 HU
- −1000 HU
- 0 HU (Correct answer)
- +100 HU
Correct answer: 0 HU
Water is defined as 0 HU in the Hounsfield scale, which serves as the reference point with air at −1000 HU and dense cortical bone at +1000 HU or greater.
Question 78: On a CT brain, the pineal gland is noted to be shifted 5 mm to the left of midline. This finding most directly suggests:
- A left-sided mass compressing the right hemisphere
- A right-sided mass or edema causing midline shift (Correct answer)
- Normal variant in elderly patients
- Congenital aplasia of the left hemisphere
Correct answer: A right-sided mass or edema causing midline shift
The pineal gland is a reliable midline marker; shift away from the normal position indicates a mass or edema on the contralateral side pushing it.
Question 79: A patient presents with thunderclap headache. CT brain shows hyperdense material in the basal cisterns. The most likely diagnosis is:
- Subarachnoid hemorrhage (Correct answer)
- Cerebral venous thrombosis
- Basilar meningitis
- Bilateral subdural hematomas
Correct answer: Subarachnoid hemorrhage
Subarachnoid hemorrhage classically presents with thunderclap headache and appears as hyperdensity filling the basal cisterns on CT.
Question 80: A CT brain shows a well-defined, homogeneously hyperdense extra-axial mass that broadly attaches to the inner table of the skull and causes hyperostosis of the adjacent bone. This is most consistent with:
- Meningioma (Correct answer)
- Lymphoma
- Dural metastasis
- Epidural hematoma
Correct answer: Meningioma
Meningiomas are characteristically hyperdense on non-contrast CT, show homogeneous enhancement, have a dural base, and frequently cause adjacent bony hyperostosis.
Question 81: What CT sign in the chest is used to identify a pulmonary embolism on CTPA?
- Pleural effusion only
- Hyperdense pulmonary artery on non-contrast CT
- Consolidation with air bronchograms
- Filling defect (hypodense clot) within a pulmonary artery on contrast-enhanced images (Correct answer)
Correct answer: Filling defect (hypodense clot) within a pulmonary artery on contrast-enhanced images
A PE appears as a hypodense (dark) filling defect within the opacified (bright) pulmonary artery lumen on CTPA, representing non-enhancing thrombus.
Question 82: Which type of artifact is caused by high-density materials like metal in the CT scan's field of view?
- Motion artifact
- Ring artifact
- Beam-hardening artifact (Correct answer)
- Partial volume artifact
Correct answer: Beam-hardening artifact
Beam-hardening artifact occurs when the X-ray beam passes through dense materials, such as metal implants or thick bone. The lower-energy photons in the beam are absorbed more readily, leaving a 'harder' (higher-energy) beam. This causes streaks and dark bands in the image, often appearing between two dense objects or at the edges of very dense objects.
Question 83: On a CT urogram, which phase of imaging is used to evaluate the collecting system, ureters, and bladder?
- Non-contrast phase
- Excretory (delayed) phase at 5–10 minutes (Correct answer)
- Arterial phase at 25 seconds
- Portal venous phase at 70 seconds
Correct answer: Excretory (delayed) phase at 5–10 minutes
The excretory phase captures opacified urine within the renal collecting system, ureters, and bladder after contrast is filtered and excreted by the kidneys.
Question 84: Which part of a CT scanner rotates around the patient during the scan?
- Detector array
- Gantry (Correct answer)
- X-ray tube
- Console
Correct answer: Gantry
The gantry is the large, donut-shaped component of a CT scanner that houses the X-ray tube and the detector array. During a CT scan, the X-ray tube and detectors rotate rapidly within the gantry around the patient. This continuous rotation allows for the acquisition of X-ray projections from multiple angles, which are then used to reconstruct detailed cross-sectional images.
Question 85: What is the typical oral contrast agent used to opacify the GI tract during abdominal CT?
- Dilute barium or water-soluble iodinated contrast (Correct answer)
- Gadolinium
- Iohexol injection
- Normal saline
Correct answer: Dilute barium or water-soluble iodinated contrast
Dilute barium sulfate or water-soluble iodinated contrast is given orally to distend and opacify the bowel, distinguishing it from adjacent pathology.
Question 86: Contrast extravasation during CT injection most commonly occurs at which site?
- The pulmonary veins
- The renal pelvis
- The aorta
- The IV catheter insertion site in the antecubital fossa (Correct answer)
Correct answer: The IV catheter insertion site in the antecubital fossa
Extravasation occurs when contrast leaks from a dislodged or infiltrated IV catheter into surrounding soft tissue, most often at the antecubital injection site.
Question 87: What is the typical Hounsfield Unit (HU) range for normal brain tissue on a non-contrast CT scan?
- Computed Tomography (CT) (Correct answer)
- Magnetic Resonance Imaging (MRI)
- Positron Emission Tomography (PET)
- Ultrasound
Correct answer: Computed Tomography (CT)
The Hounsfield Unit (HU) is the standard unit of measurement used to quantify tissue density in Computed Tomography (CT) scans. It is a scale that assigns a value of 0 HU to water and approximately -1000 HU to air, allowing for the differentiation of various tissues based on their X-ray attenuation properties. This unit is fundamental to interpreting CT images.
Question 88: Increasing the mAs in CT by a factor of 4 will have what effect on image noise?
- Eliminate noise entirely
- Quadruple the noise
- Reduce noise by 50% (Correct answer)
- Double the noise
Correct answer: Reduce noise by 50%
Since noise is proportional to 1/√mAs, quadrupling mAs reduces noise by half (√4 = 2).
Question 89: Which principle of radiation protection states that doses should be kept as low as reasonably achievable?
- BERT
- ALARA (Correct answer)
- LNT
- NCRP
Correct answer: ALARA
ALARA (As Low As Reasonably Achievable) is the guiding principle for minimizing radiation exposure to patients and staff.
Question 90: What technique is used to time the CT acquisition to arterial contrast peak in CT angiography?
- Breathing command protocol
- Fixed-delay scanning
- ECG gating
- Bolus tracking (SmartPrep) or test bolus technique (Correct answer)
Correct answer: Bolus tracking (SmartPrep) or test bolus technique
Bolus tracking uses real-time low-dose monitoring of a target vessel; when attenuation reaches a threshold, the diagnostic scan is automatically triggered.
Question 91: What is the primary indication for CT perfusion imaging of the brain?
- Assessment of cerebrovascular ischemia and infarct core vs. penumbra (Correct answer)
- Evaluation of skull fractures
- Detection of intracranial hemorrhage
- Diagnosis of brain tumors
Correct answer: Assessment of cerebrovascular ischemia and infarct core vs. penumbra
CT perfusion maps cerebral blood flow, blood volume, and mean transit time to differentiate irreversibly infarcted tissue from salvageable ischemic penumbra in acute stroke.
Question 92: Which CT finding is the hallmark of carotid artery stenosis evaluation on CT angiography?
- Calcification of the aortic arch
- Percentage of luminal narrowing at the carotid bifurcation (Correct answer)
- Presence of contralateral plaques
- Wall thickening only
Correct answer: Percentage of luminal narrowing at the carotid bifurcation
Carotid CTA quantifies the degree of luminal stenosis (percent narrowing) at the bifurcation and proximal internal carotid artery, which determines treatment decisions for stroke risk reduction.
Question 93: In CT angiography, what timing method uses real-time monitoring of contrast arrival in a target vessel?
- Fixed delay scanning
- Bolus tracking (SmartPrep/Care Bolus) (Correct answer)
- Saline flush technique
- Test bolus technique
Correct answer: Bolus tracking (SmartPrep/Care Bolus)
Bolus tracking monitors a region of interest in real time and automatically triggers scanning when contrast attenuation reaches a preset threshold.
Question 94: Which CT post-processing technique is specifically used to evaluate low-attenuation structures like the airways and lung parenchyma?
- Maximum Intensity Projection (MIP)
- Minimum Intensity Projection (MinIP) (Correct answer)
- Volume Rendering (VR)
- Curved planar reformation (CPR)
Correct answer: Minimum Intensity Projection (MinIP)
MinIP projects the lowest attenuation voxel along each ray, making it ideal for visualizing air-filled structures like bronchi and emphysematous lung.
Question 95: What effect does increasing CT tube voltage (kVp) have on patient dose?
- Has no effect on patient dose
- Decreases dose by reducing scan time
- Reduces dose by hardening the beam
- Increases dose due to higher photon energy and greater beam penetration (Correct answer)
Correct answer: Increases dose due to higher photon energy and greater beam penetration
Higher kVp increases photon energy and quantity (via space charge effect), generally increasing patient dose, though it may allow mAs reduction in large patients.
Question 96: What is the purpose of the CT reconstruction process?
- To increase the radiation dose
- To enhance the spatial resolution
- To remove artifacts from the images
- To convert raw data into an image (Correct answer)
Correct answer: To convert raw data into an image
The CT reconstruction process is the computational step where the raw X-ray attenuation data, collected by the detectors from multiple angles, is processed by complex algorithms. These algorithms convert the projection data into a cross-sectional image (slice) that can be viewed and interpreted by radiologists. Without reconstruction, the raw data is just a collection of measurements, not an image.
Question 97: What is the primary advantage of dual-energy CT (DECT) over single-energy CT?
- Material decomposition allowing tissue characterization (e.g., uric acid vs. calcium) (Correct answer)
- Better spatial resolution
- Lower radiation dose always
- Faster scan acquisition
Correct answer: Material decomposition allowing tissue characterization (e.g., uric acid vs. calcium)
DECT acquires data at two kVp levels and uses material decomposition algorithms to distinguish substances by their different attenuation characteristics at different energies.
Question 98: What does the term 'partial volume averaging' refer to in CT imaging?
- Averaging of multiple scan acquisitions to reduce noise
- Blending of attenuation values from different tissues within a single voxel (Correct answer)
- Reduction in FOV to improve spatial resolution
- Use of partial rotation for faster data acquisition
Correct answer: Blending of attenuation values from different tissues within a single voxel
Partial volume averaging occurs when a voxel contains more than one tissue type, resulting in an averaged attenuation value that misrepresents both tissues.
Question 99: In CT imaging, what does the term "slice thickness" refer to?
- The distance between adjacent slices (Correct answer)
- The diameter of the gantry
- The width of the X-ray beam
- The height of the detector array
Correct answer: The distance between adjacent slices
In CT imaging, 'slice thickness' refers to the thickness of the anatomical cross-section being imaged. However, the question asks about the term in the context of the options, and 'the distance between adjacent slices' is the most accurate description provided for how slices are spaced. This parameter is crucial for ensuring complete coverage of the area of interest and for subsequent 3D reconstructions.
Question 100: Which type of CT scan is commonly used to assess bone density and diagnose osteoporosis?
- Helical CT
- Multidetector CT
- Dual-energy CT (Correct answer)
- Dynamic CT
Correct answer: Dual-energy CT
Dual-energy CT (DECT) uses two different X-ray energy spectra simultaneously or sequentially to acquire data. This allows for the differentiation of materials based on their atomic number, making it particularly useful for assessing bone density, characterizing kidney stones, and distinguishing between different tissue types. Its ability to characterize material composition makes it ideal for bone density assessment.
Question 101: CT angiography (CTA) of the brain is primarily used to evaluate:
- Intraparenchymal calcifications
- Demyelinating lesions
- Intracranial aneurysms and vascular malformations (Correct answer)
- Cortical atrophy
Correct answer: Intracranial aneurysms and vascular malformations
CTA uses rapidly acquired post-contrast images to visualize the cerebral vasculature, detecting aneurysms, AVM, and large-vessel occlusions.
Question 102: On CT of the abdomen, the normal appendix should have a maximum diameter of:
- Less than 1 cm
- Less than 6 mm (Correct answer)
- Less than 2 cm
- Size is not diagnostically relevant
Correct answer: Less than 6 mm
A normal appendix is typically less than 6 mm in outer wall diameter; a distended appendix ≥6 mm with periappendiceal inflammation is diagnostic criteria for acute appendicitis.
Question 103: What type of CT scan is often used for assessing coronary artery disease and cardiac anatomy?
- Axial CT
- Helical CT
- Perfusion CT
- Cardiac CT (Correct answer)
Correct answer: Cardiac CT
Cardiac CT, specifically CT angiography (CTA) of the coronary arteries, is a specialized application of CT imaging. It is designed to visualize the heart and its vessels, making it ideal for assessing coronary artery disease, congenital heart anomalies, and other cardiac pathologies. This technique often involves precise timing of contrast injection and ECG gating to minimize motion artifacts from the beating heart.
Question 104: What does multiplanar reformation (MPR) allow clinicians to do with CT data?
- Generate 3D models for printing
- Remove contrast enhancement from images
- Increase spatial resolution after acquisition
- View reconstructed images in coronal, sagittal, and oblique planes from axial data (Correct answer)
Correct answer: View reconstructed images in coronal, sagittal, and oblique planes from axial data
MPR uses isotropic or near-isotropic volumetric CT data to generate reconstructions in any desired imaging plane without additional scanning.
Question 105: Which population group requires the most careful dose optimization in CT due to higher radiation sensitivity?
- Male patients
- Athletic adults
- Pediatric patients (Correct answer)
- Elderly patients
Correct answer: Pediatric patients
Children are more radiosensitive than adults and have more years of life remaining in which radiation-induced cancers could develop, requiring strict dose minimization.
Question 106: Which CT artifact is caused by two high-density objects lying in the same scan plane?
- Partial volume artifact
- Motion artifact
- Ring artifact
- Beam hardening artifact (Correct answer)
Correct answer: Beam hardening artifact
Beam hardening occurs when polychromatic X-rays are differentially absorbed, causing dark streaks between dense structures like metal implants or dense bone.
Question 107: The DLP (Dose Length Product) in CT is calculated by multiplying CTDI_vol by which parameter?
- Rotation time
- Pitch factor
- Scan length (Correct answer)
- Slice thickness
Correct answer: Scan length
DLP = CTDIvol × scan length (in cm), providing an estimate of total energy deposited during the entire CT examination.
Question 108: What is the primary advantage of iterative reconstruction over filtered back projection (FBP) in CT?
- Significant noise reduction at equivalent or lower dose (Correct answer)
- Elimination of metal artifacts
- Faster reconstruction time
- Better spatial resolution at high kVp
Correct answer: Significant noise reduction at equivalent or lower dose
Iterative reconstruction repeatedly refines the image estimate against raw data, reducing noise substantially, which allows lower dose acquisition without sacrificing image quality.
Question 109: In CT, what does 'z-axis resolution' refer to?
- Spatial resolution along the longitudinal (head-to-foot) axis of the patient (Correct answer)
- In-plane pixel size in the axial plane
- Contrast resolution between adjacent soft tissue structures
- Temporal resolution of the rotating gantry
Correct answer: Spatial resolution along the longitudinal (head-to-foot) axis of the patient
Z-axis (longitudinal) resolution determines the ability to distinguish structures stacked along the patient's long axis and improves with thinner collimation.
Question 110: The patient should be questioned about the following matters before receiving an intravenous injection of iodinated contrast material: <br> <br> 1. Renal Function- One kidney dose is cut <br> 2. Allergic History- Premedicated <br> 3. Cardiac History- Heart disease
- 2 only
- 1 only
- 1,2, and 3 (Correct answer)
- 3 only
Correct answer: 1,2, and 3
Before administering iodinated contrast material, it is crucial to assess the patient's renal function, as the kidneys excrete the contrast, and impairment can lead to contrast-induced nephropathy. An allergic history is vital to identify potential hypersensitivity reactions to iodine, which can range from mild to severe. Cardiac history is also important because contrast media can affect heart function, especially in patients with pre-existing cardiac conditions, increasing the risk of adverse events.
Question 111: Which term refers to the ability of a CT scanner to differentiate between objects with similar densities?
- Temporal resolution
- Contrast resolution (Correct answer)
- Image resolution
- Spatial resolution
Correct answer: Contrast resolution
Contrast resolution is the ability of a CT system to distinguish between tissues that have only slightly different X-ray attenuation coefficients (i.e., similar densities). High contrast resolution is crucial for detecting subtle lesions or differentiating between soft tissues like gray and white matter in the brain. It is influenced by factors such as mAs, slice thickness, and reconstruction algorithms.
Question 112: Which unit is used to express the effective radiation dose received by a patient during a CT scan?
- Rad
- Millisievert (mSv) (Correct answer)
- Milligray (mGy)
- Roentgen (R)
Correct answer: Millisievert (mSv)
Millisievert (mSv) is the SI unit used to express effective dose, which accounts for the biological effect of radiation on different tissues.
Question 113: Which CT scan view provides the best visualization of the interhemispheric fissure?
- Axial
- Oblique
- Coronal (Correct answer)
- Sagittal
Correct answer: Coronal
The interhemispheric fissure is the deep groove that separates the two cerebral hemispheres, running from the front to the back of the brain. A coronal CT scan view, which slices the brain from ear to ear (like looking at the face), provides an optimal cross-sectional image along the length of this fissure. This orientation allows for a clear and extensive visualization of its entire extent and any associated pathology.
Question 114: What is the purpose of a CT scout image (localizer) in patient positioning?
- To provide a preliminary diagnosis
- To determine optimal contrast dose
- To assess motion artifacts
- To outline the scanning region (Correct answer)
Correct answer: To outline the scanning region
The CT scout image, also known as a localizer or topogram, is a preliminary low-dose X-ray image taken before the diagnostic scan. It serves as a reference image for the technologist to accurately define the anatomical boundaries and precise start and end points for the subsequent diagnostic CT scan. This ensures that the correct area of interest is included in the scan and minimizes unnecessary radiation exposure.
Question 115: Which type of CT scan involves acquiring images at multiple time points to visualize the distribution of contrast agent in tissues?
- Dynamic CT
- Perfusion CT
- Multiplanar CT
- Virtual CT
Dynamic CT involves acquiring a series of images over time after the administration of a contrast agent. This technique allows for the visualization of the contrast agent's uptake, distribution, and washout kinetics within tissues and organs. It is used to assess vascularity, perfusion, and the enhancement patterns of lesions, providing valuable functional information.
Question 116: What is the purpose of CT dose modulation techniques?
- To reduce patient discomfort
- To adjust the tube voltage
- To optimize radiation exposure (Correct answer)
- To eliminate motion artifacts
Correct answer: To optimize radiation exposure
CT dose modulation techniques are advanced features designed to automatically adjust the X-ray tube current (mAs) during a scan based on the patient's anatomy. By reducing the dose in thinner or less attenuating regions and increasing it in thicker or denser areas, these techniques optimize radiation exposure. This maintains image quality while significantly reducing the overall patient dose.
Question 117: Which phase of contrast enhancement is best for evaluating hypervascular liver tumors such as hepatocellular carcinoma (HCC)?
- Delayed phase (3–5 minutes)
- Portal venous phase (60–70 seconds)
- Non-contrast phase
- Arterial phase (25–35 seconds post-injection) (Correct answer)
Correct answer: Arterial phase (25–35 seconds post-injection)
HCC derives its blood supply primarily from the hepatic artery and enhances dramatically during the arterial phase, a hallmark feature used for diagnosis.
Question 118: Which type of iodinated contrast agent has the lowest risk of adverse reactions?
- Low-osmolality nonionic monomers (Correct answer)
- High-osmolality ionic monomers
- High-osmolality nonionic agents
- Low-osmolality ionic dimers
Correct answer: Low-osmolality nonionic monomers
Low-osmolality nonionic monomers (e.g., iohexol, iopamidol) have the lowest rates of adverse reactions due to reduced chemotoxicity and near-physiologic osmolality.
Question 119: Which phase of contrast enhancement is optimal for detecting hypervascular liver lesions such as hepatocellular carcinoma?
- Delayed equilibrium phase (3–5 minutes)
- Non-contrast phase
- Portal venous phase (60–70 seconds)
- Arterial phase (20–35 seconds post-injection) (Correct answer)
Correct answer: Arterial phase (20–35 seconds post-injection)
Hypervascular lesions like HCC receive predominantly arterial blood supply and enhance brightly relative to normal liver parenchyma during the arterial phase.
Question 120: What type of adverse reaction to iodinated contrast media involves an immune-mediated hypersensitivity response?
- Anaphylactoid (allergy-like) reaction (Correct answer)
- Contrast extravasation
- Vasovagal syncope
- Contrast nephropathy
Correct answer: Anaphylactoid (allergy-like) reaction
Anaphylactoid reactions mimic true anaphylaxis (urticaria, bronchospasm, hypotension) but do not require prior sensitization and are not IgE-mediated.
Question 121: Which windowing preset is most appropriate for evaluating pulmonary nodules on CT?
- Bone window (W:2000, L:400)
- Brain window (W:80, L:40)
- Soft tissue window (W:400, L:40)
- Lung window (W:1500, L:-600) (Correct answer)
Correct answer: Lung window (W:1500, L:-600)
Lung window settings (wide width, negative level) optimize contrast for air-filled lung parenchyma and nodule detection.
Question 122: A ring-enhancing lesion in the brain on post-contrast CT with surrounding edema in an immunocompromised patient most likely represents:
- Glioblastoma multiforme
- Meningioma
- Cavernous malformation
- Toxoplasma abscess (Correct answer)
Correct answer: Toxoplasma abscess
Toxoplasmosis is the most common ring-enhancing lesion in immunocompromised (HIV/AIDS) patients, presenting as multiple ring-enhancing lesions.
Question 123: In CT angiography, what is the purpose of a test bolus or bolus tracking technique?
- To optimize scan timing relative to contrast arrival in the target vessel (Correct answer)
- To assess patient allergy risk before full injection
- To determine patient weight-based contrast volume
- To calibrate the CT detector gain before contrast injection
Correct answer: To optimize scan timing relative to contrast arrival in the target vessel
Bolus tracking and test bolus methods synchronize scan initiation with peak contrast enhancement in the target vessel to ensure optimal opacification.
Question 124: The CT finding of 'cytotoxic edema' is characterized by:
- Hypodensity that respects vascular territory boundaries and involves both gray and white matter (Correct answer)
- Enhancement along the cortical ribbon
- Hyperdensity within the cortex after contrast
- Hypodensity confined to white matter that crosses vascular territories
Correct answer: Hypodensity that respects vascular territory boundaries and involves both gray and white matter
Cytotoxic edema from ischemia causes swelling of neurons and glia within a vascular territory, appearing as hypodensity involving both cortex and white matter.
Question 125: Which component of a CT scanner rotates around the patient during image acquisition?
- Detector array
- Gantry (Correct answer)
- X-ray tube
- Console
Correct answer: Gantry
In CT imaging, 'slice thickness' refers to the width of the individual cross-sectional images acquired by the scanner. It essentially defines the dimension of the tissue volume represented by each image. While it can influence the detail within a single slice, it more accurately describes the spacing or thickness of the anatomical section being imaged.
Question 126: A radiopaque contrast agent, a material that can be seen on X-rays to aid in identifying one tissue from another, may be administered to patients before a CT scan. Which of the following contrast agents is used with CT the most frequently?
- Gadolinium
- Barium
- Iodine (Correct answer)
- Microbubbles
Correct answer: Iodine
Metal objects, such as jewelry and zippers, are radiopaque, meaning they absorb X-rays. During a CT scan, these items can block the X-ray beam, creating artifacts or streaks on the images. These artifacts can obscure important anatomical structures, making it difficult for radiologists to accurately interpret the scan results and potentially leading to misdiagnosis.
Question 127: In CT imaging, 'beam hardening artifact' in the posterior fossa is caused by:
- Excessive mAs settings
- Metal implants in the cervical spine
- Patient motion during scanning
- Preferential absorption of low-energy X-ray photons by the dense temporal bones (Correct answer)
Correct answer: Preferential absorption of low-energy X-ray photons by the dense temporal bones
Beam hardening occurs when dense bone absorbs low-energy photons, causing dark streaks between the temporal bones that obscure posterior fossa structures.
Question 128: During a dual-energy CT scan, what primary advantage does dual-energy provide over single-energy CT?
- Material differentiation and virtual non-contrast imaging (Correct answer)
- Improved temporal resolution
- Lower radiation dose in all cases
- Faster image acquisition
Correct answer: Material differentiation and virtual non-contrast imaging
Dual-energy CT acquires data at two different kVp settings, enabling material decomposition to distinguish iodine, calcium, and uric acid, and to create virtual non-contrast images.
Question 129: What is the primary shielding material used in CT scanning room walls?
- Aluminum
- Fiberglass
- Lead or lead-equivalent concrete (Correct answer)
- Steel
Correct answer: Lead or lead-equivalent concrete
CT room walls are constructed with lead or high-density concrete to attenuate primary and scatter radiation to acceptable levels for adjacent areas.
Question 130: At what injection rate (mL/sec) is intravenous contrast typically administered for a CT pulmonary angiography (CTPA)?
- 10–15 mL/sec
- 3–5 mL/sec (Correct answer)
- Less than 0.5 mL/sec
- 0.5–1 mL/sec
Correct answer: 3–5 mL/sec
CTPA requires a brisk injection rate of 3–5 mL/sec to achieve adequate pulmonary artery opacification during the brief arterial phase.
Question 131: What does the term 'pitch' refer to in helical CT scanning?
- The number of detector rows
- The angle of the X-ray beam relative to the patient
- The speed of gantry rotation in RPM
- The table distance traveled per gantry rotation divided by beam collimation (Correct answer)
Correct answer: The table distance traveled per gantry rotation divided by beam collimation
Pitch is calculated as table feed per rotation divided by total beam collimation width; a pitch of 1 means the table moves exactly one beam width per rotation.
Question 132: A pregnant patient requires abdominal CT for suspected appendicitis. Which statement about fetal radiation risk is most accurate?
- Radiation risk is identical regardless of gestational age
- The fetus is fully shielded by the uterus and receives no radiation dose
- CT is absolutely contraindicated in all trimesters
- Fetal risk is negligible at typical diagnostic doses below 50 mGy, but benefits must outweigh risks (Correct answer)
Correct answer: Fetal risk is negligible at typical diagnostic doses below 50 mGy, but benefits must outweigh risks
At diagnostic CT doses typically below 25 mGy, deterministic fetal effects are extremely unlikely; however, the ALARA principle still applies and clinical benefit must justify the exposure.
Question 133: What is the primary purpose of retrospective gating in cardiac CT?
- To reduce radiation dose
- To eliminate the need for beta-blockers
- To reconstruct images at any phase of the cardiac cycle (Correct answer)
- To increase spatial resolution
Correct answer: To reconstruct images at any phase of the cardiac cycle
Retrospective gating records ECG data throughout the scan and allows reconstruction at any desired cardiac phase after acquisition.
Question 134: Which artifact appears as a dark band between two high-density objects on CT images?
- Motion artifact
- Ring artifact
- Beam hardening artifact (Correct answer)
- Partial volume artifact
Correct answer: Beam hardening artifact
Beam hardening causes dark streaks or bands between dense objects like bone or metal because lower-energy photons are preferentially absorbed.
Question 135: What is the purpose of a CT collimator?
- To adjust image brightness
- To enhance contrast resolution
- To shape the X-ray beam (Correct answer)
- To reduce patient dose
Correct answer: To shape the X-ray beam
A CT collimator is a device that shapes and restricts the X-ray beam emitted from the tube. Its primary purpose is to define the slice thickness and ensure that only the necessary area is exposed to radiation. This precise beam shaping reduces scatter radiation, improves image quality, and significantly lowers the patient's radiation dose.
Question 136: What technology allows modern helical CT scanners to rotate continuously without stopping?
- Electromagnetic braking
- Cable drive system
- Gear reduction transmission
- Slip ring technology (Correct answer)
Correct answer: Slip ring technology
Slip ring technology enables continuous electrical contact between the rotating gantry and stationary components, allowing uninterrupted 360° rotation.
Question 137: What is the primary advantage of multislice (multidetector) CT over single-slice CT?
- Lower radiation dose
- Improved image contrast
- Faster scan times (Correct answer)
- Higher spatial resolution
Correct answer: Faster scan times
Multislice (or multidetector) CT scanners use multiple rows of detectors, allowing them to acquire several slices simultaneously with each rotation of the X-ray tube. This significantly reduces scan times compared to single-slice CT, where only one slice is acquired per rotation. Faster scan times are beneficial for trauma, cardiac imaging, and reducing motion artifacts, especially in uncooperative patients.
Question 138: What is the recommended lead apron shielding thickness to protect gonads from scattered CT radiation?
- 0.5 mm lead equivalent (Correct answer)
- 0.1 mm lead equivalent
- 2.0 mm lead equivalent
- No shielding is needed
Correct answer: 0.5 mm lead equivalent
A 0.5 mm lead equivalent apron is standard for gonadal shielding, though its effectiveness in CT is debated due to scatter from all directions.
Question 139: Which of the following technical aspects directly affects the dose given to patients?
- Algorithm
- Matrix size
- Milliampere-second {MAS} (Correct answer)
- Window level
Correct answer: Milliampere-second {MAS}
Milliampere-second (mAs) directly controls the quantity of X-ray photons produced during a CT scan. A higher mAs setting means more X-rays are generated and delivered to the patient, resulting in a proportionally higher radiation dose. Therefore, adjusting the mAs is a primary method for controlling and optimizing the radiation dose given to patients in CT imaging.
Question 140: What is the Hounsfield unit (HU) scale used for in CT imaging?
- Determining patient weight
- Measuring radiation dose
- Quantifying image noise
- Describing tissue density (Correct answer)
Correct answer: Describing tissue density
The Hounsfield unit (HU) scale is a quantitative scale used in CT imaging to represent the relative density of tissues. Water is assigned a value of 0 HU, air is approximately -1000 HU, and dense bone can be +1000 HU or more. This scale allows radiologists to differentiate between various tissues and pathologies based on their X-ray attenuation properties.
Question 141: In CT imaging, what is a "Hounsfield Unit" (HU) a measure of?
- Pixel size
- Image brightness
- Tissue density (Correct answer)
- Radiation dose
Correct answer: Tissue density
The gantry is the large, donut-shaped component of a CT scanner that houses the X-ray tube and the detector array. During image acquisition, the gantry rotates rapidly around the patient, allowing the X-ray tube to emit radiation from multiple angles and the detectors to capture the transmitted X-rays. This rotation is essential for collecting the data needed to reconstruct cross-sectional images.
Question 142: On CT brain, the falx cerebri normally appears:
- Invisible without contrast
- Isodense to brain parenchyma
- Hypodense relative to brain parenchyma
- Mildly hyperdense relative to brain parenchyma (Correct answer)
Correct answer: Mildly hyperdense relative to brain parenchyma
The falx cerebri is a dural reflection that appears mildly hyperdense on CT due to its fibrous dural composition and sometimes calcification.
Question 143: A thyroid cancer patient requires radioiodine ablation therapy. Why must iodinated CT contrast be avoided before this treatment?
- It permanently destroys the thyroid gland
- Iodinated contrast saturates the sodium-iodide symporter, blocking therapeutic radioiodine uptake (Correct answer)
- It causes thyroid storm
- It interferes with TSH levels only
Correct answer: Iodinated contrast saturates the sodium-iodide symporter, blocking therapeutic radioiodine uptake
Excess stable iodine from contrast saturates thyroid transporters and competes with therapeutic I-131, significantly reducing radioiodine ablation efficacy for 2–8 weeks.
Question 144: What part of the CT gantry measures photons that are transmitted and pass through the patient?
- Detectors (Correct answer)
- Slip rings
- Collimators
- Filters
Correct answer: Detectors
Solid state crystalline detectors, such as those made from cadmium tungstate or ceramic scintillators, are widely used in modern CT scanners. They are highly efficient at converting X-ray photons into light, which is then converted into electrical signals. Their high efficiency and fast response times contribute to improved image quality and reduced patient radiation dose.
Question 145: What is the typical unit of measurement used for the density of tissues in CT scans?
- Hounsfield Unit (HU) (Correct answer)
- Tissue Attenuation Index (TAI)
- X-ray Absorption Unit (XAU)
- Tissue Density Unit (TDU)
Correct answer: Hounsfield Unit (HU)
The Hounsfield Unit (HU) is the standard quantitative scale used in CT imaging to measure and display the radiodensity of tissues. This scale is normalized, with water assigned a value of 0 HU and air approximately -1000 HU. Different tissues, such as bone, fat, and various organs, have characteristic HU values, allowing for their differentiation and analysis on CT images.
Question 146: Which reconstruction algorithm provides the sharpest images but also the most noise in CT?
- Iterative reconstruction
- High-frequency (bone/edge) kernel (Correct answer)
- Soft-tissue kernel
- Lung kernel
Correct answer: High-frequency (bone/edge) kernel
High-frequency (bone/detail) kernels sharpen edges and fine structures but amplify image noise, making them ideal for bone but not soft tissue evaluation.
Question 147: A CT of the head shows enlargement of all four ventricles with periventricular hypodensity. The most likely diagnosis is:
- Diffuse axonal injury
- Communicating hydrocephalus with transependymal edema (Correct answer)
- Cerebral atrophy
- Bilateral subdural hematomas
Correct answer: Communicating hydrocephalus with transependymal edema
Communicating hydrocephalus causes symmetric dilation of all ventricles and periventricular hypodensity from transependymal CSF migration.
Question 148: What is the CT angiography finding that indicates aortic dissection?
- Linear calcification within the aortic lumen
- Focal outpouching of the aortic wall
- An intimal flap dividing the aortic lumen into true and false channels (Correct answer)
- Diffuse aortic wall thickening
Correct answer: An intimal flap dividing the aortic lumen into true and false channels
Aortic dissection is characterized by a longitudinal intimal tear creating a flap that separates the aorta into two lumens — the true and false channels — visible on contrast CT.
Question 149: Which serum creatinine-based calculation is most commonly used to screen for contrast-induced nephropathy risk before CT?
- BUN/creatinine ratio
- eGFR (estimated Glomerular Filtration Rate) (Correct answer)
- Creatinine clearance via timed urine collection
- Urine specific gravity
Correct answer: eGFR (estimated Glomerular Filtration Rate)
eGFR calculated from serum creatinine is the standard screening tool; an eGFR <30 mL/min/1.73m² generally indicates high risk for contrast nephrotoxicity.
Question 150: What is the main advantage of using iterative reconstruction over filtered back projection (FBP)?
- Improved temporal resolution
- Reduced image noise at lower radiation doses (Correct answer)
- Faster reconstruction speed
- Elimination of all motion artifacts
Correct answer: Reduced image noise at lower radiation doses
Iterative reconstruction algorithms reduce noise more effectively than FBP, enabling diagnostic-quality images at significantly lower radiation doses.
Question 151: Which factor contributes to the occurrence of motion artifacts in CT imaging?
- Slow heart rate
- Fast scan times (Correct answer)
- High radiation dose
- Sedation of the patient
Correct answer: Fast scan times
While fast scan times are primarily implemented to *reduce* the likelihood of motion artifacts by minimizing the time available for patient movement, they are often employed in imaging scenarios where involuntary motion (e.g., cardiac or respiratory motion) is inherently present. In these dynamic situations, even rapid data acquisition may not entirely eliminate the effects of motion, leading to residual artifacts. Thus, fast scan times are a response to, rather than a direct contributor to, the *occurrence* of motion, but motion can still be a challenge despite them.
Question 152: What is the term for the minimum detectable difference in attenuation between two adjacent structures in CT?
- Spatial resolution
- Temporal resolution
- Contrast resolution (Correct answer)
- Noise equivalent quanta
Correct answer: Contrast resolution
Contrast resolution refers to the ability to distinguish between tissues with similar attenuation values, and it is primarily affected by noise and radiation dose.
Question 153: Which factor most significantly increases a patient's risk for a severe contrast reaction?
- Asthma controlled with inhalers
- History of a prior severe contrast reaction (Correct answer)
- Mild shellfish allergy
- Age over 65
Correct answer: History of a prior severe contrast reaction
A prior severe contrast reaction is the single strongest risk factor for a repeat severe reaction, increasing risk approximately 5-fold.
Question 154: On CT brain, the normal attenuation of CSF within the ventricles approximates:
- 35–45 HU
- 20–30 HU
- 0–10 HU (Correct answer)
- 50–60 HU
Correct answer: 0–10 HU
CSF has nearly the same density as water and measures 0–10 HU on CT, appearing dark (hypodense) relative to brain parenchyma.
Question 155: Which type of CT scan is often used for imaging larger areas of the body with minimal patient movement?
- Axial CT
- ultidetector CT
- Perfusion CT
- Helical CT (Correct answer)
Correct answer: Helical CT
Helical CT, also known as spiral CT, involves continuous rotation of the X-ray tube and detectors while the patient table moves through the gantry. This allows for rapid, continuous acquisition of volumetric data over large anatomical areas. This method minimizes motion artifacts and reduces scan time, making it ideal for imaging extensive body regions with minimal patient movement.
Question 156: What is the typical gantry rotation time for a modern high-speed CT scanner?
- 0.27–0.5 seconds per rotation (Correct answer)
- 0.05–0.1 seconds per rotation
- 2–5 seconds per rotation
- 1–2 seconds per rotation
Correct answer: 0.27–0.5 seconds per rotation
Modern multi-detector CT scanners complete a full 360° rotation in as little as 0.27 seconds, enabling fast cardiac and dynamic imaging.
Question 157: When performing a CT brain for evaluation of acute stroke, the primary reason to acquire a non-contrast series BEFORE contrast-enhanced sequences is:
- To reduce radiation dose from overlapping sequences
- To measure baseline Hounsfield units of the tumor
- To detect acute hemorrhage, which is a contraindication to thrombolytic therapy (Correct answer)
- To assess the degree of atherosclerotic calcification
Correct answer: To detect acute hemorrhage, which is a contraindication to thrombolytic therapy
Identifying acute hemorrhage on non-contrast CT is critical because thrombolytics (tPA) are absolutely contraindicated if hemorrhage is present.
Question 158: Which parameter determines the amount of radiation exposure delivered to the patient during a CT scan?
- Reconstruction algorithm
- Slice thickness
- Milliampere-seconds (mAs) (Correct answer)
- Kilovoltage (kV)
Correct answer: Milliampere-seconds (mAs)
Milliampere-seconds (mAs) is a direct determinant of the quantity of X-ray photons produced by the CT tube, and thus directly controls the radiation exposure to the patient. A higher mAs setting results in more X-ray photons, leading to increased patient dose but also reduced image noise. Kilovoltage (kV) also affects dose, but mAs is the primary parameter for controlling the overall photon quantity.
Question 159: What does the size-specific dose estimate (SSDE) correct for compared to CTDIvol?
- Scanner manufacturer differences
- Contrast agent type
- Scan protocol complexity
- Patient body size and diameter (Correct answer)
Correct answer: Patient body size and diameter
SSDE adjusts CTDIvol based on the patient's effective diameter to provide a more accurate estimate of actual patient dose.
Question 160: What is the advantage of using a smaller CT field of view (FOV)?
- Faster scan time
- Wider anatomical coverage
- Reduced patient dose
- Improved spatial resolution for the targeted anatomy (Correct answer)
Correct answer: Improved spatial resolution for the targeted anatomy
A smaller FOV assigns more pixels per unit area, increasing spatial resolution and displaying finer anatomical detail in the region of interest.
ARRT Computed Tomography (CT) Exam
This exam certifies technologists in the specialized imaging modality of Computed Tomography, demonstrating competency in patient care, safety, and image acquisition.
Exam Rules
- You can skip questions and return to them later
- Flag questions for review before submitting
- No feedback shown until you submit the entire exam
- Unanswered questions count as wrong — answer everything
- 10 pretest questions are mixed in and don't affect your score
- Timer auto-submits when time runs out
- Your progress is auto-saved every 30 seconds
Practice questions written with AI assistance and published by Nathan Beja, founder of Practice Test Geeks. They are not real exam questions. Exam facts are checked against the official sources linked below. Editorial policy