  1. During a tonsillectomy, a fire erupts in the patient's airway. What should the team do immediately?

    Answer: Remove the endotracheal tube and stop the flow of all airway gases

    For an airway fire, the endotracheal tube is removed and all airway gases are stopped at the same time, then flammables are removed and saline is poured into the airway.

  2. Which item is an oxidizer in the surgical fire triangle?

    Answer: Nitrous oxide

    Oxygen and nitrous oxide both support combustion, so they are the oxidizers in the fire triangle.

  3. How long should an alcohol-based skin antiseptic dry before drapes are applied?

    Answer: Until it looks dry, per the manufacturer's instructions (at least 3 minutes on hairless skin)

    Flammable prep must dry fully, following the manufacturer's minimum drying time, so vapors don't build up under the drapes.

  4. A fire starts on the drapes and the circulator cannot put it out with saline. Which extinguisher is generally recommended in the OR because it is safe to use around patients?

    Answer: Carbon dioxide (CO2) extinguisher

    CO2 extinguishers leave no residue and work on Class A, B, and C fires near patients, so they are preferred in the OR.

  5. What is the main purpose of the return electrode monitoring (REM) system on an electrosurgical unit?

    Answer: To find poor contact between the dispersive pad and the patient and stop output

    REM checks the impedance at the pad and stops the generator if contact drops, which helps prevent burns at the pad site.

  6. Where should the electrosurgical dispersive pad be placed?

    Answer: Over a clean, dry, well-vascularized muscle mass close to the surgical site

    Muscle with good blood flow spreads current safely, and placing the pad close to the site shortens the current path.

  7. When the active electrode is not in use during a procedure, where should it be kept?

    Answer: In a clean, nonconductive safety holster

    A holster keeps the active electrode from being activated by accident and burning the patient or the drapes.