Water Treatment and Dialysis Equipment Flashcards
6 cards from real CDN practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 6 Water Treatment and Dialysis Equipment flashcards as text
What is the purpose of carbon tanks in the dialysis water treatment system?
Answer: Remove chlorine and chloramines via adsorption
Activated carbon tanks remove chlorine and chloramines through adsorption, protecting the RO membrane and preventing patient exposure to these hemolytic agents.
During a dialysis machine's self-test, a conductivity alarm activates indicating high conductivity. What does this most likely indicate?
Answer: Incorrect proportioning of dialysate concentrate
High conductivity in the dialysate circuit indicates improper proportioning of the acid or bicarbonate concentrate, resulting in a solution that is too concentrated.
What action should be taken if the dialysis machine's venous pressure alarm activates with a sudden increase in venous pressure?
Answer: Check for venous needle kinking, clotting, or access obstruction
A sudden increase in venous pressure suggests obstruction in the venous return pathway — check for needle kinking, clotting in the venous chamber, or access stenosis.
What does the dialysis machine's blood leak detector monitor?
Answer: Presence of blood in the spent dialysate
The blood leak detector uses a photocell to detect blood in the spent dialysate effluent, signaling a breach in the dialyzer membrane.
What is the correct response when the air detector alarm activates during hemodialysis?
Answer: Immediately clamp the venous line and stop the blood pump
An air detector alarm requires immediate clamping of the venous blood line and stopping the blood pump to prevent air embolism, which can be fatal.
Which disinfection method is most commonly used for the internal water distribution loop in a dialysis facility?
Answer: Heat disinfection or chemical disinfection with bleach or peracetic acid
Heat disinfection (hot water flushing) or chemical disinfection with sodium hypochlorite or peracetic acid are the standard methods for disinfecting dialysis water distribution loops.
How this page was made: written with AI tools and published by Nathan Beja, founder of Practice Test Geeks, who is responsible for its content. Exam facts are checked against the official sources listed below. Not reviewed by a subject-matter expert. Editorial policy
Sources & references
Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy
- cnetnurse.com — Test delivery, fees, score release · verified Oct 2026
- nncc-exam.org — Score validity, passing score, registration deadline, retake policy, test dates, exam format, time limit · verified Oct 2026