  1. Which biological indicator organism is used to validate ethylene oxide (EtO) sterilization cycles?

    Answer: Bacillus atrophaeus

    Bacillus atrophaeus spores are the standard biological indicator for EtO sterilization validation.

  2. A steam autoclave fails its Bowie-Dick test. What is the most likely cause?

    Answer: Air entrapment in the chamber

    The Bowie-Dick test detects air removal failures; a failed test indicates residual air is preventing steam penetration.

  3. What is the primary advantage of low-temperature hydrogen peroxide plasma sterilization over EtO?

    Answer: No toxic residuals requiring aeration

    Hydrogen peroxide plasma leaves no toxic residuals, eliminating the lengthy aeration period required after EtO sterilization.

  4. Which parameter is NOT directly monitored during a standard steam sterilization cycle?

    Answer: Humidity

    Standard steam sterilization monitoring focuses on temperature, pressure, and time; humidity is a critical parameter for EtO, not steam cycles.

  5. An ultrasonic cleaner used for pre-sterilization cleaning operates primarily through which mechanism?

    Answer: Acoustic cavitation

    Ultrasonic cleaners work via acoustic cavitation, where rapidly forming and collapsing bubbles dislodge debris from instrument surfaces.

  6. When performing a spore test on a tabletop autoclave, how often does AAMI recommend routine biological monitoring?

    Answer: Weekly

    AAMI ST55 recommends at minimum weekly spore testing for tabletop steam sterilizers used in healthcare settings.

  7. Which type of packaging material is incompatible with hydrogen peroxide plasma sterilization?

    Answer: Paper/cellulose-based wraps

    Cellulose-based materials absorb hydrogen peroxide, depleting the sterilant and causing cycle failures in plasma sterilizers.