Image Acquisition and Evaluation Flashcards
7 cards from real ARRT practice questions. Tap to flip, then mark Knew It or Still Learning โ missed cards come back until you master them.
Read the first 7 Image Acquisition and Evaluation flashcards as text
A grid with a ratio of 16:1 is most appropriate for which clinical scenario?
Answer: Large body part requiring high kVp technique
High-ratio grids (12:1 or 16:1) are used for large anatomical areas where significant scatter is produced, especially at higher kVp settings.
Quantum mottle (noise) in digital radiography is primarily caused by which factor?
Answer: Too few X-ray photons reaching the detector
Quantum mottle results from insufficient photon flux reaching the detector, producing a grainy, noisy appearance on the image.
Which positioning error causes grid cutoff on a cross-table lateral projection?
Answer: Angling the tube across the grid lines
Angling the tube perpendicular to the grid lines (off-axis in the wrong direction) causes photons to be absorbed, resulting in grid cutoff.
On a PA chest radiograph evaluated for proper exposure, which anatomical landmark is used to verify adequate penetration?
Answer: Faint visibility of the thoracic spine through the cardiac shadow
Adequate penetration on a PA chest is confirmed when the thoracic vertebrae are faintly visible through the cardiac silhouette.
A radiograph exhibits a wide latitude (long scale) of contrast. Which exposure factor most directly caused this?
Answer: High kVp
Higher kVp produces more penetrating, homogenous radiation that results in low contrast (long gray scale) across the image.
Which CR/DR artifact appears as a bright or dark line running across the image parallel to the scan direction?
Answer: Dropout line artifact
A dropout line artifact occurs when a row or column of detector elements fails, creating a bright or dark line across the image.
Which action best reduces magnification of a structure on a radiograph?
Answer: Decrease OID
Decreasing OID (object-to-image distance) brings the object closer to the receptor, reducing magnification.