โ† All ABO NOCE Basic Opticianry Flashcard Decks

Dispensing Instrumentation Questions and Answers Flashcards

7 cards from real ABO NOCE Basic Opticianry practice questions. Tap to flip, then mark Knew It or Still Learning โ€” missed cards come back until you master them.

Read the first 7 Dispensing Instrumentation Questions and Answers flashcards as text
  1. Which of the following tools is specifically designed to measure the thickness of a spectacle lens?

    Answer: Lens thickness gauge (caliper)

    A lens thickness gauge or caliper directly measures the physical thickness of a lens at any specified point.

  2. When using a lensmeter to measure a bifocal lens, the reading through the segment gives the:

    Answer: Near power only

    Reading through the bifocal segment on the lensometer yields the total near power; the add power is calculated as near power minus distance power.

  3. The marking dots applied by the lensometer's ink marking device indicate the:

    Answer: Optical center location and axis orientation

    The three dots marked by the lensometer's marker show the optical center (center dot) and the horizontal reference axis (outer dots) of the lens.

  4. A frame aligner (frame alignment tool) is most useful for correcting:

    Answer: Pantoscopic tilt and face-form angle of frames

    Frame aligners allow the dispenser to measure and correct pantoscopic tilt and face-form wrap to ensure proper fit and optical performance.

  5. In lensometry, the difference in drum readings between the two focal lines of a cylindrical lens indicates the:

    Answer: Cylinder power

    The numerical difference between the two power drum readings when each focal line is clear represents the cylinder power of the lens.

  6. A PD ruler (millimeter rule) is most appropriately used to measure:

    Answer: Monocular and binocular pupillary distances using a reference point

    A PD ruler placed across the bridge of the nose allows the dispenser to measure pupillary distances using a corneal reflex or pupil center as a reference.

  7. Which instrument would best verify that a finished lens has the correct amount of induced prism?

    Answer: Lensometer with prism reticle

    The lensometer can measure prism power and direction by observing displacement of the target from the center of the reticle.