← All ABA Flashcard Decks

Cochlear Implants & Implantable Hearing Devices Flashcards

21 cards from real ABA practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 20 Cochlear Implants & Implantable Hearing Devices flashcards as text
  1. Which structure in the cochlea is the target for cochlear implant electrode array insertion?

    Answer: Scala tympani

    The electrode array is inserted into the scala tympani because it is more accessible via the round window or cochleostomy and is closest to the spiral ganglion neurons.

  2. What is the minimum unaided pure-tone average (PTA) criterion typically used for adult cochlear implant candidacy in the better-hearing ear?

    Answer: 70 dB HL

    Current FDA guidelines and most centers require a PTA of 70 dB HL or greater in the better ear as part of adult cochlear implant candidacy.

  3. Which speech recognition score threshold is commonly used as part of adult CI candidacy evaluation?

    Answer: Less than 60% on aided CNC word recognition

    Adults scoring less than 60% on aided CNC monosyllabic word recognition tests in the best-aided condition typically meet speech perception criteria for CI candidacy.

  4. What does the term 'neural response telemetry' (NRT) or 'neural response imaging' (NRI) measure during cochlear implant programming?

    Answer: The electrically evoked compound action potential of the auditory nerve

    NRT/NRI measures the electrically evoked compound action potential (ECAP) of the auditory nerve, providing objective data to guide initial CI programming.

  5. In cochlear implant programming, what does 'T-level' (threshold level) refer to?

    Answer: The minimum level of electrical stimulation that produces a just-audible percept

    The T-level is the minimum electrical stimulation level at which the CI recipient just barely perceives sound, analogous to a threshold in behavioral audiometry.

  6. Which condition is considered a contraindication to cochlear implantation?

    Answer: Absent or severely deficient auditory nerve (cochlear nerve aplasia)

    Cochlear nerve aplasia or severe deficiency renders a cochlear implant ineffective because there are insufficient auditory nerve fibers to transmit the electrical signal to the brain.

  7. What is the primary advantage of a 'bimodal' fitting for a cochlear implant recipient?

    Answer: Combines a CI in one ear with a hearing aid in the opposite ear to improve speech perception, especially in noise

    Bimodal hearing pairs a CI with a conventional hearing aid in the contralateral ear, leveraging residual acoustic hearing to improve speech understanding in noise and music perception.

  8. Which imaging study is most commonly ordered prior to cochlear implant surgery to evaluate cochlear anatomy and patency?

    Answer: Computed tomography (CT) of the temporal bones

    High-resolution CT of the temporal bones is the standard pre-operative imaging to assess cochlear anatomy, detect ossification, malformations, and surgical landmarks.

  9. In CI programming, the 'C-level' (comfort level or MCL) represents:

    Answer: The loudest comfortable level of electrical stimulation before discomfort

    The C-level (also called MCL or M-level) is the maximum comfortable loudness level — the highest electrical stimulation that is loud but not uncomfortable for the CI user.

  10. Which cochlear implant processing strategy uses a 'n-of-m' approach where only the channels with the highest amplitude outputs are stimulated each cycle?

    Answer: Advanced Combination Encoder (ACE)

    ACE (Advanced Combination Encoder) selects the n channels with the greatest energy from a bank of m filters each cycle, providing a balance between spectral and temporal information.

  11. Electrode impedance checks in CI programming primarily assess:

    Answer: The integrity of the electrical circuit between the electrode and the tissue interface

    Impedance measurements evaluate the integrity of the electrode–tissue interface; abnormally high impedances may indicate an open circuit, and very low impedances can indicate a short circuit.

  12. A child receives a cochlear implant at age 18 months. According to research on critical periods, when should auditory rehabilitation be most intensively initiated?

    Answer: Immediately following device activation

    Auditory rehabilitation should begin immediately at device activation to take advantage of the sensitive period for auditory cortical development and maximize speech/language outcomes.

  13. What does 'stapedius reflex threshold' measurement provide in the context of cochlear implant programming?

    Answer: An objective way to estimate C-levels by detecting the reflexive middle ear muscle contraction

    Electrically evoked stapedial reflex thresholds (ESRT) can serve as objective landmarks for setting C-levels, as the reflex typically occurs near the upper end of the comfortable range.

  14. Bilateral cochlear implantation is most supported by evidence for which outcome benefit?

    Answer: Enhanced sound localization and speech understanding in noise

    Bilateral CI provides binaural hearing, which significantly improves sound source localization and speech understanding in background noise compared to unilateral implantation.

  15. Bone-anchored hearing aids (BAHA/OSIA) are most appropriate for which type of hearing loss?

    Answer: Conductive or mixed hearing loss, or single-sided deafness

    Bone-anchored devices transmit sound directly through the skull to the cochlea via bone conduction, making them ideal for conductive/mixed loss or single-sided deafness where air conduction is insufficient.

  16. During the cochlear implant evaluation process, which professional is typically responsible for device programming (mapping)?

    Answer: Audiologist certified in CI programming

    A certified audiologist specializing in cochlear implants performs the programming (mapping) sessions, setting T-levels, C-levels, and adjusting processing parameters to optimize the patient's hearing experience.

  17. What is 'residual hearing preservation' in the context of contemporary cochlear implantation?

    Answer: Using surgical techniques and electrode designs that minimize cochlear trauma to preserve low-frequency acoustic hearing post-implant

    Residual hearing preservation involves atraumatic surgical technique and flexible 'soft' electrode arrays to maintain low-frequency hearing, enabling 'electric-acoustic stimulation' (EAS/hybrid) after implantation.

  18. The primary reason cochlear implant electrodes are designed to be placed as close as possible to the modiolus (perimodiolar design) is to:

    Answer: Reduce current spread and improve frequency selectivity by being closer to the spiral ganglion neurons

    Perimodiolar electrodes are positioned closer to the spiral ganglion neurons, reducing the current required for stimulation and improving channel selectivity, which can enhance speech perception.

  19. Which outcome measure is most commonly used to assess speech perception benefit in adult cochlear implant recipients in the United States?

    Answer: Consonant-Nucleus-Consonant (CNC) word recognition in quiet

    CNC monosyllabic word recognition in quiet is the standard FDA-approved outcome measure used to assess post-implant speech perception benefit in adult CI recipients.

  20. Auditory neuropathy spectrum disorder (ANSD) is characterized by which audiologic finding that may still support cochlear implant candidacy?

    Answer: Present otoacoustic emissions (OAEs) with absent or abnormal auditory brainstem response (ABR)

    ANSD typically shows present OAEs (indicating outer hair cell function) but absent or severely abnormal ABR (indicating disrupted auditory nerve synchrony); CI can bypass the dyssynchronous neural pathway.

Written and edited by , founder of Practice Test Geeks, with AI assistance. Exam facts are checked against the official sources linked below. Editorial policy

Sources & references

Exam facts on this page (fees, format, dates, passing scores) are checked against these official sources; the date is our last check. Editorial policy

  1. audiology.org — Renewal, eligibility, score validity, test delivery, score release, passing score, time limit, registration, test dates · verified Nov 2026