A CSSD technician is packaging a multi-level instrument tray for steam sterilization. According to AAMI ST79, where should the internal chemical indicator be placed to present the greatest challenge to the sterilant?
-
A
On the top level, in the corner closest to the sterilizer door.
-
B
In the geometric center of each level of the tray.
-
C
Taped to the outside of the wrapped tray.
-
D
Between the two layers of the sterilization wrap.