โ† All HSPA Flashcard Decks

Processing of Patient Care Equipment Flashcards

6 cards from real HSPA practice questions. Tap to flip, then mark Knew It or Still Learning โ€” missed cards come back until you master them.

Read the first 6 Processing of Patient Care Equipment flashcards as text
  1. How should a blood pressure cuff be reprocessed per Spaulding classification?

    Answer: Low-level disinfection as a non-critical item

    Blood pressure cuffs contact only intact skin, classifying them as non-critical items requiring low-level disinfection.

  2. What reprocessing level is required for flexible GI endoscopes?

    Answer: High-level disinfection minimum, sterilization preferred

    Endoscopes contacting mucous membranes are semi-critical, requiring HLD minimum with sterilization preferred.

  3. How should reusable laryngoscope blades be reprocessed?

    Answer: High-level disinfection or sterilization

    Laryngoscope blades contact oropharyngeal mucous membranes, making them semi-critical devices.

  4. What special considerations apply to robotic surgical instrument reprocessing?

    Answer: Complex design requires specific manufacturer IFU, specialized cleaning tools, and careful reassembly with function testing

    Robotic instruments have complex multi-component designs requiring manufacturer-specific reprocessing.

  5. What is the correct procedure for C. difficile contaminated instruments?

    Answer: Standard cleaning followed by sterilization, as C. diff spores resist most disinfectants

    C. difficile spores resist most disinfectants but are killed by sterilization.

  6. How should single-use disposable devices be managed?

    Answer: Never reprocessed in-house; dispose after one use per FDA regulations unless using FDA-registered third-party reprocessor

    FDA prohibits in-house SUD reprocessing. Only FDA-registered third-party reprocessors may do so.