  1. Which infusion pump technology uses peristaltic action to propel fluid?

    Answer: Linear peristaltic pump using sequential finger-like compressions on the tubing

    Linear peristaltic pumps use a series of cam-driven fingers that sequentially compress the IV tubing in a wave-like motion, propelling fluid forward. This mechanism allows precise flow rate control using standard IV tubing. Rotary peristaltic pumps use a similar principle with rotating rollers compressing tubing around a curved track.

  2. What is the primary advantage of an elastomeric infusion device for home infusion therapy?

    Answer: No batteries, electricity, or programming required — provides constant-rate infusion through elastic recoil

    Elastomeric pumps contain a pre-filled elastomeric balloon inside a rigid container. The balloon's elastic recoil provides a constant driving pressure that pushes fluid through a flow restrictor at a predetermined rate. This design requires no power source or programming, making it ideal for simple home infusions like IV antibiotics.

  3. A nurse discovers that the wrong drug library profile was selected on a smart pump. The infusion has been running for 2 hours. What is the immediate action?

    Answer: Stop the infusion, assess the patient, verify the actual medication and dose delivered, correct the drug library selection, and notify the provider

    Using the wrong drug library profile means all safety dose limits were incorrect for the actual medication being infused. The nurse must stop the infusion to prevent further incorrect dosing, assess the patient for adverse effects, calculate the actual dose delivered, correct the profile, and notify the provider. An incident report should be filed.

  4. What is the purpose of an in-line air detector on an infusion pump?

    Answer: To detect air bubbles in the tubing and stop the infusion to prevent air embolism

    In-line air detectors use ultrasonic technology to continuously monitor the tubing for air bubbles. When air is detected above a threshold volume (typically 50-100 microliters for most pumps), the pump stops the infusion and alarms. This prevents venous air embolism, a potentially fatal complication of IV infusion.

  5. Which factor most affects the accuracy of an elastomeric infusion device?

    Answer: Temperature — cold temperatures slow the flow rate and warm temperatures increase it

    Temperature significantly affects elastomeric pump flow rates because the viscosity of the solution and the elasticity of the balloon change with temperature. Cold environments or refrigerated solutions flow slower than labeled rates, while warm environments increase flow rates. Patients should be educated to keep the device close to body temperature.

  6. What is the recommended frequency for updating smart pump drug libraries?

    Answer: At least quarterly, or whenever significant drug safety information changes

    Smart pump drug libraries should be updated at least quarterly and whenever there are changes to the formulary, new drug safety information, or analysis of override data reveals needed adjustments. Regular updates ensure dose limits reflect current evidence and practice. Compliance data from pump downloads should guide library refinements.