  1. Which type of washer-disinfector uses a thermal disinfection cycle as defined by the A0 concept?

    Answer: Thermal washer-disinfector (automated)

    Automated thermal washer-disinfectors use the A0 concept, which measures the lethal effect of heat on microorganisms at a reference temperature of 80°C.

  2. What hazard is associated with using high-pressure air to dry instruments in the decontamination area?

    Answer: It can aerosolize pathogens and contaminate the environment

    Using compressed air to blow off water can aerosolize infectious droplets, creating an inhalation hazard for CS staff.

  3. Which term describes the physical removal of foreign material (soil, blood, tissue) from instruments, which is NOT sterilization?

    Answer: Cleaning

    Cleaning is the physical removal of soil and foreign material; it reduces the microbial load but is not the same as sterilization or disinfection.

  4. Why is it important to keep instrument sets submerged below the water surface in an ultrasonic cleaner?

    Answer: To ensure cavitation action reaches all surfaces and to prevent aerosol generation

    Keeping instruments below the water surface ensures full cavitation exposure and prevents contaminated aerosols from being released into the air.

  5. A CS technician notices a brown discoloration on stainless steel instruments after washing. This is most likely caused by:

    Answer: Hard water minerals or inadequate rinsing

    Brown discoloration (staining) on stainless steel is most often caused by mineral deposits from hard water or inadequate rinsing after washing.

  6. When transporting soiled instruments from the OR to the decontamination area, containers must be:

    Answer: Closed, leak-proof containers labeled as biohazardous

    Soiled instruments must be transported in closed, leak-proof containers with biohazard labeling to prevent staff exposure and environmental contamination.

  7. The decontamination area should maintain a negative air pressure relative to adjacent areas. Why?

    Answer: To prevent contaminated air from flowing into clean processing areas

    Negative air pressure in the decontamination area ensures airflow moves inward, preventing potentially contaminated air from migrating to clean or sterile areas.