Medical Software Development & Integration Flashcards

7 cards from real CPMS practice questions. Tap to flip, then mark Knew It or Still Learning โ€” missed cards come back until you master them.

Read the first 7 Medical Software Development & Integration flashcards as text
  1. Which IEC 62304 software safety class requires the most rigorous development and maintenance processes?

    Answer: Class C

    IEC 62304 Class C applies to software where a failure could result in death or serious injury, requiring the highest level of rigor.

  2. In HL7 FHIR, what is the primary purpose of a 'Bundle' resource?

    Answer: Group multiple resources into a single transaction or message

    A FHIR Bundle groups multiple resources together, enabling batch operations, transactions, and document assembly.

  3. What is 'software of unknown provenance' (SOUP) in the context of IEC 62304?

    Answer: Pre-existing software not developed under a quality system for the current device

    SOUP refers to software items used in medical device software that were not developed under a quality system specifically for the device being developed.

  4. Which cryptographic standard is recommended for protecting ePHI in transit under HIPAA Security Rule guidance?

    Answer: TLS 1.2 or higher

    TLS 1.2 or higher is the current recommended standard for encrypting ePHI in transit, as older protocols are considered insecure.

  5. In a risk-based testing approach for medical software, what determines the depth of testing for a given module?

    Answer: The software's safety classification and associated risk

    Risk-based testing prioritizes test depth based on the software's safety class and the severity of potential failures.

  6. When integrating a medical device with an EHR via an HL7 v2 ADT message, what does the 'A01' event code signify?

    Answer: Patient admit

    ADT^A01 is the HL7 v2 event code for a patient admit or visit notification.

  7. Which software development lifecycle model is most commonly required by FDA for high-risk medical device software?

    Answer: A documented, risk-managed lifecycle with design controls

    FDA requires a documented software lifecycle with design controls per 21 CFR Part 820, regardless of whether the team uses agile or waterfall.