  1. Which IOL formula is generally preferred for eyes with axial lengths greater than 26 mm?

    Answer: Barrett Universal II

    Barrett Universal II performs well across a wide range of axial lengths, including long eyes over 26 mm.

  2. In immersion A-scan biometry, the probe is:

    Answer: Submerged in a saline-filled shell over the eye

    Immersion technique uses a saline-filled scleral shell to eliminate corneal compression artifacts.

  3. What does the term 'applanation' refer to in contact A-scan biometry?

    Answer: Compressing the cornea with the probe

    Applanation biometry involves the probe touching and slightly compressing the cornea, which can shorten axial length readings.

  4. The typical ultrasound frequency used in ophthalmic A-scan biometry is:

    Answer: 10 MHz

    Ophthalmic A-scans use 10 MHz ultrasound to achieve adequate resolution for ocular structures.

  5. Which structure produces the posterior lens spike in an A-scan waveform?

    Answer: Posterior lens capsule

    The posterior lens capsule generates the second large spike after the anterior lens spike in A-scan.

  6. A patient has a short axial length of 20.5 mm. Which IOL formula is most appropriate?

    Answer: Holladay 2 or Hoffer Q

    Hoffer Q and Holladay 2 are preferred for short eyes (axial length <22 mm) to minimize prediction errors.

  7. What artifact can cause a falsely low axial length reading on contact A-scan?

    Answer: Corneal compression by the probe

    Probe compression indents the cornea and shortens the measured axial length, leading to IOL power errors.