  1. Which action protects pipette accuracy when aspirating volatile reagents such as acetone?

    Answer: Pre-wet the pipette tip with the reagent before aspirating the target volume

    Pre-wetting saturates the dead air space inside the tip, reducing evaporation and improving volume accuracy for volatile liquids.

  2. How often should the HEPA filter in a Class II Type A2 biological safety cabinet typically be certified?

    Answer: Annually or whenever the cabinet is moved or serviced

    NSF/ANSI 49 requires annual certification of BSCs, plus recertification after relocation or major service, to verify airflow and HEPA filter integrity.

  3. What does the term 'carry-over' mean in the context of automated laboratory analyzers?

    Answer: Residual analyte from a previous sample contaminating the next sample result

    Carry-over occurs when residual analyte from a high-concentration sample contaminates the subsequent sample, falsely elevating or affecting that result.

  4. A freezer storing frozen plasma is found at –15°C instead of the required ≤–20°C. What is the primary concern?

    Answer: Labile coagulation factors in the plasma may have degraded

    Temperatures above –20°C accelerate degradation of labile coagulation factors such as Factor V and Factor VIII in frozen plasma.

  5. Why should a centrifuge rotor be inspected for cracks or corrosion before each use?

    Answer: A compromised rotor can fail catastrophically at high speed, causing injury

    Cracks or corrosion weaken the rotor structure and can lead to explosive rotor failure under centrifugal force, posing serious risk to laboratory personnel.

  6. When changing a reagent lot on a chemistry analyzer, which step must be completed before reporting patient results?

    Answer: Run quality control with the new lot and verify results are within acceptable ranges

    QC must pass with the new reagent lot before patient testing resumes, ensuring the new reagents perform acceptably.

  7. Which of the following best describes the purpose of an internal standard in an automated analyzer?

    Answer: A known-concentration substance added to correct for sample-to-sample variation in signal

    Internal standards compensate for matrix effects and instrument variation by normalizing analyte signal to the known signal of the internal standard in each sample.