  1. What temperature must a gravity-displacement autoclave reach to achieve sterilization of unwrapped instruments?

    Answer: 250°F (121°C) for 15–30 minutes

    Gravity-displacement autoclaves typically sterilize at 121°C (250°F) for 15–30 minutes depending on load configuration.

  2. Which of the following best describes the purpose of a biological indicator (spore test) in autoclave monitoring?

    Answer: It verifies that the sterilization process actually killed the most resistant bacterial spores

    Biological indicators use heat-resistant spores (e.g., Geobacillus stearothermophilus) to confirm that lethal conditions were actually achieved inside the load.

  3. In an electrology setting, which surface would be classified as a clinical contact surface requiring disinfection between clients?

    Answer: The treatment table and lamp handles touched during the procedure

    Clinical contact surfaces are those touched during treatment (treatment table, lamp, equipment controls) and must be disinfected between each client.

  4. A client presents with an active cold sore (herpes labialis) near the upper lip treatment area. What is the appropriate course of action?

    Answer: Postpone treatment to that area until the lesion has fully healed

    Active herpes lesions are highly contagious and electrology treatment over or near the site could spread the virus; the area must be avoided until fully healed.

  5. Which personal protective equipment (PPE) is considered mandatory for electrologists during all hair removal procedures?

    Answer: Gloves and eye protection

    At minimum, gloves and eye protection are required during electrology procedures to protect against bloodborne pathogen exposure from splatter.

  6. How long can Hepatitis B virus remain infectious on a dry environmental surface?

    Answer: Up to 7 days

    HBV can survive on dry surfaces for at least 7 days at room temperature, underscoring the need for thorough environmental disinfection.

  7. What is the primary purpose of using an EPA-registered disinfectant labeled 'tuberculocidal' in an electrology practice?

    Answer: It is the minimum required level for clinical contact surfaces under standard precautions

    Tuberculocidal disinfectants meet the minimum intermediate-level standard required for clinical contact surfaces and instruments that contact intact skin.