  1. A laboratory receives results with a coefficient of variation (CV) of 18% for a quality control sample where the acceptable limit is 10%. What is the most appropriate corrective action?

    Answer: Investigate calibration, reagents, and instrument performance before releasing results

    A CV exceeding the acceptable limit indicates imprecision, requiring root-cause investigation of calibration, reagents, and instrument performance before results can be released.

  2. Which statistical method is most appropriate for detecting a gradual upward drift in a laboratory's daily QC values over a two-week period?

    Answer: Linear regression trend analysis

    Linear regression trend analysis is used to detect systematic directional drift in QC data over time by modeling the slope of the data series.

  3. In laboratory reporting, a delta check flags a patient's sodium result as 168 mEq/L compared to 138 mEq/L three days prior. What is the primary purpose of this flag?

    Answer: To identify analytically erroneous results or specimen mix-ups by comparing sequential patient values

    Delta checks compare current results to previous results for the same patient to detect specimen mix-ups, transcription errors, or analytical errors.

  4. A lab consultant is reviewing a Levey-Jennings chart and notices 10 consecutive QC points all falling on the same side of the mean. Which Westgard rule is violated?

    Answer: 10-x rule

    The 10-x rule is triggered when 10 consecutive QC measurements fall on the same side of the mean, indicating systematic bias.

  5. When preparing a laboratory performance report for hospital administration, which metric best demonstrates the laboratory's contribution to patient throughput?

    Answer: Turnaround time (TAT) from order to result

    Turnaround time directly reflects how quickly the laboratory delivers actionable results, making it the key metric for demonstrating contribution to patient care throughput.

  6. A CLC is asked to validate a new automated hematology analyzer. Which statistical comparison is most appropriate for method correlation?

    Answer: Passing-Bablok regression

    Passing-Bablok regression is a non-parametric method used for method comparison that is not affected by outliers or non-normally distributed data.

  7. In an external proficiency testing (PT) program, a laboratory's result falls more than 3 standard deviations from the peer group mean. What is the required response?

    Answer: Perform a root-cause investigation and submit a corrective action plan

    An unacceptable PT result requires a formal root-cause investigation and corrective action plan to identify and resolve the source of the error.