Equipment Operation & Maintenance Flashcards
7 cards from real CHT practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 7 Equipment Operation & Maintenance flashcards as text
Which type of valve is most commonly used as a primary shut-off on hyperbaric chamber oxygen supply lines due to its clean-service compatibility?
Answer: Ball valve
Ball valves are preferred on oxygen lines because they open fully with a quarter-turn, leave no dead space, and are available in oxygen-clean configurations.
What is the minimum recommended frequency for changing the desiccant in a hyperbaric chamber's air drying system under normal operational use?
Answer: Per manufacturer schedule or when saturation indicator changes
Desiccant replacement frequency is driven by the manufacturer's schedule and color-indicator saturation, not a fixed calendar interval.
A hyperbaric technician notices the chamber's built-in breathing system (BIBS) mask check valve is leaking gas into the chamber environment. The IMMEDIATE corrective action is to:
Answer: Remove the mask from service and replace the check valve before next use
A leaking BIBS check valve enriches the chamber oxygen concentration and creates a fire hazard, so the mask must be removed from service immediately.
What is the purpose of a pressure-relief valve (PRV) on a hyperbaric chamber hull?
Answer: To prevent the chamber from exceeding its maximum allowable working pressure
The PRV is a safety device that automatically vents gas if chamber pressure exceeds the MAWP, protecting structural integrity.
When lubricating O-ring seals on a hyperbaric chamber door, which lubricant is acceptable for use?
Answer: Oxygen-compatible, silicone-based lubricant approved for the seal material
Only oxygen-compatible, material-compatible lubricants (typically silicone-based) may be used to prevent seal degradation and fire risk.
During multiplace chamber operations, the intercommunication system fails mid-treatment. Per NFPA 99 and UHMS guidelines, the operator should:
Answer: Establish an alternative communication method before continuing and follow facility emergency protocols
Facilities must have backup communication plans; an alternative means must be established and verified before continuing patient treatment.
The purpose of a hyperbaric chamber's ventilation system during an oxygen treatment is primarily to:
Answer: Maintain chamber oxygen concentration below flammable levels by purging with air
Ventilation dilutes and exhausts oxygen that leaks from BIBS masks, keeping the ambient O2 level below the fire-risk threshold (typically ≤23.5%).