ECMO Equipment & Technology Flashcards
6 cards from real CES practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 6 ECMO Equipment & Technology flashcards as text
Pre-membrane pressure (P1), post-membrane pressure (P2), and venous return pressure (P3) are used to calculate which ECMO parameter?
Answer: Transmembrane pressure gradient (ΔP), indicating oxygenator resistance
Transmembrane pressure (ΔP = P1 − P2) reflects resistance across the oxygenator — rising ΔP indicates developing oxygenator thrombosis or fiber fouling.
Which ECMO system feature allows simultaneous hemodialysis circuit integration?
Answer: An inline CRRT port or circuit tap point allowing CRRT machine connection to the ECMO circuit
CRRT machines can be connected to dedicated ports or tap points on the ECMO circuit, allowing concurrent renal replacement therapy without additional vascular access.
The rated flow of an ECMO oxygenator refers to:
Answer: The blood flow rate at which outflow blood oxygen saturation reaches 95% with inlet saturation of 65%
Oxygenator rated flow is defined as the blood flow at which outlet saturation reaches 95% given a standardized inlet saturation of 65%, indicating gas transfer efficiency.
Why is a roller pump generally NOT preferred over centrifugal pumps for neonatal ECMO circuits today?
Answer: Roller pumps cannot generate negative pressure and require gravity drainage, but centrifugal pumps offer better flow control and safety features in modern systems
While roller pumps have been used historically, centrifugal pumps offer better flow control, afterload sensitivity, and automated safety features that have made them preferred in current practice.
What does 'biocompatible circuit coating' (e.g., heparin-bonded tubing) accomplish in ECMO?
Answer: Reduces contact activation of coagulation and platelet adhesion on circuit surfaces, potentially reducing required systemic heparin
Heparin or other biocompatible coatings reduce surface thrombogenicity, decreasing contact activation and platelet consumption, often allowing lower systemic heparin requirements.
Which pressure monitor location in the ECMO circuit provides the best early warning of drainage cannula obstruction?
Answer: Pre-pump (venous inlet) negative pressure gauge
Pre-pump (inlet/venous) pressure monitors drainage line negative pressure — increasingly negative values indicate drainage obstruction or hypovolemia.