CER Study Guide 2026

Everything you need to pass the CER exam in one place: the exam format, every topic to study, real practice questions with explanations, flashcards, and full-length practice tests. Free, no sign-up needed.

📋 CER Exam Format at a Glance

100
Questions
120 min
Time Limit
70.00%
Passing Score

📚 CER Topics to Study (131)

Endoscope Reprocessing Fundamentals · 11 cardsPrinciples and Maintenance · 11 cardsAutomated Endoscope Reprocessors · 7 cardsAutomated Endoscope Reprocessors · 7 cardsAutomated Endoscope Reprocessors · 7 cardsDecontamination and Manual Cleaning · 7 cardsDecontamination and Manual Cleaning · 7 cardsDecontamination and Manual Cleaning · 7 cardsDecontamination and Manual Cleaning · 7 cardsDecontamination and Manual Cleaning Questions and Answers · 7 cardsDecontamination and Manual Cleaning Questions and Answers · 7 cardsDecontamination and Manual Cleaning Questions and Answers · 7 cardsEndoscope Design and Structure · 7 cardsEndoscope Design and Structure · 7 cardsEndoscope Design and Structure · 7 cardsEndoscope Design and Structure · 7 cardsEndoscope Design and Structure Questions and Answers · 7 cardsEndoscope Design and Structure Questions and Answers · 7 cardsEndoscope Design and Structure Questions and Answers · 7 cardsEndoscope Repair and Maintenance · 7 cardsEndoscope Repair and Maintenance · 7 cardsEndoscope Repair and Maintenance · 7 cardsEndoscope Repair and Maintenance · 7 cardsEndoscope Repair and Maintenance Questions and Answers · 7 cardsEndoscope Repair and Maintenance Questions and Answers · 7 cardsEndoscope Repair and Maintenance Questions and Answers · 7 cardsEndoscope Reprocessing Fundamentals · 7 cardsEndoscope Reprocessing Fundamentals · 7 cardsEndoscope Reprocessing Fundamentals · 7 cardsEndoscope Reprocessing Fundamentals · 7 cards

✍️ Sample CER Questions & Answers

1. What is the significance of maintaining adequate ventilation (minimum 10 ACH) in an endoscope reprocessing room?
To reduce staff exposure to toxic chemical vapors from HLD agents

HLD chemicals like glutaraldehyde release toxic fumes; OSHA and AORN require a minimum of 10 air changes per hour in reprocessing rooms to protect staff from vapor inhalation.

2. Which of the following is a key preventive maintenance practice performed by reprocessing staff to avoid costly repairs to the light guide connector?
Attaching the waterproof cap securely before any immersion in fluids.

The waterproof cap is essential for protecting the sensitive electrical contacts and components within the light guide connector from fluid invasion during cleaning and disinfection. Failure to secure the cap properly is a leading cause of major, expensive repairs due to fluid damage.

3. What is the primary purpose of performing biological monitoring (e.g., spore testing) on sterilization equipment used in endoscopy units?
To verify that the sterilization cycle is achieving the intended kill of the most resistant organisms

Biological indicators containing resistant spores confirm that the sterilization process achieved sufficient lethality to kill the hardest-to-kill organisms, validating cycle efficacy.

4. What corrective action is required when a sterilization cycle experiences a confirmed parameter deviation, such as an unplanned temperature drop during the cycle?
Quarantine the entire load and investigate before any items are released

Any cycle parameter deviation requires quarantine of the affected load until a thorough investigation is completed, because compromised parameters may result in inadequate sterilization.

5. How should a reprocessing 'near-miss' event be handled?
Documented in the facility's event reporting system and investigated to identify root cause and prevent recurrence

Near-miss events provide valuable learning opportunities and must be formally documented and investigated to prevent future failures that could harm patients.

6. What is the primary purpose of the water-resistant (leak-test) cap applied to the light guide connector before reprocessing?
To seal the connector so the scope can be fully submerged during reprocessing

The water-resistant cap seals the light guide connector, allowing complete immersion of the scope body in detergent and high-level disinfectant solutions.

🎯 Free CER Practice Tests

📖 CER Guides & Articles

Your CER Study Path
1. Learn with Flashcards → 2. Drill Practice Tests → 3. Take the Full Exam Simulation