  1. During pre-cleaning inspection, a technician notices the insertion tube has a slight spiral twist. What is the correct action?

    Answer: Immediately remove from service and notify the biomedical department

    A spiral twist in the insertion tube indicates internal channel or coil damage and the scope must be removed from service pending inspection by a qualified repair technician.

  2. Which tool is used to verify that suction and biopsy channel lumens meet the minimum diameter specification during scope inspection?

    Answer: Channel diameter gauge or brush compatibility chart

    A channel diameter gauge or the manufacturer's brush compatibility chart is used to confirm lumen diameter is within specification so correct-sized brushes can be selected.

  3. When performing a leak test on a videoscope, the endoscope must be tested:

    Answer: Dry, before any contact with water or cleaning solution

    The leak test must be performed on a dry scope before immersion to detect existing damage before fluids can enter and worsen internal damage.

  4. A steady stream of bubbles rising from the universal cord during a leak test most likely indicates:

    Answer: A breach in the outer sheath of the universal cord

    Continuous bubbling from the universal cord during a leak test indicates a breach in the outer sheath, allowing pressurized air to escape into water.

  5. What is the purpose of bending the insertion tube in all four directions during the functional inspection?

    Answer: To verify angulation wire and control wheel integrity

    Flexing in all directions tests the angulation wires, control wheels, and bending section for damage that may restrict full tip deflection.

  6. Which of the following findings during visual inspection would NOT typically require immediate removal from service?

    Answer: Minor surface scratches on the insertion tube with no breach

    Minor surface scratches that do not breach the outer sheath do not compromise scope integrity, though they should be documented and monitored.

  7. After reprocessing, an endoscope is found to have a dried, whitish residue inside the instrument channel. This most likely indicates:

    Answer: Inadequate flushing of cleaning solution or disinfectant

    Dried white residue in channels typically results from inadequate flushing, leaving detergent or disinfectant residue that can harm patients.